Radiation dose in neuroangiography using image noise reduction technology: a population study based on 614 patients

被引:73
|
作者
Soderman, Michael [1 ,2 ]
Mauti, Maria [3 ]
Boon, Sjirk [3 ]
Omar, Artur [4 ]
Marteinsdottir, Maria [4 ]
Andersson, Tommy [1 ,2 ]
Holmin, Staffan [1 ,2 ]
Hoornaert, Bart [3 ]
机构
[1] Karolinska Univ Hosp Solna, Karolinska Inst, Dept Clin Neurosci, S-17176 Stockholm, Sweden
[2] Karolinska Univ Hosp Solna, Dept Neuroradiol, S-17176 Stockholm, Sweden
[3] Philips Healthcare, Best, Netherlands
[4] Karolinska Univ Hosp, Sect Imaging Phys, Dept Med Phys, Stockholm, Sweden
关键词
Angiography; Radiation dose; Radiation physics; Imaging technology; Interventional neuroradiology; INTERVENTIONAL NEURORADIOLOGY; EXPOSURE;
D O I
10.1007/s00234-013-1276-0
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The purpose of this study was to quantify the reduction in patient radiation dose by X-ray imaging technology using image noise reduction and system settings for neuroangiography and to assess its impact on the working habits of the physician. Radiation dose data from 190 neuroangiographies and 112 interventional neuroprocedures performed with state-of-the-art image processing and reference system settings were collected for the period January-June 2010. The system was then configured with extra image noise reduction algorithms and system settings, which enabled radiation dose reduction without loss of image quality. Radiation dose data from 174 neuroangiographies and 138 interventional neuroprocedures were collected for the period January-June 2012. Procedures were classified as diagnostic or interventional. Patient radiation exposure was quantified using cumulative dose area product and cumulative air kerma. Impact on working habits of the physician was quantified using fluoroscopy time and number of digital subtraction angiography (DSA) images. The optimized system settings provided significant reduction in dose indicators versus reference system settings (p < 0.001): from 124 to 47 Gy cm(2) and from 0.78 to 0.27 Gy for neuroangiography, and from 328 to 109 Gy cm(2) and from 2.71 to 0.89 Gy for interventional neuroradiology. Differences were not significant between the two systems with regard to fluoroscopy time or number of DSA images. X-ray imaging technology using an image noise reduction algorithm and system settings provided approximately 60% radiation dose reduction in neuroangiography and interventional neuroradiology, without affecting the working habits of the physician.
引用
收藏
页码:1365 / 1372
页数:8
相关论文
共 50 条
  • [31] Reduction of radiation dose during facet joint injection using the new image guidance system SabreSource™: a prospective study in 60 patients
    Dirk Proschek
    K. Kafchitsas
    M. A. Rauschmann
    A. A. Kurth
    T. J. Vogl
    Florian Geiger
    European Spine Journal, 2009, 18
  • [32] Radiation dose reduction in CT of the brain: can advanced noise filtering compensate for loss of image quality?
    Siemund, Roger
    Love, A.
    van Westen, D.
    Stenberg, L.
    Petersen, C.
    Bjorkman-Burtscher, I. M.
    ACTA RADIOLOGICA, 2012, 53 (04) : 468 - 472
  • [33] Hepatic CT perfusion measurements: A feasibility study for radiation dose reduction using new image reconstruction method
    Negi, Noriyuki
    Yoshikawa, Takeshi
    Ohno, Yoshiharu
    Somiya, Yuichiro
    Sekitani, Toshinori
    Sugihara, Naoki
    Koyama, Hisanobu
    Kanda, Tomonori
    Kanata, Naoki
    Murakami, Tohru
    Kawamitsu, Hideaki
    Sugimura, Kazuro
    EUROPEAN JOURNAL OF RADIOLOGY, 2012, 81 (11) : 3048 - 3054
  • [34] Evaluation of the image quality and dose reduction in digital radiography with an advanced spatial noise reduction algorithm in pediatric patients
    Oh, Saelin
    Kim, Ji Hye
    Yoo, So-Young
    Jeon, Tae Yeon
    Kim, Yu Jin
    EUROPEAN RADIOLOGY, 2021, 31 (12) : 8937 - 8946
  • [35] Evaluation of the image quality and dose reduction in digital radiography with an advanced spatial noise reduction algorithm in pediatric patients
    Saelin Oh
    Ji Hye Kim
    So-Young Yoo
    Tae Yeon Jeon
    Yu Jin Kim
    European Radiology, 2021, 31 : 8937 - 8946
  • [36] Improvement of Image Registration Using Total-Variation Based Noise Reduction Algorithms for Low-Dose CBCT
    Mukherjee, S.
    Farr, J.
    Merchant, T.
    Yao, W.
    MEDICAL PHYSICS, 2016, 43 (06) : 3388 - 3388
  • [37] Dramatic Noise Reduction and Potential Radiation Dose Reduction in Breast Cone-Beam CT Imaging Using Prior Image Constrained Compressed Sensing (PICCS)
    Qi, Z.
    Huang, S.
    Nett, B.
    Tang, J.
    Yang, K.
    Boone, J.
    Chen, G.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [38] Radiation Dose Reduction Using Real-Time Image Processing in Interventional Radiology
    Kanal, K.
    Moirano, J.
    Zamora, D.
    Stewart, B.
    MEDICAL PHYSICS, 2016, 43 (06) : 3347 - 3347
  • [39] NOISE REDUCTION IN LOW-DOSE X-RAY FLUOROSCOPY FOR IMAGE-GUIDED RADIATION THERAPY
    Jing Wang
    Lei Zhu
    Lei Xing
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 74 (02): : 637 - 643
  • [40] Effects of Image Receptor Technology and Dose Reduction Software On Radiation Dose Estimates for Fluoroscopically-Guided Interventional (FGI) Procedures
    Merritt, Z.
    Dave, J.
    Eschelman, D.
    Gonsalves, C.
    MEDICAL PHYSICS, 2016, 43 (06) : 3667 - 3667