3D Monte Carlo dosimetry of intraoperative electron radiation therapy (IOERT)

被引:11
|
作者
Alhamada, Husein [1 ,2 ]
Simon, Stephane [1 ,2 ]
Philippson, Catherine [2 ]
Vandekerkhove, Christophe [2 ]
Jourani, Younes [2 ]
Pauly, Nicolas [1 ]
VanGestel, Dirk [2 ]
Reynaert, Nick [1 ,2 ]
机构
[1] Univ Libre Bruxelles, Ecole Polytech, Nucl Metrol, Brussels, Belgium
[2] Inst Jules Bordet, Radiotherapy Dept, Brussels, Belgium
关键词
DOSE CALCULATIONS; RADIOTHERAPY; IORT; DISTRIBUTIONS; APPLICATOR; PHOTON;
D O I
10.1016/j.ejmp.2018.12.037
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: This paper studies the feasibility of using Monte Carlo (MC) for treatment planning of intraoperative electron radiation therapy (IOERT) procedure to get 3D dose by using patient's CT images. Methods: The IOERT treatment planning was performed using the following successive steps: I) The Mobetron 1000 degrees machine was modelled with the EGSnrc MC codes. II) The MC model was validated with measurements of percentage depth doses and profiles for three energies (12, 9, 6) MeV. III) CT images were imported as DICOM files. IV) Contouring of the planning target volume (PTV) and the organs at risk was done by the radiation oncologist. V) The medical physicist with the radiation oncologist, had chosen the same parameters of IOERT procedures like energy, applicator (type, size) and using or not bolus. VI) Finally, dose calculation and analysis of 3D maps was carried out. Results: The tuning process of the MC model provides good results, as the maximum value of the root mean square deviation (RMSD) was less than 3% between the MC simulated PDDs and the measured PDDs. The contouring and dose analysis review were easy to conduct for the classical treatment planning system. The radiation oncologist had many tools for dose analysis such as DVH and color wash for all the slides. Summation of the 3D dose of IOERT with other radiotherapy plans is possible and helpful for total dose estimation. Archiving and documentation is as good as treatment planning system (TPS). Conclusions: The method displayed in this paper provides a step forward for IOERT Dosimetry and allows to obtain accurate dosimetry of treated volumes.
引用
收藏
页码:207 / 214
页数:8
相关论文
共 50 条
  • [21] Assessment of intraoperative 3D imaging alternatives for IOERT dose estimation
    Garcia-Vazquez, Veronica
    Marinetto, Eugenio
    Guerra, Pedro
    Valdivieso-Casique, Manlio Fabio
    Angel Calva, Felipe
    Alvarado-Vasquez, Eduardo
    Sole, Claudio Vicente
    Vosburgh, Kirby Gannett
    Desco, Manuel
    Pascau, Javier
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2017, 27 (03): : 218 - 231
  • [22] A practical and accurate 3D dosimetry system for radiation therapy
    Guo, P.
    Adamovics, J.
    Oldham, M.
    MEDICAL PHYSICS, 2006, 33 (06) : 2193 - 2193
  • [23] RAPIDARCTM TREATMENT VERIFICATION IN 3D USING GEL DOSIMETRY AND MONTE CARLO CALCULATIONS
    Ceberg, S.
    Gagne, I.
    Scherman, J. Bengtsson
    Gustafsson, H.
    Korreman, S.
    Kjaer-Kristoffersen, F.
    Hilts, M.
    Back, S.
    RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S149 - S150
  • [24] Clinical Applications and Monte Carlo Validation of a fully integrated 3D dosimetry system
    Lakrad, Kawtar
    Quinn, Benjamin
    Oldham, Mark
    Adamson, Justus
    13TH INTERNATIONAL CONFERENCE ON 3D DOSIMETRY, IC3DDOSE 2024, 2024, 2799
  • [25] INITIAL CLINICAL EXPERIENCE OF PENCIL BEAM DOSE MODELLING FOR INTRAOPERATIVE ELECTRON RADIATION THERAPY (IOERT)
    Santos Miranda, J. A.
    Pascau, J.
    Lardies, M. D.
    Desco, M.
    Calvo, F.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S246 - S246
  • [26] Feasibility assessment of the interactive use of a Monte Carlo algorithm in treatment planning for intraoperative electron radiation therapy
    Guerra, Pedro
    Udias, Jose M.
    Herranz, Elena
    Antonio Santos-Miranda, Juan
    Herraiz, Joaquin L.
    Valdivieso, Manlio F.
    Rodriguez, Raul
    Calama, Juan A.
    Pascau, Javier
    Calvo, Felipe A.
    Illana, Carlos
    Ledesma-Carbayo, Maria J.
    Santos, Andres
    PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (23): : 7159 - 7179
  • [27] A Novel Treatment Planning Design for Intraoperative Electron Radiation Therapy (IOERT) using an Electronic Board
    Pouyanrad Z.
    Zafarghandi M.S.
    Setayeshi S.
    Journal of Biomedical Physics and Engineering, 2022, 12 (05): : I - X
  • [28] Intraoperative Electron Beam Radiation Therapy (IOERT) in the Management of Locally Advanced or Recurrent Cervical Cancer
    Barney, B. M.
    Petersen, I. A.
    Haddock, M. G.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 84 (03): : S426 - S426
  • [29] Monte Carlo application in radiation dosimetry and radiotherapy
    Ma, CM
    Jiang, SB
    Pawlicki, T
    Li, JS
    Deng, J
    Kapur, A
    Mora, G
    Lee, MC
    Mok, E
    Findley, D
    Boyer, AL
    PROCEEDINGS OF THE EIGHTH INTERNATIONAL CONFERENCE ON CALORIMETRY IN HIGH ENERGY PHYSICS, 2000, : 681 - 692
  • [30] RapidArc treatment verification in 3D using polymer gel dosimetry and Monte Carlo simulation
    Ceberg, Sofie
    Gagne, Isabelle
    Gustafsson, Helen
    Scherman, Jonas Bengtsson
    Korreman, Stine S.
    Kjaer-Kristoffersen, Flemming
    Hilts, Michelle
    Back, Sven A. J.