3D Monte Carlo dosimetry of intraoperative electron radiation therapy (IOERT)

被引:11
|
作者
Alhamada, Husein [1 ,2 ]
Simon, Stephane [1 ,2 ]
Philippson, Catherine [2 ]
Vandekerkhove, Christophe [2 ]
Jourani, Younes [2 ]
Pauly, Nicolas [1 ]
VanGestel, Dirk [2 ]
Reynaert, Nick [1 ,2 ]
机构
[1] Univ Libre Bruxelles, Ecole Polytech, Nucl Metrol, Brussels, Belgium
[2] Inst Jules Bordet, Radiotherapy Dept, Brussels, Belgium
关键词
DOSE CALCULATIONS; RADIOTHERAPY; IORT; DISTRIBUTIONS; APPLICATOR; PHOTON;
D O I
10.1016/j.ejmp.2018.12.037
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: This paper studies the feasibility of using Monte Carlo (MC) for treatment planning of intraoperative electron radiation therapy (IOERT) procedure to get 3D dose by using patient's CT images. Methods: The IOERT treatment planning was performed using the following successive steps: I) The Mobetron 1000 degrees machine was modelled with the EGSnrc MC codes. II) The MC model was validated with measurements of percentage depth doses and profiles for three energies (12, 9, 6) MeV. III) CT images were imported as DICOM files. IV) Contouring of the planning target volume (PTV) and the organs at risk was done by the radiation oncologist. V) The medical physicist with the radiation oncologist, had chosen the same parameters of IOERT procedures like energy, applicator (type, size) and using or not bolus. VI) Finally, dose calculation and analysis of 3D maps was carried out. Results: The tuning process of the MC model provides good results, as the maximum value of the root mean square deviation (RMSD) was less than 3% between the MC simulated PDDs and the measured PDDs. The contouring and dose analysis review were easy to conduct for the classical treatment planning system. The radiation oncologist had many tools for dose analysis such as DVH and color wash for all the slides. Summation of the 3D dose of IOERT with other radiotherapy plans is possible and helpful for total dose estimation. Archiving and documentation is as good as treatment planning system (TPS). Conclusions: The method displayed in this paper provides a step forward for IOERT Dosimetry and allows to obtain accurate dosimetry of treated volumes.
引用
收藏
页码:207 / 214
页数:8
相关论文
共 50 条
  • [41] Bone and mucosal dosimetry in skin radiation therapy: a Monte Carlo study using kilovoltage photon and megavoltage electron beams
    Chow, James C. L.
    Jiang, Runqing
    PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (12): : 3885 - 3899
  • [42] 3D dosimetry for proton therapy
    Beddar, S.
    10TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2019, 1305
  • [43] Comparison of Electron Monte Carlo and Pencil-Beam Algorithms in Clinical Electron Beam Radiation Dosimetry
    Liu, Z.
    Hargrove, I.
    Tazi, A.
    Barrett, J.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [44] 3D Range Modulator for Proton Therapy: Development and Monte Carlo Simulations
    Simeonov, Y.
    Penchev, P.
    Printz, Ringbk T.
    Engenhart-Cabillic, R.
    Weber, U.
    Zink, K.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2017, 193 : S10 - S10
  • [45] MONTE CARLO DOSIMETRY TO PREPARE THE PRE-CLINICAL TRIALS IN MINIBEAM RADIATION THERAPY
    Alonso, Y. Prezado
    Thengumpallil, S.
    Martinez-Rovira, I.
    Renier, M.
    Bravin, A.
    RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S221 - S221
  • [46] MONTE-CARLO DOSIMETRY FOR STEREOTACTIC SYNCHROTRON RADIATION THERAPY (SSRT) FOR BRAIN TUMORS
    Alonso, Y. Prezado
    Adam, J. F.
    Bravin, A.
    Edouard, M.
    Ellaume, H.
    Esteve, F.
    Nemoz, C.
    Renier, M.
    RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S115 - S115
  • [47] A CUDA Monte Carlo simulator for radiation therapy dosimetry based on Geant4
    Henderson, N.
    Murakami, K.
    Amako, K.
    Asai, M.
    Aso, T.
    Dotti, A.
    Kimura, A.
    Gerritsen, M.
    Kurashige, H.
    Perl, J.
    Sasaki, T.
    SNA + MC 2013 - JOINT INTERNATIONAL CONFERENCE ON SUPERCOMPUTING IN NUCLEAR APPLICATIONS + MONTE CARLO, 2014,
  • [48] Minibeam radiation therapy: A micro- and nano-dosimetry Monte Carlo study
    Dos Santos, M.
    Delorme, R.
    Salmon, R.
    Prezado, Y.
    MEDICAL PHYSICS, 2020, 47 (03) : 1379 - 1390
  • [49] Towards a Guidance System to Aid in the Dosimetry Calculation of Intraoperative Electron Radiation Therapy
    Portales, Cristina
    Gimeno, Jesus
    Vera, Lucia
    Fernandez, Marcos
    JOURNAL OF IMAGING, 2015, 1 (01) : 180 - 192
  • [50] Comparative dosimetry of diode and diamond detectors in electron beams for intraoperative radiation therapy
    Björk, P
    Knöös, T
    Nilsson, P
    MEDICAL PHYSICS, 2000, 27 (11) : 2580 - 2588