An investigation of using log-file analysis for automated patient-specific quality assurance in MRgRT

被引:8
|
作者
Lim, Seng Boh [1 ]
Godoy Scripes, Paola [1 ]
Napolitano, Mary [2 ]
Subashi, Ergys [1 ]
Tyagi, Neelam [1 ]
Cervino Arriba, Laura [1 ]
Lovelock, Dale Michael [1 ]
机构
[1] Mem Sloan Kettering Canc Ctr, 1275 York Ave, New York, NY 10021 USA
[2] Standard Imaging Inc, Middleton, WI USA
来源
关键词
elekta unity; linacview; log-file analysis; MRI guided adaptive radiotherapy; MRgRT; patient specific quality assurance; MLC PERFORMANCE; IMRT; QA;
D O I
10.1002/acm2.13361
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective Adaptive radiation therapy (ART) is an integral part of MR-guided RT (MRgRT), requiring a new RT plan for each treatment fraction and resulting in a significant increase in patient-specific quality assurance (PSQA). This study investigates the possibility of using treatment log-file for automated PSQA. Method All treatment plans were delivered in 1.5T Unity MR-Linac (Elekta). A Unity compatible version of LinacView (Standard Imaging) was commissioned to automatically monitor and analyze the log-files. A total of 220 fields were delivered and measured by ArcCheck-MR (Sun Nuclear) and LinacView. Thirty incorrectly matched fields were also delivered to check for error detection sensitivity. The gamma analysis, gamma, with 3%, 3 mm criteria was used in both ArcCheck-MR and LinacView. Additionally, the gantry angle, jaws, and multileaf collimators (MLC) positions reported in the log-file were compared with plan positions using TG-142 criteria. Result The gamma (3%, 3 mm) for the 190 plans were found to be between the range of 72.5%-100.0% and 95.4%-100.0% for ArcCheck-MR and LinacVeiw, respectively. All the delivered gantry angle and jaws were found to be within 0.2 degrees and 2 mm. MLCs that were outside the guard leaves or under the diaphragms were found to have more than 1.0 mm discrepancy. This was attributed to the linac internal override for these MLCs and had no dosimetric impact. Excluding these discrepancies, all MLC positions were found to be within 1.0 mm. The gamma (3%, 3 mm) for the 30 incorrectly matched fields were found to be 3.9%-84.8% and 0.1%-64.4% for ArcCheck-MR and LinacVeiw, respectively. Conclusion Significant ranked correlation demonstrates the automated log-file analysis can be used for PSQA and expedite the ART workflow. Ongoing PSQA will be compared with log-file analysis to investigate the longer term reproducibility and correlation.
引用
收藏
页码:183 / 188
页数:6
相关论文
共 50 条
  • [1] An Investigation of Using Logfile Analysis for Automated Patient Specific Quality Assurance in MRgRT
    Lim, S.
    Scripes, P. G.
    Napolitano, M.
    Subashi, E.
    Tyagi, N.
    Cervino, L.
    Lovelock, D.
    MEDICAL PHYSICS, 2020, 47 (06) : E450 - E450
  • [2] Comparing log file to measurement-based patient-specific quality assurance
    Li Ting Chan
    Yun Inn Tan
    Poh Wee Tan
    Yuh Fun Leong
    Jong Shin Khor
    Mun Woan Teh
    Joan Faith Loria Cruz
    Sháun Peter Baggarley
    Kiat Huat Ooi
    Yiat Horng Leong
    Physical and Engineering Sciences in Medicine, 2023, 46 : 303 - 311
  • [3] Comparing log file to measurement-based patient-specific quality assurance
    Ting, Chan Li
    Inn, Tan Yun
    Wee, Tan Poh
    Leong, Leong Yuh
    Shin, Khor Jong
    Woan, Teh Mun
    Cruz, Joan Faith Loria
    Baggarley, Shaun Peter
    Huat, Ooi Kiat
    Leong, Leong Yiat
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2023, 46 (01) : 303 - 311
  • [4] Patient-Specific Quality Assurance Using Trajectory Log File-Based Dose Reconstruction for SRS VMAT
    Park, H.
    Park, J.
    Artz, M.
    Huh, S.
    MEDICAL PHYSICS, 2020, 47 (06) : E826 - E826
  • [5] MLCs Performance and Patient-Specific IMRT QA Using Log File Analysis
    Osman, A.
    Maalej, N.
    Jayesh, K.
    Abdel-Rahman, W.
    MEDICAL PHYSICS, 2016, 43 (06) : 3530 - 3530
  • [6] Deep Learning for Patient-Specific Quality Assurance : Predicting Gamma Passing Rates for IMRT Based Log File
    Huang, Y.
    Huang, Y.
    Xu, Z.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [7] Automated patient-specific IMRT quality assurance using exit detector data
    Seibert, R.
    Ramsey, C.
    MEDICAL PHYSICS, 2006, 33 (06) : 2168 - 2168
  • [8] Challenges and Insights from a Year of Delivery Log-File Analysis for Patient-Specific QA in a Multi-Vendor Linac Enterprise
    Andersen, G.
    Jermain, P.
    Crawford, B. T.
    Foote, C. M.
    McClatchy, D., III
    MEDICAL PHYSICS, 2024, 51 (10) : 7813 - 7813
  • [9] Comparison of Log-File Analysis and Verification Measurement for Quality Assurance of Intensity-modulated Radiation Plans
    Kowatsch, M.
    Szeverinski, P.
    Meinschad, M.
    Kuenzler, T.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2018, 194 (11) : 1085 - 1085
  • [10] An Integrated Log-File Based Quality Assurance Platform for Proton Flash Radiation Therapy
    Tsai, P.
    Pfeiler, T.
    Krieger, M.
    Spiessens, S.
    Folkerts, M. M.
    Simone, C. B., II
    Lin, H.
    MEDICAL PHYSICS, 2024, 51 (10) : 7807 - 7807