Parallel transmission RF pulse design with strict temperature constraints

被引:8
|
作者
Deniz, Cem M. [1 ,2 ,3 ,4 ,5 ]
Carluccio, Giuseppe [1 ,2 ,3 ]
Collins, Christopher [1 ,2 ,3 ,4 ]
机构
[1] NYU, Sch Med, Ctr Adv Imaging Innovat & Res CAI2R, New York, NY USA
[2] NYU, Sch Med, Dept Radiol, Bernard & Irene Schwartz Ctr Biomed Imaging, New York, NY USA
[3] NYU, Sch Med, Sackler Inst Grad Biomed Sci, New York, NY USA
[4] NYU, Tandon Sch Engn, NYU WIRELESS, Brooklyn, NY USA
[5] RF Test Labs Inc, New York, NY USA
关键词
parallel transmission; RF pulse design; temperature constraints; temperature correlation matrices; WHOLE-BODY SAR; IN-VIVO; ABSORPTION RATE; ANATOMICAL MODELS; LOCAL SAR; MRI; RESOLUTION; TISSUE; POWER; COIL;
D O I
10.1002/nbm.3694
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
RF safety in parallel transmission (pTx) is generally ensured by imposing specific absorption rate (SAR) limits during pTx RF pulse design. There is increasing interest in using temperature to ensure safety in MRI. In this work, we present a local temperature correlation matrix formalism and apply it to impose strict constraints on maximum absolute temperature in pTx RF pulse design for head and hip regions. Electromagnetic field simulations were performed on the head and hip of virtual body models. Temperature correlation matrices were calculated for four different exposure durations ranging between 6 and 24min using simulated fields and body-specific constants. Parallel transmission RF pulses were designed using either SAR or temperature constraints, and compared with each other and unconstrained RF pulse design in terms of excitation fidelity and safety. The use of temperature correlation matrices resulted in better excitation fidelity compared with the use of SAR in parallel transmission RF pulse design (for the 6min exposure period, 8.8% versus 21.0% for the head and 28.0% versus 32.2% for the hip region). As RF exposure duration increases (from 6min to 24min), the benefit of using temperature correlation matrices on RF pulse design diminishes. However, the safety of the subject is always guaranteed (the maximum temperature was equal to 39 degrees C). This trend was observed in both head and hip regions, where the perfusion rates are very different.
引用
收藏
页数:10
相关论文
共 50 条
  • [31] Additive angle method for fast large-tip-angle RF pulse design in parallel excitation
    Grissom, William A.
    Yip, Chun-Yu
    Wright, Steven M.
    Fessler, Jeffrey A.
    Noll, Douglas C.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2008, 59 (04) : 779 - 787
  • [32] Parallel RF transmission with eight channels at 3 Tesla
    Setsompop, Kawin
    Wald, Lawrence L.
    Alagappan, Vijayanand
    Gagoski, Borjan
    Hebrank, Franz
    Fontius, Ulrich
    Schmitt, Franz
    Adalsteinsson, Elfar
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2006, 56 (05) : 1163 - 1171
  • [33] Comprehensive RF safety concept for parallel transmission MR
    Graesslin, Ingmar
    Vernickel, Peter
    Boernert, Peter
    Nehrke, Kay
    Mens, Giel
    Harvey, Paul
    Katscher, Ulrich
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2015, 74 (02) : 589 - 598
  • [34] Parallel Transmission Pulse Design with Explicit Control for the Specific Absorption Rate in the Presence of Radiofrequency Errors
    Martin, Adrian
    Schiavi, Emanuele
    Eryaman, Yigitcan
    Herraiz, Joaquin L.
    Gagoski, Borjan
    Adalsteinsson, Elfar
    Wald, Lawrence L.
    Guerin, Bastien
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2016, 75 (06) : 2493 - 2504
  • [35] Sparsity in MRI RF excitation pulse design
    Zelinski, Adam C.
    Goyal, Vivek K.
    Adalsteinsson, Elfar
    Wald, Lawrence L.
    [J]. 2008 42ND ANNUAL CONFERENCE ON INFORMATION SCIENCES AND SYSTEMS, VOLS 1-3, 2008, : 252 - +
  • [36] MACROMOLECULAR ANISOTROPIC TEMPERATURE-FACTOR REFINEMENT WITH STRICT CONSTRAINTS.
    Kelly, Jennifer A.
    Yeates, Todd O.
    [J]. ACTA CRYSTALLOGRAPHICA A-FOUNDATION AND ADVANCES, 1996, 52 : C87 - C87
  • [37] Parallel Excitation in the Human Brain at 9.4 T Counteracting k-Space Errors With RF Pulse Design
    Wu, Xiaoping
    Vaughan, J. Thomas
    Ugurbil, Kamil
    Van de Moortele, Pierre-Francois
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2010, 63 (02) : 524 - 529
  • [38] RF TRANSMISSION LINE FOR LOW TEMPERATURE APPLICATIONS
    LAING, RA
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1965, 36 (07): : 1045 - &
  • [39] Pulse Sharpening and RF Generation Using Nonlinear Transmission Lines
    Silva Neto, L. P.
    Rossi, J. O.
    Barroso, J. J.
    [J]. 2015 SBMO/IEEE MTT-S INTERNATIONAL MICROWAVE AND OPTOELECTRONICS CONFERENCE (IMOC), 2015,
  • [40] Theoretical Analysis of RF Pulse Termination in Nonlinear Transmission Lines
    Nikoo, Mohammad Samizadeh
    Hashemi, Seyed Morad-Ali
    Farzaneh, Forouhar
    [J]. IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2018, 66 (11) : 4757 - 4764