Local SAR in parallel transmission pulse design

被引:71
|
作者
Lee, Joonsung [1 ]
Gebhardt, Matthias [2 ]
Wald, Lawrence L. [3 ,4 ]
Adalsteinsson, Elfar [1 ,4 ]
机构
[1] MIT, Dept Elect Engn & Comp Sci, Cambridge, MA 02139 USA
[2] Siemens Healthcare, Erlangen, Germany
[3] AA Martinos Ctr Biomed Imaging, Dept Radiol, Charlestown, MA USA
[4] MIT, Harvard MIT Hlth Sci & Technol, Cambridge, MA 02139 USA
关键词
local SAR; pTx RF design; parallel transmission; RF PULSES; ABSORPTION RATE; BIRDCAGE COIL; DOMAIN METHOD; HUMAN HEAD; 8; CHANNELS; EXCITATION; INHOMOGENEITY; MITIGATION; MODEL;
D O I
10.1002/mrm.23140
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The management of local and global power deposition in human subjects (specific absorption rate, SAR) is a fundamental constraint to the application of parallel transmission (pTx) systems. Even though the pTx and single channel have to meet the same SAR requirements, the complex behavior of the spatial distribution of local SAR for transmission arrays poses problems that are not encountered in conventional single-channel systems and places additional requirements on pTx radio frequency pulse design. We propose a pTx pulse design method which builds on recent work to capture the spatial distribution of local SAR in numerical tissue models in a compressed parameterization in order to incorporate local SAR constraints within computation times that accommodate pTx pulse design during an in vivo magnetic resonance imaging scan. Additionally, the algorithm yields a protocol-specific ultimate peak in local SAR, which is shown to bound the achievable peak local SAR for a given excitation profile fidelity. The performance of the approach was demonstrated using a numerical human head model and a 7 Tesla eight-channel transmit array. The method reduced peak local 10 g SAR by 1466% for slice-selective pTx excitations and 2D selective pTx excitations compared to a pTx pulse design constrained only by global SAR. The primary tradeoff incurred for reducing peak local SAR was an increase in global SAR, up to 34% for the evaluated examples, which is favorable in cases where local SAR constraints dominate the pulse applications. Magn Reson Med, 2011. (c) 2011 Wiley Periodicals, Inc.
引用
收藏
页码:1566 / 1578
页数:13
相关论文
共 50 条
  • [1] Minimal-SAR RF Pulse Optimization for Parallel Transmission in MRI
    Liu, Yinan
    Ji, Jim X.
    [J]. 2008 30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Vols 1-8, 2008, : 5774 - 5777
  • [2] SAR in parallel transmission
    Graesslin, I.
    Glaesel, D.
    Biederer, S.
    Schweser, F.
    Vernickel, P.
    Boernert, P.
    Annighoefer, B.
    Stahl, H.
    Dingemans, H.
    Mens, G.
    Harvey, P.
    Katscher, U.
    [J]. PIERS 2008 CAMBRIDGE, PROCEEDINGS, 2008, : 705 - +
  • [3] Parallel transmission RF pulse design with strict temperature constraints
    Deniz, Cem M.
    Carluccio, Giuseppe
    Collins, Christopher
    [J]. NMR IN BIOMEDICINE, 2017, 30 (05)
  • [4] Radiofrequency Pulse Design in Parallel Transmission Under Strict Temperature Constraints
    Boulant, Nicolas
    Massire, Aurelien
    Amadon, Alexis
    Vignaud, Alexandre
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2014, 72 (03) : 679 - 688
  • [5] On Variant Strategies to Solve the Magnitude Least Squares Optimization Problem in Parallel Transmission Pulse Design and Under Strict SAR and Power Constraints
    Hoyos-Idrobo, A.
    Weiss, P.
    Massire, A.
    Amadon, A.
    Boulant, N.
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2014, 33 (03) : 739 - 748
  • [6] Local Specific Absorption Rate (SAR), Global SAR, Transmitter Power, and Excitation Accuracy Trade-Offs in Low Flip-Angle Parallel Transmit Pulse Design
    Guerin, Bastien
    Gebhardt, Matthias
    Cauley, Steven
    Adalsteinsson, Elfar
    Wald, Lawrence L.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2014, 71 (04) : 1446 - 1457
  • [7] System and SAR characterization in parallel RF transmission
    Zhu, Yudong
    Alon, Leeor
    Deniz, Cem M.
    Brown, Ryan
    Sodickson, Daniel K.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2012, 67 (05) : 1367 - 1378
  • [8] Adapted RF pulse design for SAR reduction in parallel excitation with experimental verification at 9.4 T
    Wu, Xiaoping
    Akgun, Can
    Vaughan, J. Thomas
    Andersen, Peter
    Strupp, John
    Ugurbil, Kamil
    Van de Moortele, Pierre-Francois
    [J]. JOURNAL OF MAGNETIC RESONANCE, 2010, 205 (01) : 161 - 170
  • [9] A Mathematical Analysis of Clustering-Free Local SAR Compression Algorithms for MRI Safety in Parallel Transmission
    Gras, V.
    Boulant, N.
    Luong, M.
    Morel, L.
    Le Touz, N.
    Adam, J. -P.
    Joly, J. -C.
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2024, 43 (02) : 714 - 722
  • [10] Parallel transmission RF pulse design for eddy current correction at ultra high field
    Zheng, Hai
    Zhao, Tiejun
    Qian, Yongxian
    Ibrahim, Tamer
    Boada, Fernando
    [J]. JOURNAL OF MAGNETIC RESONANCE, 2012, 221 : 139 - 146