Performance evaluation of a whole-body prototype PET scanner with four-layer DOI detectors

被引:37
|
作者
Akamatsu, Go [1 ]
Tashima, Hideaki [1 ]
Iwao, Yuma [1 ]
Wakizaka, Hidekatsu [1 ]
Maeda, Takamasa [1 ]
Mohammadi, Akram [1 ]
Takyu, Sodai [1 ]
Nitta, Munetaka [1 ]
Nishikido, Fumihiko [1 ]
Rutherford, Harley [2 ,3 ]
Chacon, Andrew [2 ,3 ]
Safavi-Naeini, Mitra [3 ]
Yoshida, Eiji [1 ]
Yamaya, Taiga [1 ]
机构
[1] Natl Inst Quantum & Radiol Sci & Technol QST, NIRS, Chiba, Japan
[2] Univ Wollongong, Ctr Med Radiat Phys, Wollongong, NSW, Australia
[3] ANSTO, Sydney, NSW, Australia
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2019年 / 64卷 / 09期
关键词
PET; DOI; whole-body; NEMA NU 2; OF-FLIGHT PET/CT; AXIAL-FIELD; RESOLUTION; CRYSTALS; SYSTEM; DESIGN; IMPACT; VIEW; PMT;
D O I
10.1088/1361-6560/ab18b2
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Parallax error caused by the detector crystal thickness degrades spatial resolution at the peripheral regions of the field-of-view (FOV) of a scanner. To resolve this issue, depth-of-interaction (DOI) measurement is a promising solution to improve the spatial resolution and its uniformity over the entire FOV. Even though DOI detectors have been used in dedicated systems with a small ring diameter such as for the human brain, breast and small animals, the use of DOI detectors for a large bore whole-body PET system has not been demonstrated yet. We have developed a four-layered DOI detector, and its potential for a brain dedicated system has been proven in our previous development. In the present work, we investigated the use of the four-layer DOI detector for a large bore PET system by developing the world's first whole-body prototype. We evaluated its performance characteristics in accordance with the NEMA NU 2 standard. Furthermore, the impact of incorporating DOI information was evaluated with the NEMA NU 4 image quality phantom. Point source images were reconstructed with a filtered back projection (FBP), and an average spatial resolution of 5.2 +/- 0.7 mm was obtained. For the FBP image, the four-layer DOI information improved the radial spatial resolution by 48% at the 20 cm offset position. The peak noise-equivalent count rate (NECR) was 22.9 kcps at 7.4 kBq ml(-1) and the scatter fraction was 44%. The system sensitivity was 5.9 kcps MBq(-1). For the NEMA NU 2 image quality phantom, the 10 mm sphere was clearly visualized without any artifacts. For the NEMA NU 4 image quality phantom, we measured the phantom at 0, 10 and 20 cm offset positions. As a result, we found the image with four-layer DOI could visualize the 2 mm-diameter hot cylinder although it could not be recognized on the image without DOI. The average improvements in the recovery coefficients for the five hot rods (1-5 mm) were 0.3%, 4.4% and 26.3% at the 0, 10 and 20 cm offset positions, respectively (except for the 1 mm-diameter rod at the 20 cm offset position). Although several practical issues (such as adding end-shields) remain to be addressed before the scanner is ready for clinical use, we showed that the four-layer DOI technology provided higher and more uniform spatial resolution over the FOV and improved contrast for small uptake regions located at the peripheral FOV, which could improve detectability of small and distal lesions such as nodal metastases, especially in obese patients.
引用
收藏
页数:11
相关论文
共 50 条
  • [21] Performance of a whole-body PET scanner using curve-plate NaI(Tl) detectors
    Adam, LE
    Karp, JS
    Daube-Witherspoon, ME
    Smith, RJ
    JOURNAL OF NUCLEAR MEDICINE, 2001, 42 (12) : 1821 - 1830
  • [22] Spatial Resolution Measured by a Prototype System of Two Four-Layer DOI Detectors for jPET-RD
    Nishikido, Fumihiko
    Tsuda, Tomoaki
    Inadama, Naoko
    Yoshida, Eiji
    Takahashi, Kei
    Shibuya, Kengo
    Yamaya, Taiga
    Kitamura, Keishi
    Murayama, Hideo
    2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 3041 - 3044
  • [23] The jPET-D4: Performance evaluation of four-layer DOI-PET scanner using the NEMA NU2-2001 standard
    Yoshida, Eiji
    Kobayashi, Ayako
    Yamaya, Taiga
    Watanabe, Mitsuo
    Nishikido, Fumihiko
    Kitamura, Keishi
    Hasegawa, Tomoyuki
    Fukushi, Masahiro
    Murayama, Hideo
    2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 2532 - 2536
  • [24] Energy spectra and their calibration of four-layer DOI detector for brain PET scanner: jPET-D4
    Yoshida, E
    Kitamura, K
    Tsuda, T
    Inadama, N
    Murayama, H
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 4172 - 4176
  • [25] Energy spectra analysis of the four-layer DOI detector for the brain PET scanner: jPET-D4
    Yoshida, E
    Kitamura, K
    Tsuda, T
    Shibuya, K
    Yamaya, T
    Inadama, N
    Hasegawa, T
    Murayama, H
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 557 (02): : 664 - 669
  • [26] PERFORMANCE EVALUATION OF WHOLE-BODY NMR SCANNER ANTENNA SYSTEMS
    SERGIADIS, G
    MAGNETIC RESONANCE IN MEDICINE, 1985, 2 (04) : 328 - 335
  • [27] Four-layer DOI detector with a relative offset in animal PET system
    Ito, Mikiko
    Hong, Seong Jong
    Lee, Jae Sung
    Kwon, Sun Il
    Lee, Geon Song
    Park, Kwang Suk
    Hong, Byungsik
    Lee, Kyong Sei
    Lee, Seok Jae
    Rhee, June Tak
    Sim, Kwang-Souk
    2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 4296 - +
  • [28] Four-layer DOI-PET detector with a silicon photomultiplier array
    Nishikido, Fumihiko
    Inadama, Naoko
    Shibuya, Kengo
    Yoshida, Eiji
    Yamaya, Taiga
    Oda, Ichiro
    Kitamura, Keishi
    Murayama, Hideo
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 3198 - +
  • [29] Investigation of a whole-body DOI-PET system
    Ohi, Junichi
    Tonami, Hiromichi
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 571 (1-2): : 223 - 226
  • [30] A Dual-Layer TOF-DOI Detector Block for Whole-Body PET
    Liu, Shitao
    An, Shaohui
    Li, Hongdi
    Wang, Chao
    Ramirez, Rocio A.
    Zhang, Yuxuan
    Baghaei, Hossain
    Wong, Wai-Hoi
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2012, 59 (05) : 1805 - 1808