Evaluation of noninvasive oscillometric blood pressure monitoring in anesthetized boid snakes

被引:9
|
作者
Chinnadurai, Sathya K. [1 ,2 ]
Wrenn, Amy [3 ]
DeVoe, Ryan S. [2 ]
机构
[1] N Carolina State Univ, Coll Vet Med, Dept Clin Sci, Raleigh, NC 27606 USA
[2] N Carolina State Univ, Coll Vet Med, Environm Med Consortium, Raleigh, NC 27606 USA
[3] Chatham Anim Hosp, Cary, NC 27513 USA
关键词
DOGS; CATS; ANESTHESIA; REPTILES; MANAGEMENT; MORTALITY;
D O I
10.2460/javma.234.5.625
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Objective-To determine the accuracy of a noninvasive oscillometric monitor in the measurement of arterial blood pressure in anesthetized boid snakes. Design-Evaluation study. Animals-4 boa constrictors (Boa constrictor), 2 carpet pythons (Morelia spilota), and 2 reticulated pythons (Python reticulatus). Procedures-After induction of anesthesia with isoflurane, each snake was instrumented with an arterial catheter connected to a pressure transducer and oscilloscope to obtain invasive measurements of systolic (SAP), diastolic (DAP), and mean (MAP) arterial blood pressure as well as a pressure waveform. A cuff connected to an oscillometric device was placed on the tail immediately distal to the vent for noninvasive measurements. Heart rate, respiratory rate, and invasive and noninvasive measurements of SAP DAP and MAP were obtained every 5 minutes for 45 minutes. Delivered isoflurane concentration was increased in 15-minute increments to induce hypotension. Repeatability of each device and fixed and proportional biases between devices were calculated. Results-Throughout most of the measured ranges of blood pressures, the oscillometric unit overestimated the SAP and underestimated the DAP and MAP compared with respective direct measurements. When the invasively determined SAP was > 100 mm Hg, the oscillometric unit underestimated all 3 variables. Fixed bias was significant for SAP and DAP and proportional bias was significant for SAP and MAP Conclusions and Clinical Relevance-When using an oscillometric blood pressure monitor on anesthetized boid snakes, veterinarians can potentially monitor changes in blood pressure, although the displayed readings may underestimate DAP and MAP and overestimate SAP Indirect measurements of blood pressure made with the oscillometric device cannot substitute for direct measurements. (J Am Vet Med Assoc 2009;234:625-630)
引用
收藏
页码:625 / 630
页数:6
相关论文
共 50 条
  • [31] Accuracy of oscillometric blood pressure monitoring with concurrent auscultatory blood pressure in hemodialysis patients
    Semret, M
    Zidehsarai, M
    Agarwal, R
    BLOOD PRESSURE MONITORING, 2005, 10 (05) : 249 - 255
  • [32] Noninvasive mean blood pressure monitoring
    Lakhanko, L
    Polonetsky, L
    Sidorenko, G
    Vorobiov, A
    EUROPEAN HEART JOURNAL, 2000, 21 : 103 - 103
  • [33] Noninvasive continuous blood pressure monitoring
    Poghosyan, Armen
    Mouradian, Vahram
    Hovhannisyan, Levon
    OPTICAL FIBERS AND SENSORS FOR MEDICAL DIAGNOSTICS AND TREATMENT APPLICATIONS XV, 2015, 9317
  • [34] Perioperative Noninvasive Blood Pressure Monitoring
    Kuck, Kai
    Baker, Philip D.
    ANESTHESIA AND ANALGESIA, 2018, 127 (02): : 408 - 411
  • [35] A Noninvasive Blood Pressure Monitoring System
    Balamurugan
    Reddy, S. Ramasubba
    Muthan, S.
    2016 IEEE INTERNATIONAL CONFERENCE ON RECENT TRENDS IN ELECTRONICS, INFORMATION & COMMUNICATION TECHNOLOGY (RTEICT), 2016, : 1999 - 2001
  • [36] EVALUATION OF A NEW NONINVASIVE SEMIAUTOMATIC BLOOD-PRESSURE MONITORING DEVICE
    FUKUOKA, M
    YAMORI, Y
    UENO, T
    KODAMA, S
    PARATI, G
    POMIDOSSI, G
    MANCIA, G
    ZANCHETTI, A
    CLINICAL AND EXPERIMENTAL HYPERTENSION PART A-THEORY AND PRACTICE, 1987, 9 (01) : 141 - 152
  • [37] Evaluation of the oscillometric method for noninvasive blood pressure measurement during cuff deflation and cuff inflation with reference to intra-arterial blood pressure
    Celler, Branko G.
    Yong, Andy
    Rubenis, Imants
    Butlin, Mark
    Argha, Ahmadreza
    Rehan, Rajan
    Avolio, Alberto
    JOURNAL OF HYPERTENSION, 2024, 42 (07) : 1235 - 1247
  • [38] Comparison of noninvasive estimations of central blood pressure by tonometric radial pressure waveform calibrated with oscillometric brachial blood pressure
    Hirata, K.
    Sagawa, Y.
    Kojima, I.
    Yamazaki, M.
    Shinomiya, N.
    Miyawaki, Y.
    Momomura, S. -I.
    JOURNAL OF HYPERTENSION, 2008, 26 : S490 - S490
  • [39] Method for Evaluation of Trustworthiness of Oscillometric Blood Pressure Measurements
    Koohi, Iraj
    Ahmad, Saif
    Batkin, Izmail
    Groza, Voicu Z.
    Shirmohammadi, Shervin
    Dajani, Hilmi R.
    Petriu, Emil M.
    2015 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS (MEMEA) PROCEEDINGS, 2015, : 267 - 272
  • [40] OPTIMAL SITES FOR FOREHEAD OSCILLOMETRIC BLOOD-PRESSURE MONITORING
    LEE, TK
    SILVA, FH
    EGBERT, TP
    WESTENSKOW, DR
    JOURNAL OF CLINICAL MONITORING, 1995, 11 (05): : 298 - 304