Comparison of multiplexed sensitivity encoding and single-shot echo-planar imaging for diffusion-weighted imaging of the liver

被引:16
|
作者
Kim, Yeun-Yoon [1 ]
Kim, Myeong-Jin [1 ]
Gho, Sung-Min [2 ]
Seo, Nieun [1 ]
机构
[1] Yonsei Univ, Severance Hosp, Dept Radiol, Coll Med, 50-1 Yonsei Ro, Seoul 03722, South Korea
[2] GE Healthcare, MR Collaborat & Dev, 416 Hangang Daero, Seoul 04637, South Korea
关键词
Diagnostic techniques and procedures; Magnetic resonance imaging; Comparative study; Artefacts; Signal-to-noise ratio; LESION DETECTION; MRI; EPI; ARTIFACTS; IMAGES; DWI;
D O I
10.1016/j.ejrad.2020.109292
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To compare multiplexed sensitivity encoding (MUSE) and conventional diffusion-weighted magnetic resonance imaging (cDWI) techniques in liver MRI. Methods: Fifty-nine patients who underwent both two-shot echo-planar DWI using MUSE and single-shot echoplanar cDWI at a 3.0-T MRI system were included. Qualitative parameters were independently evaluated by three radiologists, and quantitative parameters were calculated on the basis of region of interest measurements. Receiver operating characteristic curve analysis and McNemar's test were used to compare solid lesion characterization results and lesion detectability, respectively. Results: All reviewers found less image noise, sharper liver contours, milder susceptibility artifacts, and better lesion conspicuity in MUSE-DWI than in cDWI (reader average mean, 4.1-4.5 vs. 3.5-4.0; p < 0.05). The signalto-noise ratio (SNR) of the liver was significantly higher in MUSE-DWI than in cDWI (right lobe: mean, 9.39 vs. 8.10, p < 0.001; left lobe: mean, 8.34 vs. 7.19, p < 0.001), while the SNR of the lesion (mean, 23.72 vs. 23.88, p = 0.911) and lesion-to-liver contrast-to-noise ratio (mean, 14.65 vs. 15.41, p = 0.527) were comparable between MUSE-DWI and cDWI. Solid lesion characterization results were comparably accurate between MUSE-DWI and cDWI (reader average area under the receiver operating characteristic curve, 0.985 vs. 0.986, p = 0.480). The detectability of lesions was better in MUSE-DWI than in cDWI (reader consensus, 83.7 % [41/49] vs. 67.3 % [33/49], p = 0.021). Conclusion: MUSE-DWI can provide multi-shot liver DWI with less noise, fewer distortions, improved SNR of the liver, and better lesion detectability.
引用
收藏
页数:9
相关论文
共 50 条
  • [41] Characterization of focal liver lesions by ADC measurements using a respiratory triggered diffusion-weighted single-shot echo-planar MR imaging technique
    Bruegel, Melanie
    Holzapfel, Konstantin
    Gaa, Jochen
    Woertler, Klaus
    Waldt, Simone
    Kiefer, Berthold
    Stemmer, Alto
    Ganter, Carl
    Rummeny, Ernst J.
    EUROPEAN RADIOLOGY, 2008, 18 (03) : 477 - 485
  • [42] Characterization of focal liver lesions by ADC measurements using a respiratory triggered diffusion-weighted single-shot echo-planar MR imaging technique
    Melanie Bruegel
    Konstantin Holzapfel
    Jochen Gaa
    Klaus Woertler
    Simone Waldt
    Berthold Kiefer
    Alto Stemmer
    Carl Ganter
    Ernst J. Rummeny
    European Radiology, 2008, 18 : 477 - 485
  • [43] Comparing the clinical utility of single-shot, readout-segmented and zoomit echo-planar imaging in diffusion-weighted imaging of the kidney at 3 T
    Liu, Wenguang
    Liu, Hui
    Xie, Simin
    Masokano, Ismail Bilal
    Bai, Yu
    Wang, Xiao
    Zhong, Linhui
    Wu, Yi
    Nie, Jilin
    Zhou, Gaofeng
    Pei, Yigang
    Li, Wenzheng
    SCIENTIFIC REPORTS, 2022, 12 (01)
  • [44] Comparing the clinical utility of single-shot, readout-segmented and zoomit echo-planar imaging in diffusion-weighted imaging of the kidney at 3 T
    Wenguang Liu
    Hui Liu
    Simin xie
    Ismail Bilal Masokano
    Yu Bai
    Xiao Wang
    Linhui Zhong
    Yi Wu
    Jilin Nie
    Gaofeng Zhou
    Yigang Pei
    Wenzheng Li
    Scientific Reports, 12
  • [45] Optimising diffusion-weighted imaging in the abdomen and pelvis: comparison of image quality between monopolar and bipolar single-shot spin-echo echo-planar sequences
    Kyriazi, Stavroula
    Blackledge, Matthew
    Collins, David J.
    deSouza, Nandita M.
    EUROPEAN RADIOLOGY, 2010, 20 (10) : 2422 - 2431
  • [46] Optimising diffusion-weighted imaging in the abdomen and pelvis: comparison of image quality between monopolar and bipolar single-shot spin-echo echo-planar sequences
    Stavroula Kyriazi
    Matthew Blackledge
    David J. Collins
    Nandita M. deSouza
    European Radiology, 2010, 20 : 2422 - 2431
  • [47] Diffusion-weighted MRI in prostate cancer - comparison between single-shot fast spin echo and echo planar imaging sequences
    Kozlowski, Piotr
    Chang, Silvia D.
    Goldenberg, S. Larry
    MAGNETIC RESONANCE IMAGING, 2008, 26 (01) : 72 - 76
  • [48] Radiological diagnosis of acute stroke -: Comparison of conventional MR imaging, echo-planar diffusion-weighted imaging, and spin-echo diffusion-weighted imaging
    Geijer, B
    Brockstedt, S
    Lindgren, A
    Ståhlberg, F
    Norrving, B
    Holtås, S
    ACTA RADIOLOGICA, 1999, 40 (03) : 255 - 262
  • [49] Performance of Single-Shot Echo-Planar Imaging in Diffusion Tensor Imaging in Rat Sciatic Nerve Compared With Readout-Segmented Echo-Planar Imaging
    Chen, Yueyao
    Pan, Zhongxian
    Meng, Fanqi
    Li, Zhujing
    Hu, Yuanming
    Yu, Xuewen
    Gao, Jinyun
    Guo, Yihao
    Lyu, Hanqing
    Lin, Xiaofeng
    FRONTIERS IN NEUROSCIENCE, 2022, 16
  • [50] Focal liver masses: Characterization with diffusion-weighted echo-planar MR imaging
    Namimoto, T
    Yamashita, Y
    Sumi, S
    Tang, Y
    Takahashi, M
    RADIOLOGY, 1997, 204 (03) : 739 - 744