Radiation dose in neuroangiography using image noise reduction technology: a population study based on 614 patients

被引:73
|
作者
Soderman, Michael [1 ,2 ]
Mauti, Maria [3 ]
Boon, Sjirk [3 ]
Omar, Artur [4 ]
Marteinsdottir, Maria [4 ]
Andersson, Tommy [1 ,2 ]
Holmin, Staffan [1 ,2 ]
Hoornaert, Bart [3 ]
机构
[1] Karolinska Univ Hosp Solna, Karolinska Inst, Dept Clin Neurosci, S-17176 Stockholm, Sweden
[2] Karolinska Univ Hosp Solna, Dept Neuroradiol, S-17176 Stockholm, Sweden
[3] Philips Healthcare, Best, Netherlands
[4] Karolinska Univ Hosp, Sect Imaging Phys, Dept Med Phys, Stockholm, Sweden
关键词
Angiography; Radiation dose; Radiation physics; Imaging technology; Interventional neuroradiology; INTERVENTIONAL NEURORADIOLOGY; EXPOSURE;
D O I
10.1007/s00234-013-1276-0
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The purpose of this study was to quantify the reduction in patient radiation dose by X-ray imaging technology using image noise reduction and system settings for neuroangiography and to assess its impact on the working habits of the physician. Radiation dose data from 190 neuroangiographies and 112 interventional neuroprocedures performed with state-of-the-art image processing and reference system settings were collected for the period January-June 2010. The system was then configured with extra image noise reduction algorithms and system settings, which enabled radiation dose reduction without loss of image quality. Radiation dose data from 174 neuroangiographies and 138 interventional neuroprocedures were collected for the period January-June 2012. Procedures were classified as diagnostic or interventional. Patient radiation exposure was quantified using cumulative dose area product and cumulative air kerma. Impact on working habits of the physician was quantified using fluoroscopy time and number of digital subtraction angiography (DSA) images. The optimized system settings provided significant reduction in dose indicators versus reference system settings (p < 0.001): from 124 to 47 Gy cm(2) and from 0.78 to 0.27 Gy for neuroangiography, and from 328 to 109 Gy cm(2) and from 2.71 to 0.89 Gy for interventional neuroradiology. Differences were not significant between the two systems with regard to fluoroscopy time or number of DSA images. X-ray imaging technology using an image noise reduction algorithm and system settings provided approximately 60% radiation dose reduction in neuroangiography and interventional neuroradiology, without affecting the working habits of the physician.
引用
收藏
页码:1365 / 1372
页数:8
相关论文
共 50 条
  • [1] Radiation dose in neuroangiography using image noise reduction technology: a population study based on 614 patients
    Michael Söderman
    Maria Mauti
    Sjirk Boon
    Artur Omar
    María Marteinsdóttir
    Tommy Andersson
    Staffan Holmin
    Bart Hoornaert
    Neuroradiology, 2013, 55 : 1365 - 1372
  • [2] Impact of tube angulation on radiation dose using image noise reduction technology
    Abdo, M. Haj
    Langenbrink, L.
    Michaelsen, J.
    Wirtz, M.
    Altiok, E.
    Reimann, D.
    Hoffmann, R.
    EUROPEAN HEART JOURNAL, 2020, 41 : 2451 - 2451
  • [3] Significant reduction of radiation dose using image noise reduction technology in coronary angiography and intervention
    Kastrati, M.
    Langenbrink, L.
    Piatkowski, M.
    Michaelsen, J.
    Hoffmann, R.
    EUROPEAN HEART JOURNAL, 2016, 37 : 878 - 878
  • [4] Reducing Radiation Dose in Coronary Angiography and Angioplasty Using Image Noise Reduction Technology
    Kastrati, Mirlind
    Langenbrink, Lukas
    Piatkowski, Michal
    Michaelsen, Jochen
    Reimann, Doris
    Hoffmann, Rainer
    AMERICAN JOURNAL OF CARDIOLOGY, 2016, 118 (03): : 353 - 356
  • [5] Image Quality of the Coronary Angiography with Noise Reduction Technology to Decrease the Radiation Dose
    Kumashiro, Masayuki
    Kataoka, Takahiro
    Yokota, Shinobu
    Nakagawa, Shinobu
    Otsuki, Keisuke
    Miyake, Nobuyuki
    Osumi, Shinji
    Yamaoka, Kiyonori
    ACTA MEDICA OKAYAMA, 2018, 72 (02) : 153 - 164
  • [6] Radiation Dose Reduction in Structural Heart Disease Interventions Using Noise Reduction Technology
    Choi, Jeong Won
    Atsina, Kwame Bodor-Tsia
    Stripe, Benjamin Ray
    Wong, Garrett Bradley
    Smith, Thomas Wayne
    Rogers, Jason Hsieh
    Li, Chin-Shang
    Singh, Gagan
    Boyd, Walter Douglas
    Perry, Paul
    Southard, Jeffrey Allen
    JACC-CARDIOVASCULAR INTERVENTIONS, 2018, 11 (04) : S65 - S65
  • [7] Image Noise and Radiation Dose Reduction in Spectral CT
    Leng, S.
    McCollough, C.
    Fletcher, J.
    Yu, L.
    Mistretta, C.
    MEDICAL PHYSICS, 2010, 37 (06) : 3408 - +
  • [8] Reducing Patient Radiation Dose With Image Noise Reduction Technology in Transcatheter Aortic Valve Procedures
    Lauterbach, Michael
    Hauptmann, Karl Eugen
    AMERICAN JOURNAL OF CARDIOLOGY, 2016, 117 (05): : 834 - 838
  • [9] Significant reduction of radiation dosage in device implantation procedures using image noise reduction technology
    Hoffmann, R.
    Langenbrink, L.
    Kastrati, M.
    Piatkowski, M.
    Reimann, D.
    Michaelsen, J.
    EUROPEAN HEART JOURNAL, 2017, 38 : 692 - 693
  • [10] Radiation dose reduction utilizing noise reduction technology during uterine artery embolization: a pilot study
    Durrani, Raisa J.
    Fischman, Aaron M.
    van der Bom, Imramsjah M. J.
    Kim, Edward
    Nowakowski, F. Scott
    Patel, Rahul S.
    Lookstein, Robert A.
    CLINICAL IMAGING, 2016, 40 (03) : 378 - 381