Evaluation of noninvasive oscillometric blood pressure monitoring in anesthetized boid snakes

被引:9
|
作者
Chinnadurai, Sathya K. [1 ,2 ]
Wrenn, Amy [3 ]
DeVoe, Ryan S. [2 ]
机构
[1] N Carolina State Univ, Coll Vet Med, Dept Clin Sci, Raleigh, NC 27606 USA
[2] N Carolina State Univ, Coll Vet Med, Environm Med Consortium, Raleigh, NC 27606 USA
[3] Chatham Anim Hosp, Cary, NC 27513 USA
关键词
DOGS; CATS; ANESTHESIA; REPTILES; MANAGEMENT; MORTALITY;
D O I
10.2460/javma.234.5.625
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Objective-To determine the accuracy of a noninvasive oscillometric monitor in the measurement of arterial blood pressure in anesthetized boid snakes. Design-Evaluation study. Animals-4 boa constrictors (Boa constrictor), 2 carpet pythons (Morelia spilota), and 2 reticulated pythons (Python reticulatus). Procedures-After induction of anesthesia with isoflurane, each snake was instrumented with an arterial catheter connected to a pressure transducer and oscilloscope to obtain invasive measurements of systolic (SAP), diastolic (DAP), and mean (MAP) arterial blood pressure as well as a pressure waveform. A cuff connected to an oscillometric device was placed on the tail immediately distal to the vent for noninvasive measurements. Heart rate, respiratory rate, and invasive and noninvasive measurements of SAP DAP and MAP were obtained every 5 minutes for 45 minutes. Delivered isoflurane concentration was increased in 15-minute increments to induce hypotension. Repeatability of each device and fixed and proportional biases between devices were calculated. Results-Throughout most of the measured ranges of blood pressures, the oscillometric unit overestimated the SAP and underestimated the DAP and MAP compared with respective direct measurements. When the invasively determined SAP was > 100 mm Hg, the oscillometric unit underestimated all 3 variables. Fixed bias was significant for SAP and DAP and proportional bias was significant for SAP and MAP Conclusions and Clinical Relevance-When using an oscillometric blood pressure monitor on anesthetized boid snakes, veterinarians can potentially monitor changes in blood pressure, although the displayed readings may underestimate DAP and MAP and overestimate SAP Indirect measurements of blood pressure made with the oscillometric device cannot substitute for direct measurements. (J Am Vet Med Assoc 2009;234:625-630)
引用
收藏
页码:625 / 630
页数:6
相关论文
共 50 条
  • [1] EVALUATION OF NONINVASIVE OSCILLOMETRIC BLOOD PRESSURE MONITORING IN ANESTHETIZED BENNETT'S WALLABIES (MACROPUS RUFOGRISEUS)
    Watson, Megan K.
    Chinnadurai, Sathya K.
    JOURNAL OF ZOO AND WILDLIFE MEDICINE, 2019, 50 (02) : 389 - 395
  • [2] A clinical evaluation of an oscillometric blood pressure monitor on anesthetized horses
    Branson, KR
    JOURNAL OF EQUINE VETERINARY SCIENCE, 1997, 17 (10) : 537 - 540
  • [3] Evaluation of an oscillometric blood pressure monitor for use in anesthetized cats
    Pedersen, KM
    Butler, MA
    Ersboll, AK
    Pedersen, HD
    JOURNAL OF THE AMERICAN VETERINARY MEDICAL ASSOCIATION, 2002, 221 (05) : 646 - 650
  • [4] Evaluation of an oscillometric blood pressure monitor for use in anesthetized sheep
    Trim, Cynthia M.
    Hofmeister, Erik H.
    Peroni, John F.
    Thoresen, Merrilee
    VETERINARY ANAESTHESIA AND ANALGESIA, 2013, 40 (06) : E31 - E39
  • [5] Evaluation of an indirect oscillometric blood pressure monitor in normotensive and hypotensive anesthetized dogs
    Shih, Andre
    Robertson, Sheilah
    Vigani, Alessio
    da Cunha, Anderson
    Pablo, Luisito
    Bandt, Carsten
    JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2010, 20 (03) : 313 - 318
  • [6] Evaluation of a new oscillometric blood pressure monitor in isoflurane-anesthetized dogs
    Sawyer, DC
    Guikema, AH
    Siegel, EM
    VETERINARY ANAESTHESIA AND ANALGESIA, 2004, 31 (01) : 27 - 39
  • [7] DOPPLER ULTRASONOGRAPHIC, OSCILLOMETRIC SPHYGMOMANOMETRIC, AND PHOTOPLETHYSMOGRAPHIC TECHNIQUES FOR NONINVASIVE BLOOD-PRESSURE MEASUREMENT IN ANESTHETIZED CATS
    BINNS, SH
    SISSON, DD
    BUOSCIO, DA
    SCHAEFFER, DJ
    JOURNAL OF VETERINARY INTERNAL MEDICINE, 1995, 9 (06) : 405 - 414
  • [8] 24-HOUR NONINVASIVE OSCILLOMETRIC BLOOD-PRESSURE MONITORING - EVALUATION OF THE ANTIHYPERTENSIVE CIRCADIAN PROFILE OF NITRENDIPINE
    MEYERSABELLEK, W
    SCHULTE, KL
    STREITBERG, B
    GOTZEN, R
    DISTLER, A
    JOURNAL OF CARDIOVASCULAR PHARMACOLOGY, 1988, 12 : S22 - S28
  • [9] Accuracy of an oscillometric blood pressure monitor in anesthetized pigs
    Reed, Rachel
    Barletta, Michele
    Grimes, Janet
    Mumaw, Jennifer
    Park, Hea Jin
    Giguere, Steeve
    Azain, Michael
    Fang, Xi
    Quandt, Jane
    LABORATORY ANIMALS, 2018, 52 (05) : 490 - 496
  • [10] Evaluation of oscillometric and Doppler ultrasonic devices for blood pressure measurements in anesthetized and conscious dogs
    Vachon, Catherine
    Belanger, Marie C.
    Burns, Patrick M.
    RESEARCH IN VETERINARY SCIENCE, 2014, 97 (01) : 111 - 117