Performance evaluation of siemens CTI ECAT EXACT 47 PET scanner using NEMA NU2-2001

被引:0
|
作者
Kim, JS [1 ]
Lee, JS [1 ]
Lee, DS [1 ]
Jung, JK [1 ]
Lee, MC [1 ]
机构
[1] Seoul Natl Univ, Coll Med, Dept Nucl Med, Seoul, South Korea
关键词
D O I
暂无
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
In this study, system performance of Siemens CTI ECAT EXACT 47 PET scanner was measured using NEMA NU2-2001 standard. ECAT EXACT 47 is a BGO crystal based PET scanner. Coincidence window are set to 12 ns, and energy window are set to 250 similar to 650 keV. For the spatial resolution measurement, F-18 point source was placed at 1cm and 10 cm offset from the center of transaxial FOV. Images were reconstructed using FBP with ramp filter. To measure the system sensitivity, NEMA sensitivity phantom filled with F-18 solution and surrounded by 1 similar to 5 aluminum sleeves were scanned at the center of transaxial FOV and 10 cm offset from the center. Attenuation free values of sensitivity were estimated by extrapolating data to the zero wall thickness. NEMA scatter phantom with length of 70 cm was filled with F-18 solution (2D: 2,900 MBq), and coincidence count rates were measured for 7 half-lives to obtain noise equivalent count rate (NECR) and scatter fraction. Transverse and axial resolutions at 1cm offset from the center were 9.1 and 9.8 mm (FBP, 2D and 3D), Sensitivity values were 709 (2D), 3332 (3D) cps/MBq at the center and 796(2D), 4050 (3D) cps/MBq at 10 cm offset from the center. Scatter fractions were 0.19 (2D). Peak true count rate and NECR were 64.0 kcps@37.5 kBq/mL and 49.6 kcps@37.5 kBq/mL in 2D. Information about the performance of EXACT 47 PET scanner reported in this study will be useful for the quantitative analysis of data and determination of optimal image acquisition protocols using this widely used scanner for clinical and research purposes.
引用
收藏
页码:3118 / 3120
页数:3
相关论文
共 50 条
  • [21] Performance evaluation of a small-animal PET scanner with 213 mm axis using NEMA NU 4-2008
    Wei, Qing
    Wang, Yingjie
    Huang, Xianchao
    Li, Daowu
    Zheng, Yushuang
    Wang, Peilin
    Sun, Xiaoli
    Chai, Pei
    Han, Xiaorou
    Liu, Shuangquan
    Feng, Baotong
    Zhou, Wei
    Zeng, Xiangtao
    Zhu, Meiling
    Zhang, Zhiming
    Wei, Long
    MEDICAL PHYSICS, 2024,
  • [22] Performance evaluation of the PET Sub-Systems of two PET/CT imagers according to NEMA standard NU 2-2001
    Strigl, Markus
    Martinez, Maria-Jose
    Zach, Christian
    Ziegler, Sibylle I.
    Hahn, Klaus
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2006, 16 (01): : 83 - 91
  • [23] NEMA NU 2-2007 performance measurements of the Siemens Inveon™ preclinical small animal PET system
    Kemp, Brad J.
    Hruska, Carrie B.
    McFarland, Aaron R.
    Lenox, Mark W.
    Lowe, Val J.
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (08): : 2359 - 2376
  • [24] Performance evaluation of the whole-body PET scanner ECAT EXACT HR+ following the IEC standard
    Adam, LE
    Zaers, J
    Ostertag, H
    Trojan, H
    Bellemann, ME
    Brix, G
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) : 1172 - 1179
  • [25] Performance comparison of two commercial BGO-based PET/CT scanners using NEMA NU 2-2001
    Bolard, Gregory
    Prior, John O.
    Modolo, Luca
    Delaloye, Angelika Bischof
    Kosinski, Marek
    Wastiel, Claude
    Malterre, Jerome
    Bulling, Shelley
    Bochud, Francois
    Verdun, Francis R.
    MEDICAL PHYSICS, 2007, 34 (07) : 2708 - 2717
  • [26] Comparison of NEMA NU 4-2008 vs NEMA NU 2-2001 for the performance evaluation of the microPET R4 system
    Popota, F. D.
    Aguiar, P.
    Fernandez, Y.
    Lois, C.
    Pareto, D.
    Ros, D.
    Pavia, J.
    Gispert, J. D.
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 2706 - +
  • [27] NEMA NU2-2012 performance evaluation of the Philips Vereos PET/CT System
    Rausch, I.
    Ruiz, A.
    Valverde-Pascual, I.
    Cal-Gonzalez, J.
    Beyer, T.
    Carrio, I.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2018, 45 : S715 - S715
  • [28] NEMA NU-4 2008 performance evaluation of Xtrim-PET: A prototype SiPM-based preclinical scanner
    Amirrashedi, Mahsa
    Sarkar, Saeed
    Ghafarian, Pardis
    Shahraki, Reza Hashemi
    Geramifar, Parham
    Zaidi, Habib
    Ay, Mohammad Reza
    MEDICAL PHYSICS, 2019, 46 (11) : 4816 - 4825
  • [29] Performance characteristics of a new development dedicated breast PET scanner evaluated with the NEMA NU 2-2008 standard
    Inoue, Mika
    Nakamoto, Yuji
    Matsumoto, Keiichi
    Kawase, Shigeto
    Yamakawa, Yoshiyuki
    Ohi, Junichi
    Kitamura, Keishi
    Togashi, Kaori
    Higashimura, Kyoji
    Onoguchi, Masahisa
    JOURNAL OF NUCLEAR MEDICINE, 2011, 52
  • [30] Performance evaluation of the Mediso PET scanner to image three mice simultaneously based on the NEMA NU4-2008 standards
    Li, Guiyu
    Ma, Xiaowei
    Wang, Jing
    JOURNAL OF NUCLEAR MEDICINE, 2017, 58