Fracture of the GORE HELEX Septal Occluder: Associated Factors and Clinical Outcomes

被引:27
|
作者
Fagan, Thomas [1 ]
Dreher, Dennis [2 ]
Cutright, Warren [2 ]
Jacobson, Joth [2 ]
Latson, Larry [3 ]
机构
[1] Univ Colorado, Childrens Hosp, Div Pediat Cardiol, Aurora, CO USA
[2] WL Gore & Associates, Flagstaff, AZ USA
[3] Cleveland Clin Fdn, Div Pediat Cardiol, Cleveland, OH 44195 USA
关键词
congenital heart disease; atrial septal defect; septal occluder device; frame fracture; catheterization; PATENT FORAMEN OVALE; TRANSCATHETER CLOSURE; PERCUTANEOUS CLOSURE; CATHETER ABLATION; CARDIAC CYCLE; ATRIAL; DEVICE; DEFECT; OCCLUSION; COMPLICATIONS;
D O I
10.1002/ccd.21929
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives: This study examines factors associated with, and clinical effects of, wire frame fractures (WFF) of the GORE HELEX (TM) Septal Occluder (HELEX). Methods: Investigator-reported data from every HELEX implanted in the United States between 4/2000 and 4/2005 were reviewed. Clinical and procedural data from patients who experienced WFF were compared with data from patients who did not. The echocardiographic and fluoroscopic images for HELEX with WFF were reviewed for predictors of WFF and alterations in HELEX function. Results: With 90% of subjects followed for >12 months, 19/298 (6.4%) HELEX implanted were found to have WFF. Thirty WFFs were observed, with multiple WFFs occurring in 8/19 HELEX. Univariate predictors of WFF were large device size (P = 0.0003) and balloon defect size (P = 0.001); however, large device size (P = 0.0003) was the only significant predictor of WFF by multivariate analysis. WFF of the 30-mm and 35-mm HELEX accounted for 84% (16/19) of all WFFs. Review of HELEX images with WFF revealed all WFF occurred along the circumferential wire, except one (straight portion of locking loop). Seventeen of 29 (59%) circumferential WFFs were on the right disk. Residual defect status remained clinically insignificant (6/19) or became completely occluded (13/19) during follow-up. There were no clinical sequelae due to WFF; however, secondary to right atrial disk mobility, the HELEX with locking loop WFF was percutaneously removed 6 weeks after implantation. Conclusions: WFF occurred in 6.4% of HELEX and were most common in large devices. With the exception of one device removed for theoretical risks, no clinical sequelae were related to WFF. WFF did not alter the function of the HELEX. (C) 2009 Wiley-Liss, Inc.
引用
收藏
页码:941 / 948
页数:8
相关论文
共 50 条
  • [1] Fracture of a GORE HELEX Septal Occluder Following PFO Closure in a Diver
    Scott, Paul
    Wilson, Neil
    Veldtman, Gruchen
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2009, 73 (06) : 828 - 831
  • [3] Combined prospective United States clinical study data for the GORE® HELEX® septal occluder device
    Rhodes, John F., Jr.
    Goble, Jake
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2014, 83 (06) : 944 - 952
  • [4] Gore Helex Septal Occluder for Percutaneous Closure of Patent Foramen Ovale Associated With Atrial Septal Aneurysm: Short- and Mid-Term Clinical and Echocardiographic Outcomes
    Musto, Carmine
    Cifarelli, Alberta
    Fiorilli, Rosario
    De Felice, Francesco
    Parma, Antonio
    Pandolfi, Claudia
    Confessore, Pierpaolo
    Bernardi, Leda
    Violini, Roberto
    JOURNAL OF INVASIVE CARDIOLOGY, 2012, 24 (10): : 510 - 514
  • [5] Single Centre Experience with GORE-HELEX Septal Occluder for Closure of PFO
    Ponnuthurai, Francis A.
    van Gaal, William J.
    Burchell, Amy
    Mitchell, Andrew
    Wilson, Neil
    Ormerod, Oliver
    HEART LUNG AND CIRCULATION, 2009, 18 (02): : 140 - 142
  • [6] Failure of the GORE HELEX Atrial Septal Defect Occluder to Treat Membranous VSDs in Swine
    Hagler, Donald J.
    Latson, Larry A.
    Goble, Jake
    Mathewson, James W.
    Cutright, Warren
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2010, 76 (06) : 874 - 877
  • [7] Fenestration of a Gore Helex Septal Occluder Device in a Patient with Diastolic Dysfunction of the Left Ventricle
    Kenny, Damien
    Cao, Qi-Ling
    Hijazi, Ziyad M.
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2011, 78 (04) : 594 - 598
  • [8] Closure of a secundum atrial septal defect in two infants with chronic lung disease using the Gore HELEX Septal Occluder
    Zussman, Matt E.
    Freire, Grace
    Cupp, Shawn D.
    Stapleton, Gary E.
    CARDIOLOGY IN THE YOUNG, 2016, 26 (01) : 79 - 83
  • [9] Percutaneous Retrieval of the Locked Helex Septal Occluder
    Poommipanit, Paul
    Levi, Daniel
    Shenoda, Michael
    Tobis, Jonathan
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2011, 77 (06) : 892 - 900
  • [10] US multicenter pivotal study of the HELEX Septal Occluder