Cardiac spiral dual-source CT with high pitch: a feasibility study

被引:53
|
作者
Ertel, Dirk [1 ]
Lell, Michael M. [2 ]
Harig, Frank [3 ]
Flohr, Thomas [4 ]
Schmidt, Bernhard [4 ]
Kalender, Willi A. [1 ]
机构
[1] Univ Erlangen Nurnberg, Inst Med Phys, D-91052 Erlangen, Germany
[2] Univ Erlangen Nurnberg, Dept Radiol, D-91052 Erlangen, Germany
[3] Univ Erlangen Nurnberg, Ctr Cardiac Surg, D-91052 Erlangen, Germany
[4] Siemens Healthcare, Forchheim, Germany
关键词
Dual-source CT; Cardiac imaging; Image quality; Dose; High pitch; TOMOGRAPHY CORONARY-ANGIOGRAPHY; MULTISLICE COMPUTED-TOMOGRAPHY; IMAGE QUALITY; RADIATION-EXPOSURE; SHOOT MODE; PERFORMANCE; HEART; VALIDATION; MODULATION; REDUCTION;
D O I
10.1007/s00330-009-1503-6
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Increase of pitch in spiral CT decreases data acquisition time; dual-source CT (DSCT) systems provide improved temporal resolution. We evaluated the combination of these two features. Measurements were performed using a commercial DSCT system equipped with prototype software allowing pitch factors from p=0.35 to 3.0. We measured slice sensitivity profiles as a function of pitch to assess spatial resolution in the z-direction and the contrast of structures moved periodically to measure temporal resolution. Additionally we derived modulation transfer functions to provide objective parameters; both spatial and temporal resolution were essentially unchanged even at high pitch. CT of the cardiac region of three pigs was performed at p=3.0. In vivo CT images confirmed good image quality; direct comparison with standard low-pitch phase-correlated CT image datasets showed no significant difference. For a normalized z-axis acquisition of 12 cm, the corresponding effective dose value was 2.0 mSv for the high-pitch CT protocol. We conclude that spiral DSCT imaging with a pitch of 3.0 can provide unimpaired image quality with respect to spatial and temporal resolution. Applications to cardiac and thoracic imaging with effective dose below 1 mSv are possible.
引用
收藏
页码:2357 / 2362
页数:6
相关论文
共 50 条
  • [21] High-pitch dual-source CT coronary angiography with low volumes of contrast medium
    Lembcke, Alexander
    Schwenke, Carsten
    Hein, Patrick A.
    Knobloch, Gesine
    Durmus, Tahir
    Hamm, Bernd
    Huppertz, Alexander
    EUROPEAN RADIOLOGY, 2014, 24 (01) : 120 - 127
  • [22] High-pitch dual-source CT coronary angiography with low volumes of contrast medium
    Alexander Lembcke
    Carsten Schwenke
    Patrick A. Hein
    Gesine Knobloch
    Tahir Durmus
    Bernd Hamm
    Alexander Huppertz
    European Radiology, 2014, 24 : 120 - 127
  • [23] Preliminary Evaluation of High-Pitch, Dual-Source CT Enterography for Radiation Exposure Reduction
    Horst, N.
    Boulter, D.
    Mayes, N.
    Hardie, A.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2012, 198 (05)
  • [24] Electrocardiographically triggered CT angiography of the whole aorta and coronary arteries with high-pitch dual-source CT
    张楠
    China Medical Abstracts (Internal Medicine), 2013, 30 (02) : 65 - 65
  • [25] High-pitch coronary CT angiography with third generation dual-source CT: limits of heart rate
    Gordic, Sonja
    Husarik, Daniela B.
    Desbiolles, Lotus
    Leschka, Sebastian
    Frauenfelder, Thomas
    Alkadhi, Hatem
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2014, 30 (06): : 1173 - 1179
  • [26] High-pitch coronary CT angiography with third generation dual-source CT: limits of heart rate
    Sonja Gordic
    Daniela B. Husarik
    Lotus Desbiolles
    Sebastian Leschka
    Thomas Frauenfelder
    Hatem Alkadhi
    The International Journal of Cardiovascular Imaging, 2014, 30 : 1173 - 1179
  • [27] Performance of iodine quantification through high-pitch dual-source photon-counting CT: a phantom study
    Peng Liu
    Shanshui Zhou
    Haipeng Dong
    Jiqiang Li
    Zhihan Xu
    Shushen Lin
    Wenjie Yang
    Fuhua Yan
    Le Qin
    Japanese Journal of Radiology, 2025, 43 (2) : 309 - 318
  • [28] High Pitch Dual-Source Whole Aorta CT Angiography in the Detection of Coronary Arteries: A Feasibility Study of Using lodixanol 270 and 100 kVp with Iterative Reconstruction
    Shen, Yanguang
    Fan, Zhanming
    Sun, Zhonghua
    Xu, Lei
    Li, Yu
    Zhang, Nan
    Yan, Zixu
    JOURNAL OF MEDICAL IMAGING AND HEALTH INFORMATICS, 2015, 5 (01) : 117 - 125
  • [29] High-pitch, high temporal resolution, multi-energy cardiac imaging on a dual-source photon-counting-detector CT
    Ahmed, Zaki
    Campeau, David
    Gong, Hao
    Rajendran, Kishore
    Rajiah, Prabhakar
    McCollough, Cynthia
    Leng, Shuai
    MEDICAL PHYSICS, 2023, 50 (03) : 1428 - 1435
  • [30] High-pitch dual-source CT coronary angiography: systolic data acquisition at high heart rates
    Robert Goetti
    Gudrun Feuchtner
    Paul Stolzmann
    Lotus Desbiolles
    Michael Alexander Fischer
    Christoph Karlo
    Stephan Baumueller
    Hans Scheffel
    Hatem Alkadhi
    Sebastian Leschka
    European Radiology, 2010, 20 : 2565 - 2571