Performance of electronic portal imaging devices (EPIDs) used in radiotherapy:: Image quality and dose measurements

被引:47
|
作者
Cremers, F [1 ]
Frenzel, T [1 ]
Kausch, C [1 ]
Albers, D [1 ]
Schönborn, T [1 ]
Schmidt, R [1 ]
机构
[1] Univ Hamburg Hosp, Clin Radiotherapy & Radiooncol, D-20246 Hamburg, Germany
关键词
portal imaging; image quality; amorphous silicon array; fluoroscopic portal imaging system; response; dosimetry; radiation therapy;
D O I
10.1118/1.1688212
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of our Study was to compare the image and dosimetric quality of two different imaging systems. The first one is a fluoroscopic electronic portal imaging device (first generation), while the second is based on an amorphous silicon flat-panel array (second generation). The parameters describing image quality include spatial resolution [modulation transfer function (MTF)], noise [noise power spectrum (NPS)], and signal-to-noise transfer [detective quantum efficiency (DQE)]. The dosimetric measurements were compared with ionization chamber as well as with film measurements. The response of the flat-panel imager and the fluoroscopic-optical device was determined performing a two-step Monte Carlo simulation. All measurements were performed in a 6 MV linear accelerator photon beam. The resolution (MTF) of the fluoroscopic device (f(1/2) = 0.3 mm(-1)) is larger than of the amorphous silicon based system (f(1/2) = 0.21 mm(-1)), which is due to the missing backscattered photons and the smaller pixel size. The noise measurements (NPS) show the correlation of neighboring pixels of the amorphous silicon electronic portal imaging device, whereas the NPS of the fluoroscopic system is frequency independent. At zero spatial frequency the DQE of the flat-panel imager has a value of 0.008 (0.8%). Due to the minor frequency dependency this device may be almost x-ray quantum limited. Monte Carlo simulations verified these characteristics. For the fluoroscopic imaging system the DQE at low frequencies is about 0.0008 (0.08%) and degrades with higher frequencies. Dose measurements with the flat-panel imager revealed that images can only be directly converted to portal dose images, if scatter can be neglected. Thus objects distant to the detector (e.g., inhomogeneous dose distribution generated by a modificator) can be verified dosimetrically, while objects close to a detector (e.g., a patient) cannot be verified directly and must be scatter corrected prior to verification. This is justified by the response of the flat-panel imaging device revealing a strong dependency at low energies. (C) 2004 American Association of PhYsicists in Medicine.
引用
收藏
页码:985 / 996
页数:12
相关论文
共 50 条
  • [41] AUTOMATED QUALITY CONTROL OF THE MLC PERFORMANCE USING AN ELECTRONIC PORTAL IMAGING DEVICE
    De Brabandere, M.
    Hrbacek, J.
    Swinnen, A.
    Van den Heuvel, F.
    RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S381 - S381
  • [42] TREATMENT QUALITY ASSESSMENT USING IN VIVO ELECTRONIC PORTAL IMAGING DEVICE (EPID) DOSIMETRY IN RADIOTHERAPY
    Fuangrod, Todsaporn
    Simpson, John
    Greer, Peter
    ASIA-PACIFIC JOURNAL OF CLINICAL ONCOLOGY, 2016, 12 : 13 - 14
  • [43] Performance of an edge block used in a configuration detector: Image quality measurements
    Min, Jung-Whan
    Kim, Ki-Won
    Seoung, Youl-Hun
    Kim, Jung-Min
    Choi, In-Seok
    Jeong, Hoi-Woun
    Son, Soon-Yong
    Kim, Sang-Young
    Lee, Do-Wan
    Choe, Bo-Young
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2014, 64 (05) : 732 - 739
  • [44] Performance of an edge block used in a configuration detector: Image quality measurements
    Jung-Whan Min
    Ki-Won Kim
    Youl-Hun Seoung
    Jung-Min Kim
    In-Seok Choi
    Hoi-Woun Jeong
    Soon-Yong Son
    Sang-Young Kim
    Do-Wan Lee
    Bo-Young Choe
    Journal of the Korean Physical Society, 2014, 64 : 732 - 739
  • [45] Calibration of portal imaging devices for radiotherapy in-vivo dosimetry
    Piermattei, Angelo
    Cilla, Savino
    Fidanzio, Andrea
    Greco, Francesca
    Sabatino, Domenico
    Gargiulo, Laura
    Azario, Luigi
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 623 (02): : 829 - 831
  • [46] QUALITY ASSURANCE OF IMAGE GUIDED RADIOTHERAPY DEVICES
    Stock, M.
    RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S134 - S135
  • [47] Cone beam CT versus electronic portal imaging for daily image-guidance in prostate radiotherapy
    White, E
    Moseley, D
    Catton, C
    Wiltshire, K
    Sharpe, M
    Gospodarowicz, M
    Warde, P
    Jaffray, D
    RADIOTHERAPY AND ONCOLOGY, 2005, 76 : S4 - S4
  • [48] DESIGN OF COMPENSATORS FOR BREAST RADIOTHERAPY USING ELECTRONIC PORTAL IMAGING
    EVANS, PM
    HANSEN, VN
    MAYLES, WPM
    SWINDELL, W
    TORR, M
    YARNOLD, JR
    RADIOTHERAPY AND ONCOLOGY, 1995, 37 (01) : 43 - 54
  • [49] In silico study of the electronic portal imaging devices sensitivity to parotid glands shrinkage during radiotherapy for nasopharyngeal cases
    Alzahrani H.M.
    Journal of Umm Al-Qura University for Applied Sciences, 2024, 10 (2): : 279 - 289
  • [50] Assessment of dosimetrical performance in 11 Varian a-Si500 electronic portal imaging devices
    Kavuma, Awusi
    Glegg, Martin
    Currie, Garry
    Elliott, Alex
    PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (23): : 6893 - 6909