Stereotactic body radiation therapy patient specific quality assurance using a two-dimensional array at extended source to surface distance

被引:0
|
作者
Stathakis, Sotirios [1 ]
Mavroidis, Panayiotis [1 ]
Narayanasamy, Ganesh [1 ]
Markovic, Miljenko [1 ]
Myers, Pamela [2 ]
Papanikolaou, Niko [1 ]
机构
[1] Univ Texas Hlth Sci Ctr San Antonio, Dept Radiat Oncol, San Antonio, TX 78229 USA
[2] Baylor Coll Med, Dept Radiat Oncol, Houston, TX 77030 USA
来源
JOURNAL OF BUON | 2015年 / 20卷 / 04期
关键词
DQA; IMRT; measurements at extended SSD; SBRT; ION-CHAMBER ARRAY; IMRT DOSE VERIFICATION; SPATIAL-RESOLUTION; BEAMS; RADIOTHERAPY; PHOTON; DOSIMETRY; DETECTOR; SIZE; FILM;
D O I
暂无
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Methods: Five patients with 38 fields have been analyzed in this study. The plans were optimized for the following clinical sites: one. liver, one lung, one brain, one prostate aiid one spine. The detector array used for the measurements was the PTW Seven29 array. All the plans were optimized and calculated using Eclipse v8.9. The center of the array was setup at 215 cm from the source and all the fields were measured and analyzed one by one. All the 30 measurements were performed on a NovalisTX linear accelerator equipped with a high definition multileaf collimator. The evaluation was based mainly on gamma index passing rates using 2 mm distance to agreement (DTA) and 2% dose difference. Results: The accuracy of the Eclipse Treatment Planning System (TPS) at extended Source to Surface Distances (SSDs) using an ionization chamber was measured to be within 1.0%. All the field measurements were performed and analyzed 35. individually. The percent of the points that had a gamma index of less than 1 using 3%/3 mm was >99% for all the measurements. In order to better evaluate our process and distinguish smaller differences a new set of results was obtained by applying gamma index tolerances of 2%/2mm. In this case, the gamma index passing rates ranged from 90.8 to 100% (95.5%+/- 3%). The profile comparison showed that the detector array measurements followed closely the calculated 40 profiles, even for fields optimized with multiple peaks and valleys. Conclusion: The choice of the IMRT QA device has an important role in the results of the patient specific QA of the delivered dose to the patient in the case of small targets as in the treatment of spinal targets. In this study, we demonstrated that an extended SSD measurement can improve the sampling resolution of a two-dimensional (2D) detector array, in our case the PTW 45 Seven29 array. This method was shown to be accurate and efficient for measuring highly modulated small fields for pre-treatment patient specific QA.
引用
收藏
页码:1154 / 1163
页数:10
相关论文
共 50 条
  • [1] Evaluation of a two-dimensional diode array for patient-specific quality assurance of HyperArc
    Popple, Richard A.
    Sullivan, Rodney J.
    Yuan, Yuan
    Wu, Xingen
    Covington, Elizabeth L.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (12): : 203 - 210
  • [2] Patient-specific quality assurance using machine log files analysis for stereotactic body radiation therapy (SBRT)
    Chow, Vivian U. Y.
    Kan, Monica W. K.
    Chan, Anthony T. C.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (11): : 179 - 187
  • [3] Patient and Organ Specific Quality Assurance Phantom Insert for Stereotactic Body Radiation Therapy of Prostate Cancer
    Ross, Cullen
    Donlon, Elliott
    Kessler, Alexander
    Lee, Christopher
    Xiang, Hong
    Jaffe, C. Carl
    Bloch, B. Nicolas
    JOURNAL OF MEDICAL DEVICES-TRANSACTIONS OF THE ASME, 2015, 9 (02):
  • [4] Comparison of Diode-Based Array Devices with Varying Resolution for Patient-Specific Quality Assurance of Stereotactic Body Radiation Therapy
    Skinner, A. M.
    Tan, S.
    McDonald, D. G.
    Godwin, W. J.
    Rapchak, A. K.
    Peng, J. L.
    Roles, S. A.
    MEDICAL PHYSICS, 2024, 51 (09) : 6611 - 6612
  • [5] Quality assurance procedures for stereotactic body radiation therapy
    Calvin, James M.
    Bednarz, Greg
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 71 (01): : S122 - S125
  • [6] Quality Assurance for Stereotactic Body Radiation Therapy for Gynecologic Malignancies
    Andersen, Samuel N.
    Bonnen, Mark D.
    Ludwig, Michelle S.
    Dalwadi, Shraddha M.
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2024, 16 (02)
  • [7] A national survey on technology and quality assurance for stereotactic body radiation therapy
    Marino, Carmelo
    Garibaldi, Cristina
    Veronese, Ivan
    Carbonini, Claudia
    Russo, Serenella
    Stasi, Michele
    Mancosu, Pietro
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 65 : 6 - 14
  • [8] Dosimetric characterization of a new two-dimensional diode detector array used for stereotactic radiosurgery quality assurance
    Yasui, K.
    Saito, Y.
    Ogawa, S.
    Hayashi, N.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2021, 19 (02): : 281 - 289
  • [9] Use of a two-dimensional ionization chamber array for proton therapy quality assurance checks
    Arjomandy, B.
    Sahoo, N.
    Gillin, M.
    Zhu, X.
    MEDICAL PHYSICS, 2007, 34 (06) : 2502 - 2502
  • [10] Use of a two-dimensional ionization chamber array for proton therapy beam quality assurance
    Arjomandy, Bijan
    Sahoo, Narayan
    Ding, Xiaoning
    Gillin, Michael
    MEDICAL PHYSICS, 2008, 35 (09) : 3889 - 3894