Reducing and correcting for contamination of ecosystem water stable isotopes measured by isotope ratio infrared spectroscopy

被引:25
|
作者
Schmidt, Markus [1 ]
Maseyk, Kadmiel [1 ]
Lett, Celine [1 ]
Biron, Philippe [1 ]
Richard, Patricia [1 ]
Bariac, Thierry [1 ]
Seibt, Ulli [1 ,2 ]
机构
[1] Univ Paris 06, BIOEMCO, INRA, ENS,IRD,CNRS,UMR 7618, F-78850 Thiverval Grignon, France
[2] Univ Calif Los Angeles, Dept Atmospher & Ocean Sci, Los Angeles, CA 90095 USA
基金
欧洲研究理事会;
关键词
RING-DOWN SPECTROSCOPY; MASS-SPECTROMETRY; SPATIAL VARIATION; LEAF WATER; SOIL-WATER; PLANT; EXTRACTION; O-18; DISCREPANCIES; DELTA-O-18;
D O I
10.1002/rcm.5317
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Concern exists about the suitability of laser spectroscopic instruments for the measurement of the O-18/O-16 and H-2/H-1 values of liquid samples other than pure water. It is possible to derive erroneous isotope values due to optical interference by certain organic compounds, including some commonly present in ecosystem-derived samples such as leaf or soil waters. Here we investigated the reliability of wavelength-scanned cavity ring-down spectroscopy (CRDS) O-18/O-16 and H-2/H-1 measurements from a range of ecosystem-derived waters, through comparison with isotope ratio mass spectrometry (IRMS). We tested the residual of the spectral fit S-r calculated by the CRDS instrument as a means to quantify the difference between the CRDS and IRMS delta-values. There was very good overall agreement between the CRDS and IRMS values for both isotopes, but differences of up to 2.3 parts per thousand (delta O-18 values) and 23 parts per thousand (delta H-2 values) were observed in leaf water extracts from Citrus limon and Alnus cordata. The S-r statistic successfully detected contaminated samples. Treatment of Citrus leaf water with activated charcoal reduced, but did not eliminate, delta H-2(CRDS)-delta H-2(IRMS) linearly for the tested range of 0-20 % charcoal. The effect of distillation temperature on the degree of contamination was large, particularly for delta H-2 values but variable, resulting in positive, negative or no correlation with distillation temperature. S-r and delta(CRDS) - delta(IRMS) were highly correlated, in particular for delta H-2 values, across the range of samples that we tested, indicating the potential to use this relationship to correct the delta-values of contaminated plant water extracts. We also examined the sensitivity of the CRDS system to changes in the temperature of its operating environment. We found that temperature changes >= 4 degrees C for delta O-18 values and >= 10 degrees C for delta H-2 values resulted in errors larger than the CRDS precision for the respective isotopes and advise the use of such instruments only in sufficiently temperature-stabilised environments. Copyright (C) 2011 John Wiley & Sons, Ltd.
引用
收藏
页码:141 / 153
页数:13
相关论文
共 50 条
  • [21] Analytical techniques in biomedical stable isotope applications: (isotope ratio) mass spectrometry or infrared spectrometry?
    Stellaard, F
    Elzinga, H
    ISOTOPES IN ENVIRONMENTAL AND HEALTH STUDIES, 2005, 41 (04) : 345 - 361
  • [22] Determination of inorganic chlorine stable isotopes by continuous flow isotope ratio mass spectrometry
    Shouakar-Stash, O
    Drimmie, RJ
    Frape, SK
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2005, 19 (02) : 121 - 127
  • [23] Measuring natural abundance of stable S isotopes in soil by isotope ratio mass spectrometry
    Eriksen, J
    COMMUNICATIONS IN SOIL SCIENCE AND PLANT ANALYSIS, 1996, 27 (5-8) : 1251 - 1264
  • [24] Using stable isotopes and isotope ratio mass spectrometry to study mineral metabolism in humans
    Griffin, I.J. (igriffin@bcm.tmc.edu), 1600, Royal Society of Chemistry (17):
  • [25] Online Determination of Boron Isotope Ratio in Boron Trifluoride by Infrared Spectroscopy
    Zhang, Weijiang
    Tang, Yin
    Xu, Jiao
    APPLIED SCIENCES-BASEL, 2018, 8 (12):
  • [26] Simultaneous Raman and Infrared Spectroscopy of Stable Isotope Labelled Escherichia coli
    Lima, Cassio
    Muhamadali, Howbeer
    Goodacre, Royston
    SENSORS, 2022, 22 (10)
  • [27] Triple oxygen stable isotope analysis of nitrite measured using continuous flow isotope ratio mass spectrometry *
    Walters, Wendell W.
    Hastings, Meredith G.
    METHODSX, 2023, 11
  • [28] Determination of bromine stable isotopes using continuous-flow isotope ratio mass spectrometry
    Shouakar-Stash, O
    Frape, SK
    Drimmie, RJ
    ANALYTICAL CHEMISTRY, 2005, 77 (13) : 4027 - 4033
  • [29] STABLE CARBON ISOTOPE RATIO OF SUBMERGED FRESH-WATER MACROPHYTES
    LAZERTE, BD
    SZALADOS, JE
    LIMNOLOGY AND OCEANOGRAPHY, 1982, 27 (03) : 413 - 418
  • [30] The influence of environmental water on the hydrogen stable isotope ratio in aquatic consumers
    Christopher T. Solomon
    Jonathan J. Cole
    Richard R. Doucett
    Michael L. Pace
    Nicholas D. Preston
    Laura E. Smith
    Brian C. Weidel
    Oecologia, 2009, 161 : 313 - 324