Fracture of the GORE HELEX Septal Occluder: Associated Factors and Clinical Outcomes

被引:27
|
作者
Fagan, Thomas [1 ]
Dreher, Dennis [2 ]
Cutright, Warren [2 ]
Jacobson, Joth [2 ]
Latson, Larry [3 ]
机构
[1] Univ Colorado, Childrens Hosp, Div Pediat Cardiol, Aurora, CO USA
[2] WL Gore & Associates, Flagstaff, AZ USA
[3] Cleveland Clin Fdn, Div Pediat Cardiol, Cleveland, OH 44195 USA
关键词
congenital heart disease; atrial septal defect; septal occluder device; frame fracture; catheterization; PATENT FORAMEN OVALE; TRANSCATHETER CLOSURE; PERCUTANEOUS CLOSURE; CATHETER ABLATION; CARDIAC CYCLE; ATRIAL; DEVICE; DEFECT; OCCLUSION; COMPLICATIONS;
D O I
10.1002/ccd.21929
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives: This study examines factors associated with, and clinical effects of, wire frame fractures (WFF) of the GORE HELEX (TM) Septal Occluder (HELEX). Methods: Investigator-reported data from every HELEX implanted in the United States between 4/2000 and 4/2005 were reviewed. Clinical and procedural data from patients who experienced WFF were compared with data from patients who did not. The echocardiographic and fluoroscopic images for HELEX with WFF were reviewed for predictors of WFF and alterations in HELEX function. Results: With 90% of subjects followed for >12 months, 19/298 (6.4%) HELEX implanted were found to have WFF. Thirty WFFs were observed, with multiple WFFs occurring in 8/19 HELEX. Univariate predictors of WFF were large device size (P = 0.0003) and balloon defect size (P = 0.001); however, large device size (P = 0.0003) was the only significant predictor of WFF by multivariate analysis. WFF of the 30-mm and 35-mm HELEX accounted for 84% (16/19) of all WFFs. Review of HELEX images with WFF revealed all WFF occurred along the circumferential wire, except one (straight portion of locking loop). Seventeen of 29 (59%) circumferential WFFs were on the right disk. Residual defect status remained clinically insignificant (6/19) or became completely occluded (13/19) during follow-up. There were no clinical sequelae due to WFF; however, secondary to right atrial disk mobility, the HELEX with locking loop WFF was percutaneously removed 6 weeks after implantation. Conclusions: WFF occurred in 6.4% of HELEX and were most common in large devices. With the exception of one device removed for theoretical risks, no clinical sequelae were related to WFF. WFF did not alter the function of the HELEX. (C) 2009 Wiley-Liss, Inc.
引用
收藏
页码:941 / 948
页数:8
相关论文
共 50 条
  • [31] Initial clinical experience with the GORE septal occluder for the treatment of atrial septal defects and patent foramen ovale
    Freixa, Xavier
    Ibrahim, Reda
    Chan, Jason
    Garceau, Patrick
    Dore, Annie
    Marcotte, Francois
    Mercier, Lise-Andree
    Mongeon, Francois-Pierre
    Basmadjian, Arsene
    Khairy, Paul
    Asgar, Anita W.
    EUROINTERVENTION, 2013, 9 (05) : 629 - 635
  • [32] Experience using the new GORE® septal occluder at the margins
    Lockhart, C. J.
    Johnston, N. G.
    Spence, M. S.
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2013, 81 (07) : 1244 - 1248
  • [33] Percutaneous removal of unraveled HELEX® septal occluder 4 months post deployment
    Suntharos, Patcharapong
    Komarlu, Rukmini
    Prieto, Lourdes R.
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2015, 86 (01) : 115 - 120
  • [34] Perforating the GORE® CARDIOFORM septal occluder and atrial septal defect occluder to gain access to the left atrium
    Anderson, Jason H.
    Adamson, Teagan
    Migliati, Elton
    Herlihy, Jennipher
    Danieu, Pierrette
    Daly, John
    Allen, Carissa
    Hua, Khoa
    Inglessis, Ignacio
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2020, 96 (06) : E660 - E665
  • [35] Sequential implantation of two Helex septal occluder devices in a patient with complex atrial septal anatomy
    Dobrolet, NC
    Iskowitz, S
    Lopez, L
    Whalen, R
    Zahn, EM
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2001, 54 (02) : 242 - 246
  • [36] Transcatheter closure of atrial septal defects and patent foramen ovale using the Helex septal occluder
    Ostermayer, SHU
    Wilson, N
    Billinger, K
    Reschke, M
    Bayard, Y
    Buescheck, F
    Fischer, E
    Leetz, M
    Taaffe, M
    Hein, R
    Roemer, A
    Sievert, H
    AMERICAN JOURNAL OF CARDIOLOGY, 2005, 96 (7A): : 91H - 91H
  • [37] AN UNPRECEDENTED CASE OF LATE BACTERIAL ENDOCARDITIS INVOLVING THE HELEX SEPTAL OCCLUDER DEVICE
    Thiagaraj, Ashwin
    Bloomingdale, Richard
    Herrmann, Kimberly
    Cardozo, Shaun
    Afonso, Luis
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2016, 67 (13) : 1126 - 1126
  • [38] First Experiences With the GORE® Septal Occluder in Children and Adults With Atrial Septal Defects
    Nyboe, Camilla
    Hjortdal, Vibeke E.
    Nielsen-Kudsk, Jens Erik
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2013, 82 (06) : 929 - 934
  • [39] OUTCOMES OF SECUNDUM ATRIAL SEPTAL DEFECT CLOSURE WITH THE NEW GORE CARDIOFORM ASD OCCLUDER- RESULTS FROM THE CONTINUED ACCESS GORE ASSURED CLINICAL TRIAL
    Qureshi, Athar M.
    Sommer, Robert
    Morgan, Gareth
    Gray, Robert
    Love, Barry
    Goldstein, Bryan
    Sugeng, Lissa
    Gillespie, Matthew
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2021, 77 (18) : 436 - 436
  • [40] Development and testing of the helex septal occluder, a new expanded polytetrafluoroethylene atrial septal defect occlusion system
    Zahn, EM
    Wilson, N
    Cutright, W
    Latson, LA
    CIRCULATION, 2001, 104 (06) : 711 - 716