Radiation dose in neuroangiography using image noise reduction technology: a population study based on 614 patients

被引:73
|
作者
Soderman, Michael [1 ,2 ]
Mauti, Maria [3 ]
Boon, Sjirk [3 ]
Omar, Artur [4 ]
Marteinsdottir, Maria [4 ]
Andersson, Tommy [1 ,2 ]
Holmin, Staffan [1 ,2 ]
Hoornaert, Bart [3 ]
机构
[1] Karolinska Univ Hosp Solna, Karolinska Inst, Dept Clin Neurosci, S-17176 Stockholm, Sweden
[2] Karolinska Univ Hosp Solna, Dept Neuroradiol, S-17176 Stockholm, Sweden
[3] Philips Healthcare, Best, Netherlands
[4] Karolinska Univ Hosp, Sect Imaging Phys, Dept Med Phys, Stockholm, Sweden
关键词
Angiography; Radiation dose; Radiation physics; Imaging technology; Interventional neuroradiology; INTERVENTIONAL NEURORADIOLOGY; EXPOSURE;
D O I
10.1007/s00234-013-1276-0
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The purpose of this study was to quantify the reduction in patient radiation dose by X-ray imaging technology using image noise reduction and system settings for neuroangiography and to assess its impact on the working habits of the physician. Radiation dose data from 190 neuroangiographies and 112 interventional neuroprocedures performed with state-of-the-art image processing and reference system settings were collected for the period January-June 2010. The system was then configured with extra image noise reduction algorithms and system settings, which enabled radiation dose reduction without loss of image quality. Radiation dose data from 174 neuroangiographies and 138 interventional neuroprocedures were collected for the period January-June 2012. Procedures were classified as diagnostic or interventional. Patient radiation exposure was quantified using cumulative dose area product and cumulative air kerma. Impact on working habits of the physician was quantified using fluoroscopy time and number of digital subtraction angiography (DSA) images. The optimized system settings provided significant reduction in dose indicators versus reference system settings (p < 0.001): from 124 to 47 Gy cm(2) and from 0.78 to 0.27 Gy for neuroangiography, and from 328 to 109 Gy cm(2) and from 2.71 to 0.89 Gy for interventional neuroradiology. Differences were not significant between the two systems with regard to fluoroscopy time or number of DSA images. X-ray imaging technology using an image noise reduction algorithm and system settings provided approximately 60% radiation dose reduction in neuroangiography and interventional neuroradiology, without affecting the working habits of the physician.
引用
收藏
页码:1365 / 1372
页数:8
相关论文
共 50 条
  • [21] Denoising Technology for Radiation Image with Poisson Noise Based on Shearlet Transform
    Xu, Yuting
    Wu, Zhifang
    Wang, Qiang
    Hou, Yongming
    Zhao, Bin
    Liu, Xinxia
    Yuanzineng Kexue Jishu/Atomic Energy Science and Technology, 2022, 56 (03): : 577 - 584
  • [22] Effect of acoustic noise reduction technology on image quality: a multivendor study
    Yamashiro, Takanobu
    Takatsu, Yasuo
    Morita, Kosuke
    Nakamura, Masafumi
    Yukimura, Yoshihiro
    Nakajima, Kazuhiro
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2023, 16 (02) : 235 - 243
  • [23] Effect of acoustic noise reduction technology on image quality: a multivendor study
    Takanobu Yamashiro
    Yasuo Takatsu
    Kosuke Morita
    Masafumi Nakamura
    Yoshihiro Yukimura
    Kazuhiro Nakajima
    Radiological Physics and Technology, 2023, 16 : 235 - 243
  • [24] Reduction of SPECT MPI Radiation Dose Using Contemporary Protocols and Technology
    Thompson, Randall C.
    O'Keefe, James H.
    McGhie, A. Iain
    Bybee, Kevin A.
    Thompson, Elaine C.
    Bateman, Timothy M.
    JACC-CARDIOVASCULAR IMAGING, 2018, 11 (02) : 282 - 283
  • [25] Radiation dose reduction and improvement of image quality in digital chest radiography by new spatial noise reduction algorithm
    Lee, Wonje
    Lee, Seungho
    Chong, Semin
    Lee, Kyungmin
    Lee, Jongha
    Choi, Jae Chol
    Lim, Changwon
    PLOS ONE, 2020, 15 (02):
  • [26] Noise reduction angiographic imaging technology reduces radiation dose during bronchial artery embolization
    Spink, C.
    Avanesov, M.
    Schmidt, T.
    Grass, M.
    Schoen, G.
    Adam, G.
    Koops, A.
    Ittrich, H.
    Bannas, P.
    EUROPEAN JOURNAL OF RADIOLOGY, 2017, 97 : 115 - 118
  • [27] Radiation dose reduction using deep learning-based image reconstruction for a low-dose chest computed tomography protocol: a phantom study
    Jung, Yunsub
    Hur, Jin
    Han, Kyunghwa
    Imai, Yasuhiro
    Hong, Yoo Jin
    Im, Dong Jin
    Lee, Kye Ho
    Desnoyers, Melissa
    Thomsen, Brian
    Shigemasa, Risa
    Um, Kyounga
    Jang, Kyungeun
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2023, 13 (03) : 1937 - +
  • [28] Reduction of radiation dose in pediatric patients using pulsed fluoroscopy
    Hernandez, RJ
    Goodsitt, MM
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1996, 167 (05) : 1247 - 1253
  • [29] Study on Noise Reduction Method of low Illuminance Image Based on Noise Feature
    Enze, Yuan
    Miura, Yasuyuki
    2022 IEEE INTERNATIONAL CONFERENCE ON CONSUMER ELECTRONICS - TAIWAN, IEEE ICCE-TW 2022, 2022, : 107 - 108
  • [30] Reduction of radiation dose during facet joint injection using the new image guidance system SabreSource™: a prospective study in 60 patients
    Proschek, Dirk
    Kafchitsas, K.
    Rauschmann, M. A.
    Kurth, A. A.
    Vogl, T. J.
    Geiger, Florian
    EUROPEAN SPINE JOURNAL, 2009, 18 (04) : 546 - 553