Patient-specific quality assurance using machine log files analysis for stereotactic body radiation therapy (SBRT)

被引:11
|
作者
Chow, Vivian U. Y. [1 ]
Kan, Monica W. K. [1 ,2 ]
Chan, Anthony T. C. [1 ,2 ]
机构
[1] Prince Wales Hosp, Dept Clin Oncol, Hong Kong, Peoples R China
[2] Chinese Univ Hong Kong, Dept Clin Oncol, Hong Kong, Peoples R China
来源
关键词
patient-specific QA; log file analysis; stereotactic body radiation therapy; volumetric-modulated arc therapy; plan delivery accuracy; TRAJECTORY LOG; DELIVERY; IMRT; ERRORS; IMPLEMENTATION;
D O I
10.1002/acm2.13053
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
An in-house trajectory log analysis program (LOGQA) was developed to evaluate the delivery accuracy of volumetric-modulated arc therapy (VMAT) for stereotactic body radiation therapy (SBRT). Methods have been established in LOGQA to provide analysis on dose indices, gantry angles, and multi-leaf collimator (MLC) positions. Between March 2019 and May 2020, 120 VMAT SBRT plans of various treatment sites using flattening filter-free (FFF) mode were evaluated using both LOGQA and phantom measurements. Gantry angles, dose indices, and MLC positions were extracted from log and compared with each plan. Integrated transient fluence map (ITFM) was reconstructed from log to examine the deviation of delivered fluence against the planned one. Average correlation coefficient of dose index versus gantry angle and ITFM for all patients were 1.0000, indicating that the delivered beam parameters were in good agreement with planned values. Maximum deviation of gantry angles and monitor units (MU) of all patients were less than 0.2 degree and 0.03 % respectively. Regarding MLC positions, maximum and root-mean-square (RMS) deviations from planned values were less than 0.6 mm and 0.3 mm respectively, indicating that MLC positions during delivery followed planned values in precise manner. Results of LOGQA were consistent with measurement, where all gamma-index passing rates were larger than 95 %, with 2 %/2 mm criteria. Three types of intentional errors were introduced to patient plan for software validation. LOGQA was found to recognize the introduced errors of MLC positions, gantry angles, and dose indices with magnitudes of 1 mm, 1 degree, and 5 %, respectively, which were masked in phantom measurement. LOGQA was demonstrated to have the potential to reduce or even replace patient-specific QA measurements for SBRT plan delivery provided that the frequency and amount of measurement-based machine-specific QA can be increased to ensure the log files record real values of machine parameters.
引用
收藏
页码:179 / 187
页数:9
相关论文
共 50 条
  • [1] Inverse Verification of the Dose Distribution for Intensity Modulated Radiation Therapy Patient-specific Quality Assurance Using Dynamic MLC Log Files
    Lee, Jeong-Woo
    Park, Jeong-Hoon
    Chung, Jin-Beom
    Park, Ji-Yeon
    Choe, Bo-Young
    Suh, Tae-Suk
    Lee, Doo-Hyun
    Hong, Semie
    Kang, Min-Young
    Choi, Kyoung-Sik
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2009, 55 (04) : 1649 - 1656
  • [2] Use of treatment log files in spot scanning proton therapy as part of patient-specific quality assurance
    Li, Heng
    Sahoo, Narayan
    Poenisch, Falk
    Suzuki, Kazumichi
    Li, Yupeng
    Li, Xiaoqiang
    Zhang, Xiaodong
    Lee, Andrew K.
    Gillin, Michael T.
    Zhu, X. Ronald
    MEDICAL PHYSICS, 2013, 40 (02)
  • [3] Patient-Specific Rectal Toxicity Predictor Based Plan Quality Control for Prostate Stereotactic Body Radiation Therapy (SBRT)
    Song, T.
    Li, Y.
    Zhou, L.
    Jiang, S.
    Gu, X.
    MEDICAL PHYSICS, 2015, 42 (06) : 3211 - 3211
  • [4] Patient Specific Quality Assurance Using Machine Log Files for the Mevion S250i Proton Therapy System
    Bunker, A.
    Ahmad, S.
    Chen, Y.
    MEDICAL PHYSICS, 2019, 46 (06) : E325 - E325
  • [5] Analysis of dose comparison techniques for patient-specific quality assurance in radiation therapy
    Yu, Liting
    Tang, Timothy L. S.
    Cassim, Naasiha
    Livingstone, Alexander
    Cassidy, Darren
    Kairn, Tanya
    Crowe, Scott B.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (11): : 189 - 198
  • [6] Comparison of Diode-Based Array Devices with Varying Resolution for Patient-Specific Quality Assurance of Stereotactic Body Radiation Therapy
    Skinner, A. M.
    Tan, S.
    McDonald, D. G.
    Godwin, W. J.
    Rapchak, A. K.
    Peng, J. L.
    Roles, S. A.
    MEDICAL PHYSICS, 2024, 51 (09) : 6611 - 6612
  • [7] Patient-Specific Quality Assurance Using Monte Carlo Dose Calculation and Elekta Log Files for Prostate Volumetric-Modulated Arc Therapy
    Katsuta, Yoshiyuki
    Kadoya, Noriyuki
    Fujita, Yukio
    Shimizu, Eiji
    Matsunaga, Kenichi
    Sawada, Kinya
    Matsushita, Haruo
    Majima, Kazuhiro
    Jingu, Keiichi
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2017, 16 (06) : 1220 - 1225
  • [8] Efficient quality assurance for isocentric stability in stereotactic body radiation therapy using machine learning
    Sana Salahuddin
    Saeed Ahmad Buzdar
    Khalid Iqbal
    Muhammad Adeel Azam
    Lidia Strigari
    Radiological Physics and Technology, 2024, 17 : 219 - 229
  • [9] Efficient quality assurance for isocentric stability in stereotactic body radiation therapy using machine learning
    Salahuddin, Sana
    Buzdar, Saeed Ahmad
    Iqbal, Khalid
    Azam, Muhammad Adeel
    Strigari, Lidia
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2024, 17 (01) : 219 - 229
  • [10] An investigation of using log-file analysis for automated patient-specific quality assurance in MRgRT
    Lim, Seng Boh
    Godoy Scripes, Paola
    Napolitano, Mary
    Subashi, Ergys
    Tyagi, Neelam
    Cervino Arriba, Laura
    Lovelock, Dale Michael
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (09): : 183 - 188