Parallel transmission RF pulse design with strict temperature constraints

被引:8
|
作者
Deniz, Cem M. [1 ,2 ,3 ,4 ,5 ]
Carluccio, Giuseppe [1 ,2 ,3 ]
Collins, Christopher [1 ,2 ,3 ,4 ]
机构
[1] NYU, Sch Med, Ctr Adv Imaging Innovat & Res CAI2R, New York, NY USA
[2] NYU, Sch Med, Dept Radiol, Bernard & Irene Schwartz Ctr Biomed Imaging, New York, NY USA
[3] NYU, Sch Med, Sackler Inst Grad Biomed Sci, New York, NY USA
[4] NYU, Tandon Sch Engn, NYU WIRELESS, Brooklyn, NY USA
[5] RF Test Labs Inc, New York, NY USA
关键词
parallel transmission; RF pulse design; temperature constraints; temperature correlation matrices; WHOLE-BODY SAR; IN-VIVO; ABSORPTION RATE; ANATOMICAL MODELS; LOCAL SAR; MRI; RESOLUTION; TISSUE; POWER; COIL;
D O I
10.1002/nbm.3694
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
RF safety in parallel transmission (pTx) is generally ensured by imposing specific absorption rate (SAR) limits during pTx RF pulse design. There is increasing interest in using temperature to ensure safety in MRI. In this work, we present a local temperature correlation matrix formalism and apply it to impose strict constraints on maximum absolute temperature in pTx RF pulse design for head and hip regions. Electromagnetic field simulations were performed on the head and hip of virtual body models. Temperature correlation matrices were calculated for four different exposure durations ranging between 6 and 24min using simulated fields and body-specific constants. Parallel transmission RF pulses were designed using either SAR or temperature constraints, and compared with each other and unconstrained RF pulse design in terms of excitation fidelity and safety. The use of temperature correlation matrices resulted in better excitation fidelity compared with the use of SAR in parallel transmission RF pulse design (for the 6min exposure period, 8.8% versus 21.0% for the head and 28.0% versus 32.2% for the hip region). As RF exposure duration increases (from 6min to 24min), the benefit of using temperature correlation matrices on RF pulse design diminishes. However, the safety of the subject is always guaranteed (the maximum temperature was equal to 39 degrees C). This trend was observed in both head and hip regions, where the perfusion rates are very different.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Radiofrequency Pulse Design in Parallel Transmission Under Strict Temperature Constraints
    Boulant, Nicolas
    Massire, Aurelien
    Amadon, Alexis
    Vignaud, Alexandre
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2014, 72 (03) : 679 - 688
  • [2] On Variant Strategies to Solve the Magnitude Least Squares Optimization Problem in Parallel Transmission Pulse Design and Under Strict SAR and Power Constraints
    Hoyos-Idrobo, A.
    Weiss, P.
    Massire, A.
    Amadon, A.
    Boulant, N.
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2014, 33 (03) : 739 - 748
  • [3] Parallel transmission RF pulse design for eddy current correction at ultra high field
    Zheng, Hai
    Zhao, Tiejun
    Qian, Yongxian
    Ibrahim, Tamer
    Boada, Fernando
    [J]. JOURNAL OF MAGNETIC RESONANCE, 2012, 221 : 139 - 146
  • [4] Local SAR in parallel transmission pulse design
    Lee, Joonsung
    Gebhardt, Matthias
    Wald, Lawrence L.
    Adalsteinsson, Elfar
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2012, 67 (06) : 1566 - 1578
  • [5] Signal-Domain Optimization Metrics for MPRAGE RF Pulse Design in Parallel Transmission at 7 Tesla
    Gras, V.
    Vignaud, A.
    Mauconduit, F.
    Luong, M.
    Amadon, A.
    Le Bihan, D.
    Boulant, N.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2016, 76 (05) : 1431 - 1442
  • [6] Minimal-SAR RF Pulse Optimization for Parallel Transmission in MRI
    Liu, Yinan
    Ji, Jim X.
    [J]. 2008 30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Vols 1-8, 2008, : 5774 - 5777
  • [7] Quick Reduction of Specific Absorption Rate Constraints in Parallel RF Excitation Pulse Design by Maximum Volume Inscribed Ellipsoids
    Majewski, Kurt
    [J]. CONCEPTS IN MAGNETIC RESONANCE PART A, 2019, 2019
  • [8] Magnetic Resonance RF Pulse Design by Optimal Control With Physical Constraints
    Rund, Armin
    Aigner, Christoph Stefan
    Kunisch, Karl
    Stollberger, Rudolf
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2018, 37 (02) : 461 - 472
  • [9] Optimal Design of Multiple-Channel RF Pulses Under Strict Power and SAR Constraints
    Brunner, David O.
    Pruessmann, Klaas P.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2010, 63 (05) : 1280 - 1291
  • [10] Parallel RF transmission in MRI
    Katscher, U
    Börnert, P
    [J]. NMR IN BIOMEDICINE, 2006, 19 (03) : 393 - 400