Fracture of the GORE HELEX Septal Occluder: Associated Factors and Clinical Outcomes

被引:27
|
作者
Fagan, Thomas [1 ]
Dreher, Dennis [2 ]
Cutright, Warren [2 ]
Jacobson, Joth [2 ]
Latson, Larry [3 ]
机构
[1] Univ Colorado, Childrens Hosp, Div Pediat Cardiol, Aurora, CO USA
[2] WL Gore & Associates, Flagstaff, AZ USA
[3] Cleveland Clin Fdn, Div Pediat Cardiol, Cleveland, OH 44195 USA
关键词
congenital heart disease; atrial septal defect; septal occluder device; frame fracture; catheterization; PATENT FORAMEN OVALE; TRANSCATHETER CLOSURE; PERCUTANEOUS CLOSURE; CATHETER ABLATION; CARDIAC CYCLE; ATRIAL; DEVICE; DEFECT; OCCLUSION; COMPLICATIONS;
D O I
10.1002/ccd.21929
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives: This study examines factors associated with, and clinical effects of, wire frame fractures (WFF) of the GORE HELEX (TM) Septal Occluder (HELEX). Methods: Investigator-reported data from every HELEX implanted in the United States between 4/2000 and 4/2005 were reviewed. Clinical and procedural data from patients who experienced WFF were compared with data from patients who did not. The echocardiographic and fluoroscopic images for HELEX with WFF were reviewed for predictors of WFF and alterations in HELEX function. Results: With 90% of subjects followed for >12 months, 19/298 (6.4%) HELEX implanted were found to have WFF. Thirty WFFs were observed, with multiple WFFs occurring in 8/19 HELEX. Univariate predictors of WFF were large device size (P = 0.0003) and balloon defect size (P = 0.001); however, large device size (P = 0.0003) was the only significant predictor of WFF by multivariate analysis. WFF of the 30-mm and 35-mm HELEX accounted for 84% (16/19) of all WFFs. Review of HELEX images with WFF revealed all WFF occurred along the circumferential wire, except one (straight portion of locking loop). Seventeen of 29 (59%) circumferential WFFs were on the right disk. Residual defect status remained clinically insignificant (6/19) or became completely occluded (13/19) during follow-up. There were no clinical sequelae due to WFF; however, secondary to right atrial disk mobility, the HELEX with locking loop WFF was percutaneously removed 6 weeks after implantation. Conclusions: WFF occurred in 6.4% of HELEX and were most common in large devices. With the exception of one device removed for theoretical risks, no clinical sequelae were related to WFF. WFF did not alter the function of the HELEX. (C) 2009 Wiley-Liss, Inc.
引用
收藏
页码:941 / 948
页数:8
相关论文
共 50 条
  • [21] A Comparison Between Gore Cardioform and Amplatzer Septal Occluder for Percutaneous Closure of Patent Foramen Ovale Associated With Atrial Septal Aneurysm: Clinical and Echocardiographic Outcomes
    Musto, Carmine
    Cifarelli, Alberta
    Dipasquale, Francesco
    Chin, Diana
    Nazzaro, Marco Stefano
    Stio, Rocco Edoardo
    Pennacchi, Mauro
    De Felice, Francesco
    JOURNAL OF INVASIVE CARDIOLOGY, 2021, 33 (11): : E857 - E862
  • [22] Analysis of factors related to successful transcatheter closure of secundum atrial septal defects using the HELEX septal occluder
    Latson, LA
    Jones, TK
    Jacobson, J
    Zahn, E
    Rhodes, JF
    AMERICAN HEART JOURNAL, 2006, 151 (05) : 1129.e7 - 1129.e11
  • [23] The Role of Echocardiography in the Management of an Embolized Atrial Septal Helex® Occluder
    Almeida-Chen, Gracie
    Niezgoda, Julie
    Diaz-Gomez, Jose L.
    Motta, Pablo
    ANESTHESIA AND ANALGESIA, 2010, 110 (03): : 734 - 736
  • [24] Transcranial Doppler Quantification of Residual Shunt After Percutaneous Patent Foramen Ovale Closure: Efficacy of the GORE® HELEX Septal Occluder
    Sorensen, Sherman G.
    Smout, Randall
    Spruance, Spotswood L.
    JOURNAL OF INTERVENTIONAL CARDIOLOGY, 2011, 24 (04) : 366 - 372
  • [25] Echocardiographic considerations during deployment of the Helex Septal Occluder for closure of atrial septal defects
    Lopez, L
    Ventura, R
    Welch, EM
    Nykanen, DG
    Zahn, EM
    CARDIOLOGY IN THE YOUNG, 2003, 13 (03) : 290 - 298
  • [26] Two cases of pericardial tamponade due to nitinol wire fracture of a gore septal occluder
    Kumar, Preetham
    Orford, James L.
    Tobis, Jonathan M.
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2020, 96 (01) : 219 - 224
  • [27] Atrial septal defect closure with the HELEX septal occluder: US pivotal study results
    Jones, TK
    Zahn, EM
    Vincent, RN
    Latson, LA
    CIRCULATION, 2002, 106 (19) : 485 - 485
  • [28] Percutaneous closure of patent foramen ovale using the HELEX septal occluder
    Reschke, M
    Ostermayer, S
    Büscheck, F
    Bayard, Y
    Billinger, K
    Trepels, T
    Sievert, H
    AMERICAN JOURNAL OF CARDIOLOGY, 2003, 92 (6A): : 13L - 13L
  • [29] Displacement of helex atrial septal occluder: a rare complication of percutaneous occlusion of atrial septal defect
    Unal, Yusuf
    Kula, Serdar
    Emmez, Gokcen
    Olgunturk, Rana
    Yardim, Sahin
    PEDIATRIC ANESTHESIA, 2008, 18 (06) : 568 - 569
  • [30] Secondary embolization of a Helex occluder implanted into a secundum atrial septal defect
    Peuster, M
    Reckers, J
    Fink, C
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2003, 59 (01) : 77 - 82