A Dual-Layer TOF-DOI Detector Block for Whole-Body PET

被引:6
|
作者
Liu, Shitao [1 ]
An, Shaohui [1 ]
Li, Hongdi [1 ]
Wang, Chao [1 ]
Ramirez, Rocio A. [1 ]
Zhang, Yuxuan [1 ]
Baghaei, Hossain [1 ]
Wong, Wai-Hoi [1 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Expt Diagnost Imaging, Houston, TX 77030 USA
基金
美国国家卫生研究院;
关键词
Depth of interaction (DOI); lutetium gadolinium oxyorthosilicate (LGSO); lutetium yttrium orthosilicate (LYSO); photomultiplier quadrant sharing (PQS); positron emission tomography (PET); time of flight (TOF); OF-FLIGHT PET; POSITRON-EMISSION-TOMOGRAPHY; TIME RESOLUTION; DEPTH; SCINTILLATION; PERFORMANCE; INFORMATION; MODULES; ARRAY;
D O I
10.1109/TNS.2012.2210053
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We developed a dual-layer high-resolution detector block for whole-body positron-emission tomography systems. The top layer is a 13 x 13 array of lutetium yttrium orthosilicate (LYSO) crystals (2.85 x 2.85 x 10 mm(3)) and the bottom layer a 13 x 13 array of lutetium gadolinium oxyorthosilicate (LGSO) crystals (2.85 x 2.85 x 10 mm(3)). The LGSO layer was coupled to four Hamamatsu R11194HA photomultiplier tubes (PMTs) by a photomultiplier quadrant-sharing configuration. This block has both time-of-flight and depth-of-interaction functions. Using a Na-22 point source, the performance of this block was evaluated by coinciding with a reference detector which consisted of a 4 x 4 x 10 mm(3) LYSO crystal coupled to a Hamamatsu R9779 PMT. A total of 4.7% of the counts were mixed among the two layers. The LYSO layer had 11.4% energy resolution and 329 ps coincidence time resolution against the reference detector. The LGSO layer had 14.8% energy resolution and 356 ps coincidence time resolution against the reference detector.
引用
收藏
页码:1805 / 1808
页数:4
相关论文
共 50 条
  • [1] A Dual-layer TOF-DOl Detector Block for Whole-body PET
    Liu, Shitao
    An, Shaohui
    Li, Hongdi
    Wang, Chao
    Ramirez, Rocio A.
    Zhang, Yuxuan
    Baghaei, Hossain
    Wong, Wai-Hoi
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 3008 - 3011
  • [2] A Dual-Ended Readout Detector Using a Meantime Method for SiPM TOF-DOI PET
    Kang, Han Gyu
    Ko, Guen Bae
    Rhee, June Tak
    Kim, Kyeong Min
    Lee, Jae Sung
    Hong, Seong Jong
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2015, 62 (05) : 1935 - 1943
  • [3] Investigation of a scalable phoswich crystal design for a TOF-DOI PET detector
    Chang, Chen-Ming
    Lee, Brian J.
    Sacco, Ilaria
    Levin, Craig S.
    2017 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2017,
  • [4] Novel staggered scintillator configurations for brain TOF-DOI PET detector
    Blackberg, Lisa
    Sajedi, Salar
    El Fakhri, Georges
    Sabet, Hamid
    JOURNAL OF NUCLEAR MEDICINE, 2020, 61
  • [5] Optimizing layer thickness for a TOF-DOI detector to be used in a long axial field-of-view PET scanner
    Schmall, Jeffrey
    Surti, Suleman
    Panetta, Joseph
    Karp, Joel
    JOURNAL OF NUCLEAR MEDICINE, 2015, 56 (03)
  • [6] A multi-resolution TOF-DOI detector for human brain dedicated PET scanner
    He, Wen
    Zhao, Yangyang
    Huang, Wenjie
    Zhao, Xin
    Niu, Ming
    Yang, Hang
    Zhang, Lei
    Ren, Qiushi
    Gu, Zheng
    PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (02):
  • [7] A dual layer DOI GSO block detector for a small animal PET
    Yamamoto, Seiichi
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2009, 598 (02): : 480 - 484
  • [8] Influence of detector pixel size, TOF resolution and DOI on image quality in MR-compatible whole-body PET
    Thoen, Hendrik
    Keereman, Vincent
    Mollet, Pieter
    Van Holen, Roel
    Vandenberghe, Stefaan
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (18): : 6459 - 6479
  • [9] Design of High-Resolution Brain PET Using Dual-Layer DOI Detector
    Lim, W.
    Lee, S.
    Jung, J. H.
    Choi, Y.
    INTERNAL MEDICINE JOURNAL, 2022, 52 : 33 - 34
  • [10] Crystal surface and reflector optimization for the SiPM-based dual-ended readout TOF-DOI PET detector
    Kang, Han Gyu
    Yamaya, Taiga
    Han, Young Been
    Song, Seong Hyun
    Ko, Guen Bae
    Lee, Jae Sung
    Hong, Seong Jong
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2020, 6 (06)