Fracture of the GORE HELEX Septal Occluder: Associated Factors and Clinical Outcomes

被引:27
|
作者
Fagan, Thomas [1 ]
Dreher, Dennis [2 ]
Cutright, Warren [2 ]
Jacobson, Joth [2 ]
Latson, Larry [3 ]
机构
[1] Univ Colorado, Childrens Hosp, Div Pediat Cardiol, Aurora, CO USA
[2] WL Gore & Associates, Flagstaff, AZ USA
[3] Cleveland Clin Fdn, Div Pediat Cardiol, Cleveland, OH 44195 USA
关键词
congenital heart disease; atrial septal defect; septal occluder device; frame fracture; catheterization; PATENT FORAMEN OVALE; TRANSCATHETER CLOSURE; PERCUTANEOUS CLOSURE; CATHETER ABLATION; CARDIAC CYCLE; ATRIAL; DEVICE; DEFECT; OCCLUSION; COMPLICATIONS;
D O I
10.1002/ccd.21929
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives: This study examines factors associated with, and clinical effects of, wire frame fractures (WFF) of the GORE HELEX (TM) Septal Occluder (HELEX). Methods: Investigator-reported data from every HELEX implanted in the United States between 4/2000 and 4/2005 were reviewed. Clinical and procedural data from patients who experienced WFF were compared with data from patients who did not. The echocardiographic and fluoroscopic images for HELEX with WFF were reviewed for predictors of WFF and alterations in HELEX function. Results: With 90% of subjects followed for >12 months, 19/298 (6.4%) HELEX implanted were found to have WFF. Thirty WFFs were observed, with multiple WFFs occurring in 8/19 HELEX. Univariate predictors of WFF were large device size (P = 0.0003) and balloon defect size (P = 0.001); however, large device size (P = 0.0003) was the only significant predictor of WFF by multivariate analysis. WFF of the 30-mm and 35-mm HELEX accounted for 84% (16/19) of all WFFs. Review of HELEX images with WFF revealed all WFF occurred along the circumferential wire, except one (straight portion of locking loop). Seventeen of 29 (59%) circumferential WFFs were on the right disk. Residual defect status remained clinically insignificant (6/19) or became completely occluded (13/19) during follow-up. There were no clinical sequelae due to WFF; however, secondary to right atrial disk mobility, the HELEX with locking loop WFF was percutaneously removed 6 weeks after implantation. Conclusions: WFF occurred in 6.4% of HELEX and were most common in large devices. With the exception of one device removed for theoretical risks, no clinical sequelae were related to WFF. WFF did not alter the function of the HELEX. (C) 2009 Wiley-Liss, Inc.
引用
收藏
页码:941 / 948
页数:8
相关论文
共 50 条
  • [41] Late thrombus formation on the Helex septal occluder after double-lung transplant
    Zaidi, Ali N.
    Cheatham, John P.
    Galantowicz, Mark
    Astor, Todd
    Kovalchin, John P.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2010, 29 (07): : 814 - 816
  • [42] Biatrial Thrombi on a Gore Septal Occluder Despite Early Epithelialization
    Herron, Christopher
    Iacono, Karen
    Nykanen, David
    PEDIATRIC CARDIOLOGY, 2024,
  • [43] Midterm Follow-Up of Percutaneous Closure of Secundum Atrial Septal Defect with Helex Septal Occluder
    Smith, Benjamin G.
    Wilson, Neil
    Richens, Trevor
    Knight, W. Brodie
    JOURNAL OF INTERVENTIONAL CARDIOLOGY, 2008, 21 (04) : 363 - 368
  • [44] Initial Clinical Experience With the GORE® CARDIOFORM ASD Occluder for Transcatheter Atrial Septal Defect Closure
    de Hemptinne, Quentin
    Horlick, Eric M.
    Osten, Mark D.
    Millan, Xavier
    Tadros, Victor-Xavier
    Pighi, Michele
    Barlatey, Francisco Gonzalez
    Alnasser, Sami M.
    Miro, Joaquim
    Asgar, Anita W.
    Ibrahim, Reda
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2017, 90 (03) : 495 - 503
  • [45] Transcatheter PFO Closure With GORE® Septal Occluder: Early and Mid-Term Clinical Results
    Butera, Gianfranco
    Saracino, Antonio
    Danna, Paolo
    Sganzerla, Paolo
    Chessa, Massimo
    Carminati, Mario
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2013, 82 (06) : 944 - 949
  • [46] Upsizing of GORE® Cardioform ASD Occluder for Atrial Septal Defect With Atrial Septal Aneurysm
    Hayama, Hiromasa
    Makino, Kenji
    Yazaki, Yoshiyuki
    Hara, Hidehiko
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2024, 16 (06)
  • [47] First experience with the new GORE® septal occluder for the closure of multiple atrial septal defects
    Freixa, Xavier
    Garceau, Patrick
    Asgar, Anita W.
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2013, 81 (07) : 1238 - 1242
  • [48] Recurrent Strokes After Gore Septal Occluder Device Closure of Atrial Septal Defect
    Alhadlaq, Adnan
    Hussain, Arif
    Friesen, Camille Hancock
    Dhillon, Santokh
    CANADIAN JOURNAL OF CARDIOLOGY, 2017, 33 (12) : 1736.e9 - 1736.e11
  • [49] HELEX septal occluder, a low profile and effective device for transcatheter closure of fontan fenestration
    Coc, JY
    Pepelassis, D
    Rutledge, J
    PEDIATRIC RESEARCH, 2004, 55 (04) : 51A - 51A
  • [50] Results of the US multicenter pivotal study of the HELEX septal occluder for percutaneous closure of secundum atrial septal defects
    Jones, Thomas K.
    Latson, Larry A.
    Zahn, Evan
    Fleishman, Craig E.
    Jacobson, Joth
    Vincent, Robert
    Kanter, Kirk
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2007, 49 (22) : 2215 - 2221