Comparing the clinical utility of single-shot echo-planar imaging and readout-segmented echo-planar imaging in diffusion-weighted imaging of the liver at 3 tesla

被引:5
|
作者
Xie, Simin [1 ]
Masokano, Ismail Bilal [1 ]
Liu, Wenguang [1 ]
Long, Xueying [1 ]
Li, Guijin [2 ]
Pei, Yigang [1 ,3 ]
Li, Wenzheng [1 ]
机构
[1] Cent South Univ, Xiangya Hosp, Dept Radiol, Changsha 410008, Hunan, Peoples R China
[2] Siemens Healthcare Ltd, Guangzhou Branch, Guangzhou 510620, Peoples R China
[3] Cent South Univ, Xiangya Hosp, Changsha 410008, Hunan, Peoples R China
基金
中国博士后科学基金;
关键词
Hepatic diffusion-weighted imaging; Single-shot echo-planar imaging (EPI); Readout-segmented EPI; NT-DWI; Clinical utility;
D O I
10.1016/j.ejrad.2020.109472
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To compare the clinical utility of single-shot echo-planar imaging (SS-EPI) using different breathing schemes and readout-segmented EPI (RS-EPI) in the repeatability of apparent diffusion coefficient (ADC) measurements, signal-to-noise ratio (SNR) and image quality. Methods: In this institutional review board-approved prospective study, hepatic DWIs (b = 50, 300, 600 s/mm(2)) were performed in 22 volunteers on 3.0 T MRI using SS-EPI with free-breathing diffusion-weighted imaging (FB-DWI), breath-hold (BH-DWI), respiratory-triggered (RT-DWI) and navigator-triggered (NT-DWI), and readout-segmented EPI (RS-DWI). ADC and surrogate SNR (sSNR) were measured in nine anatomic locations in the right lobe, and image quality was assessed on all FB-DWI, BH-DWI, RT-DWI, NT-DWI, and RS-DWI sequences. The sequence with the optimal clinical utility was decided by systematically comparing the ADC repeatability, sSNR and image quality of the above DWIs. Results: In all the five sequences, NT-DWI had the most reliable intra-observer agreement (intraclass correlation coefficient (ICC): 0.900-0.922; all P > 0.05), and a better interobserver agreement (ICC: 0.853 0.960; all p > 0.05) than RS-DWI (ICC:0.881-0.916; some P < 0.05). NT-DWI had the best ADC repeatability in the nine locations (mean ADC absolute differences: 38.47-56.38 x 10(-6) mm(2)/s, limits of agreement (LOA): 17.33-22.52 x 10(-6) mm(2)/s). Also, NT-DWI had the highest sSNR (Reader 1: 50.58 +/- 20.11 (Superior), 74.06 +/- 28.37 (Central), 80.99 +/- 38.11(Inferior)); Reader 2: 48.07 +/- 23.92 (Superior), 68.23 +/- 32.91 (Central), 76.78 +/- 33.07 (Inferior)) in three representative sections except for RS-DWI. Furthermore, NT-DWI had a better image quality than RS-DWI (P < 0.05) and was superior to FB-DWI and BH-DWI in sharpness of the liver (at b = 300 s/mm(2)) (P < 0.05) Conclusion: RS-DWI has the best SNR. However, NT-DWI can provide sufficient SNR, excellent image quality, and the best ADC repeatability on 3.0 T MRI. It is thus the recommended sequence for the clinical application of hepatic DWI.
引用
收藏
页数:10
相关论文
共 50 条
  • [21] Preoperative Localization of Parathyroid Adenomas with Diffusion MR Imaging: Readout-segmented versus Single-shot Echo-planar Imaging
    Yamamoto, Hirotaka
    Iima, Mami
    Kishimoto, Yo
    Kishimoto, Ayami Ohno
    Koyasu, Sho
    Yamamoto, Akira
    Kawai, Yoshitaka
    Yoshizawa, Akira
    Omori, Koichi
    MAGNETIC RESONANCE IN MEDICAL SCIENCES, 2023, 22 (01) : 79 - 85
  • [22] Comparison of Readout-Segmented Echo-Planar Imaging and Single-Shot TSE DWI for Cholesteatoma Diagnostics
    Wiesmueller, M.
    Wuest, W.
    May, M. S.
    Ellmann, S.
    Heiss, R.
    Saake, M.
    Janka, R.
    Uder, M.
    Laun, F. B.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2021, 42 (07) : 1305 - 1312
  • [23] Comparison of multiplexed sensitivity encoding and single-shot echo-planar imaging for diffusion-weighted imaging of the liver
    Kim, Yeun-Yoon
    Kim, Myeong-Jin
    Gho, Sung-Min
    Seo, Nieun
    EUROPEAN JOURNAL OF RADIOLOGY, 2020, 132
  • [24] Editorial for "Intraobserver and Interobserver Reproducibility of Breast Diffusion-Weighted Imaging Quantitative Parameters: Readout-Segmented vs. Single-Shot Echo-Planar Imaging"
    Loubrie, Stephane
    Batasin, Summer
    Rakow-Penner, Rebecca
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2023, 58 (06) : 1737 - 1738
  • [25] Differential diagnosis of pancreatic cancer by single-shot echo-planar imaging diffusion-weighted imaging
    Hong, Ben-Zu
    Li, Xin-Feng
    Lin, Jian-Qing
    WORLD JOURNAL OF GASTROENTEROLOGY, 2015, 21 (20) : 6374 - 6380
  • [26] Differential diagnosis of pancreatic cancer by single-shot echo-planar imaging diffusion-weighted imaging
    Ben-Zu Hong
    Xin-Feng Li
    Jian-Qing Lin
    World Journal of Gastroenterology, 2015, (20) : 6374 - 6380
  • [27] Histogram analysis of breast diffusion kurtosis imaging: a comparison between readout-segmented and single-shot echo-planar imaging sequence
    Huang, Cicheng
    Zhan, Chenao
    Hu, Yiqi
    Yin, Ting
    Grimm, Robert
    Ai, Tao
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2022,
  • [28] Advanced diffusion weighted imaging of the prostate: Comparison of readout-segmented multi-shot, parallel-transmit and single-shot echo-planar imaging
    Klingebiel, M.
    Ullrich, T.
    Quentin, M.
    Bonekamp, D.
    Aissa, J.
    Mally, D.
    Arsov, C.
    Albers, P.
    Antoch, G.
    Schimmoeller, L.
    EUROPEAN JOURNAL OF RADIOLOGY, 2020, 130
  • [29] Usefulness of free-breathing readout-segmented echo-planar imaging (RESOLVE) for detection of malignant liver tumors: Comparison with single-shot echo-planar imaging (SS-EPI)
    Tokoro, Hirokazu
    Fujinaga, Yasunari
    Ohya, Ayumi
    Ueda, Kazuhiko
    Shiobara, Aya
    Kitou, Yoshihiro
    Ueda, Hitoshi
    Kadoya, Masumi
    EUROPEAN JOURNAL OF RADIOLOGY, 2014, 83 (10) : 1728 - 1733
  • [30] Comparison between readout-segmented and single-shot echo-planar imaging in the evaluation of cervical cancer staging
    Qian, Weiliang
    Chen, Qian
    Zhang, Zhongshuai
    Wang, Hong
    Zhang, Jibin
    Xu, Jianming
    BRITISH JOURNAL OF RADIOLOGY, 2019, 92 (1094):