Determination of bromine stable isotopes using continuous-flow isotope ratio mass spectrometry

被引:54
|
作者
Shouakar-Stash, O [1 ]
Frape, SK [1 ]
Drimmie, RJ [1 ]
机构
[1] Univ Waterloo, Dept Earth Sci, Waterloo, ON N2L 3G1, Canada
关键词
D O I
10.1021/ac048318n
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new methodology for bromine stable isotope determination by continuous-flow isotope ratio mass spectrometry (CF-IRMS) was developed. The technique was tested on inorganic samples. Inorganic bromide was precipitated in the form of silver bromide by using silver nitrate in a standard methodology. Bromine stable isotope analysis was carried out on methyl bromide (CH3Br) after converting silver bromide to methyl bromide by reacting it with methyl iodide (CH3I). The system used in this study is an IsoPrime IRMS, with analytical capabilities of both dual-inlet and continuous-flow modes coupled with an Agilent 6890 GC equipped with a CTC Analytics Combi-PAL autosampler. This new technique measures samples as small as 0.2 mg of AgBr (1 mu mol of Br-). The bromine stable isotope analysis using continuous flow technology showed excellent precision and accuracy. The internal precision using pure methyl bromide gas is better than +/- 0.03%o (+/- SD); the external precision using seawater standard is better than +/- 0.06%o (SD) for n = 12. Moreover, the sample analysis time is 16 min, as compared to 75 min needed in previous techniques. Ibis allows for 50 samples to be analyzed in 1 day, as compared to 8 samples using the conventional techniques. A series of natural saline formation waters and brines from sedimentary and crystalline rock environments was measured by this new methodology to test the potential natural range of delta Br-81. The bromine isotopic composition of the samples ranged from 0.00 to + 1.80 parts per thousand relative to standard mean ocean bromide (SMOB). Initial trends and distinctive isotopic difference were noticed between crystalline shield brines and sedimentary formation brines.
引用
收藏
页码:4027 / 4033
页数:7
相关论文
共 50 条
  • [1] Determination of inorganic chlorine stable isotopes by continuous flow isotope ratio mass spectrometry
    Shouakar-Stash, O
    Drimmie, RJ
    Frape, SK
    [J]. RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2005, 19 (02) : 121 - 127
  • [2] An analytical system for stable isotope analysis on carbon monoxide using continuous-flow isotope-ratio mass spectrometry
    Pathirana, S. L.
    van der Veen, C.
    Popa, M. E.
    Roeckmann, T.
    [J]. ATMOSPHERIC MEASUREMENT TECHNIQUES, 2015, 8 (12) : 5315 - 5324
  • [3] Membrane permeation continuous-flow isotope ratio mass spectrometry for on-line carbon isotope ratio determination
    Tremblay, Patrice
    Savard, Martine M.
    Smirnoff, Anna
    Paquin, Real
    [J]. RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2009, 23 (14) : 2213 - 2220
  • [4] Stable chlorine isotope ratio analysis of subnanomolar level methyl chloride by continuous-flow isotope ratio mass spectrometry
    Konno, Uta
    Takai, Ken
    Kawagucci, Shinsuke
    [J]. GEOCHEMICAL JOURNAL, 2013, 47 (04) : 469 - 473
  • [5] RECENT PROGRESS IN CONTINUOUS-FLOW ISOTOPE RATIO MASS-SPECTROMETRY
    PRESTON, T
    BARRIE, A
    [J]. AMERICAN LABORATORY, 1991, 23 (18) : H32 - &
  • [6] High-precision continuous-flow isotope ratio mass spectrometry
    Brenna, JT
    Corso, TN
    Tobias, HJ
    Caimi, RJ
    [J]. MASS SPECTROMETRY REVIEWS, 1997, 16 (05) : 227 - 258
  • [7] DETERMINATION OF N-15 AT SUBMICROGRAM LEVELS OF NITROGEN USING AUTOMATED CONTINUOUS-FLOW ISOTOPE RATIO MASS-SPECTROMETRY
    OWENS, NJP
    REES, AP
    [J]. ANALYST, 1989, 114 (12) : 1655 - 1657
  • [8] High-precision, automated stable isotope analysis of atmospheric methane and carbon dioxide using continuous-flow isotope-ratio mass spectrometry
    Fisher, R
    Lowry, D
    Wilkin, O
    Sriskantharajah, S
    Nisbet, EG
    [J]. RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2006, 20 (02) : 200 - 208
  • [9] Triple oxygen stable isotope analysis of nitrite measured using continuous flow isotope ratio mass spectrometry *
    Walters, Wendell W.
    Hastings, Meredith G.
    [J]. METHODSX, 2023, 11
  • [10] Recent developments in the analysis of light isotopes by continuous flow isotope ratio mass spectrometry
    Midwood, AJ