Stereotactic body radiation therapy patient specific quality assurance using a two-dimensional array at extended source to surface distance

被引:0
|
作者
Stathakis, Sotirios [1 ]
Mavroidis, Panayiotis [1 ]
Narayanasamy, Ganesh [1 ]
Markovic, Miljenko [1 ]
Myers, Pamela [2 ]
Papanikolaou, Niko [1 ]
机构
[1] Univ Texas Hlth Sci Ctr San Antonio, Dept Radiat Oncol, San Antonio, TX 78229 USA
[2] Baylor Coll Med, Dept Radiat Oncol, Houston, TX 77030 USA
来源
JOURNAL OF BUON | 2015年 / 20卷 / 04期
关键词
DQA; IMRT; measurements at extended SSD; SBRT; ION-CHAMBER ARRAY; IMRT DOSE VERIFICATION; SPATIAL-RESOLUTION; BEAMS; RADIOTHERAPY; PHOTON; DOSIMETRY; DETECTOR; SIZE; FILM;
D O I
暂无
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Methods: Five patients with 38 fields have been analyzed in this study. The plans were optimized for the following clinical sites: one. liver, one lung, one brain, one prostate aiid one spine. The detector array used for the measurements was the PTW Seven29 array. All the plans were optimized and calculated using Eclipse v8.9. The center of the array was setup at 215 cm from the source and all the fields were measured and analyzed one by one. All the 30 measurements were performed on a NovalisTX linear accelerator equipped with a high definition multileaf collimator. The evaluation was based mainly on gamma index passing rates using 2 mm distance to agreement (DTA) and 2% dose difference. Results: The accuracy of the Eclipse Treatment Planning System (TPS) at extended Source to Surface Distances (SSDs) using an ionization chamber was measured to be within 1.0%. All the field measurements were performed and analyzed 35. individually. The percent of the points that had a gamma index of less than 1 using 3%/3 mm was >99% for all the measurements. In order to better evaluate our process and distinguish smaller differences a new set of results was obtained by applying gamma index tolerances of 2%/2mm. In this case, the gamma index passing rates ranged from 90.8 to 100% (95.5%+/- 3%). The profile comparison showed that the detector array measurements followed closely the calculated 40 profiles, even for fields optimized with multiple peaks and valleys. Conclusion: The choice of the IMRT QA device has an important role in the results of the patient specific QA of the delivered dose to the patient in the case of small targets as in the treatment of spinal targets. In this study, we demonstrated that an extended SSD measurement can improve the sampling resolution of a two-dimensional (2D) detector array, in our case the PTW 45 Seven29 array. This method was shown to be accurate and efficient for measuring highly modulated small fields for pre-treatment patient specific QA.
引用
收藏
页码:1154 / 1163
页数:10
相关论文
共 50 条
  • [21] Using a Knowledge-Based Planning Model for Quality Assurance of Liver Stereotactic Body Radiation Therapy Plans
    Gits, H. C.
    Dow, J.
    Feng, M.
    Matuszak, M. M.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 96 (02): : E645 - E646
  • [22] A two dimensional silicon detectors array for quality assurance in stereotactic radiotherapy: MagicPlate-512
    Aldosari, A. H.
    Petasecca, M.
    Espinoza, A.
    Newall, M.
    Fuduli, I.
    Porumb, C.
    Alshaikh, S.
    Alrowaili, Z. A.
    Weaver, M.
    Metcalfe, P.
    Carolan, M.
    Lerch, M. L. F.
    Perevertaylo, V.
    Rosenfeld, A. B.
    MEDICAL PHYSICS, 2014, 41 (09)
  • [23] Patient-specific Quality Assurance Using a 4-dimensional Programmable Motion Platform in Respiratory Gated Radiation Therapy
    Lee, J.
    Park, Y.
    Sung, W.
    Kim, J.
    Choi, C.
    Ye, S.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 84 (03): : S785 - S785
  • [24] Modeling of Gamma Index for Prediction of Pretreatment Quality Assurance in Stereotactic Body Radiation Therapy of the Liver
    Kamal, Rose
    Thaper, Deepak
    Singh, Gaganpreet
    Sharma, Shambhavi
    Navjeet, Arun Singh
    Oinam, Arun Singh
    Kumar, Vivek
    JOURNAL OF MEDICAL PHYSICS, 2024, 49 (02) : 232 - 239
  • [25] Verification of patient-specific dose distributions in proton therapy using a commercial two-dimensional ion chamber array
    Arjomandy, Bijan
    Sahoo, Narayan
    Ciangaru, George
    Zhu, Ronald
    Song, Xiaofei
    Gillin, Michael
    MEDICAL PHYSICS, 2010, 37 (11) : 5831 - 5837
  • [26] Patient-specific quality control for intensity-modulated radiation therapy and volumetric-modulated arc therapy using electronic portal imaging device and two-dimensional ion chamber array
    Kausar, Abu
    Mani, Karthick R.
    Azhari, Hasin A.
    Zakaria, Golam A.
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2019, 18 (01) : 26 - 31
  • [27] Film dosimetry studies for patient specific quality assurance in microbeam radiation therapy
    Ocadiz, Alexandre
    Livingstone, Jayde
    Donzelli, Mattia
    Bartzsch, Stefan
    Nemoz, Christian
    Kefs, Samy
    Pellicioli, Paolo
    Giraud, Jean-Yves
    Balosso, Jacques
    Krisch, Michael
    Brauer-Krisch, Elke
    Serduc, Raphael
    Adam, Jean-Francois
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 65 : 227 - 237
  • [28] Characterization of a Two-Dimensional Diode Detector Array for Use in Patient Specific, Pre-Treatment QA of Stereotactic Plans
    Wieczorek, D.
    Mishra, V.
    Poozhikala, T.
    Gutierrez, A.
    MEDICAL PHYSICS, 2019, 46 (06) : E321 - E321
  • [29] IMAT Patient Specific Quality Assurance Using ArcCHECK Diode Array Detector
    Thiyagarajan, Rajesh
    Sinha, S.
    Yadav, G.
    Ashokkumar, S.
    Raman, Kothanda
    Mishra, M.
    Nambiraj, N. Arunai
    MEDICAL PHYSICS, 2012, 39 (06) : 3733 - 3733
  • [30] Assessing the suitability of a two-dimensional array for routine quality assurance checks of flatness and symmetry
    Ibrahim, Salma
    Seshadri, Venkatakrishnan
    Charles, Paul
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2020, 43 (04) : 1451 - 1460