Comparative dosimetry of diode and diamond detectors in electron beams for intraoperative radiation therapy

被引:0
|
作者
Björk, P [1 ]
Knöös, T [1 ]
Nilsson, P [1 ]
机构
[1] Univ Lund Hosp, Dept Radiat Phys, S-22185 Lund, Sweden
关键词
diamond detector; silicon diode; intraoperative radiation therapy; electron beam dosimetry; dose distribution analysis;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of the present study is to examine the validity of using silicon semiconductor detectors in degraded electron beams with a broad energy spectrum and a wide angular distribution. A comparison is made with diamond detector measurements, which is the dosimeter considered to give the best results provided that dose rate effects are corrected for. Two-dimensional relative absorbed dose distributions in electron beams (6-20 MeV) for intraoperative radiation therapy (IORT) are measured in a water phantom. To quantify deviations between the detectors, a dose comparison tool that simultaneously examines the dose difference and distance to agreement (DTA) is used to evaluate the results in low- and high-dose gradient regions, respectively. Uncertainties of the experimental measurement setup (+/-1% and +/-0.5 mm) are taken into account by calculating a composite distribution that fails this dose-difference and DTA acceptance limit. Thus, the resulting area of disagreement should be related to differences in detector performance. The dose distributions obtained with the diode are generally in very good agreement with diamond detector measurements. The buildup region and the dose falloff region show good agreement with increasing electron energy, while the region outside the radiation field close to the water surface shows an increased difference with energy. The small discrepancies in the composite distributions are due to several factors: (a) variation of the silicon-to-water collision stopping-power ratio with electron energy, (b) a more pronounced directional dependence for diodes than for diamonds, and (c) variation of the electron fluence perturbation correction factor with depth. For all investigated treatment cones and energies, the deviation is within dose-difference and DTA acceptance criteria of +/-3% and +/-1 mm, respectively. Therefore, p-type silicon diodes are well suited, in the sense that they give results in close agreement with diamond detectors, for practical measurements of relative absorbed dose distributions in degraded electron beams used for IORT. (C) 2000 American Association of Physicists in Medicine. [S0094-2405(00)00611-8].
引用
收藏
页码:2580 / 2588
页数:9
相关论文
共 50 条
  • [1] The dose rate dependence of synthetic diamond detectors in the relative dosimetry of high-energy electron therapy beams
    Ade, N.
    Nam, T. L.
    Derry, T. E.
    Mhlanga, S. H.
    RADIATION PHYSICS AND CHEMISTRY, 2014, 98 : 155 - 162
  • [2] Comparison of a novel Schottky diamond diode with reference detectors for dosimetry of flattening filter free beams
    Stravato, A.
    Lobefalo, F.
    Pimpinella, M.
    Reggiori, G.
    Marinelli, M.
    Verona-Rinati, G.
    Gaudino, A.
    Mancosu, P.
    Scorsetti, M.
    Tomatis, S.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S148 - S149
  • [3] Dosimetry characteristics of degraded electron beams investigated by Monte Carlo calculations in a setup for intraoperative radiation therapy
    Björk, P
    Nilsson, P
    Knöös, T
    PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (02): : 239 - 256
  • [4] DIAMOND DETECTORS IN RELATIVE DOSIMETRY OF PHOTON, ELECTRON AND PROTON RADIATION-FIELDS
    KHRUNOV, VS
    MARTYNOV, SS
    VATNITSKY, SM
    ERMAKOV, IA
    CHERVJAKOV, AM
    KARLIN, DL
    FOMINYCH, VI
    TARBEYEV, YV
    RADIATION PROTECTION DOSIMETRY, 1990, 33 (1-4) : 155 - 157
  • [5] Operating parameters of CVD diamond detectors for radiation dosimetry
    Betzel, G. T.
    Lansley, S. P.
    Baluti, F.
    Reinisch, L.
    Meyer, J.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 614 (01): : 130 - 136
  • [6] PERFORMANCE OF PLATINUM-DOPED SILICON DIODE DETECTORS IN RADIATION-THERAPY DOSIMETRY
    DIXON, RL
    EKSTRAND, KE
    MEDICAL PHYSICS, 1987, 14 (04) : 705 - 705
  • [7] COMPUTED DOSIMETRY OF SCANNED ELECTRON AND PHOTON BEAMS FOR RADIATION-THERAPY
    BRAHME, A
    NAFSTADIUS, P
    MEDICAL PHYSICS, 1984, 11 (03) : 381 - 381
  • [8] Clinical implementation of MOSFET detectors for dosimetry in electron beams
    Bloemen-van Gurp, Esther J.
    Minken, Andre W. H.
    Mijnheer, Ben J.
    Dehing-Oberye, Cary J. G.
    Lambin, Philippe
    RADIOTHERAPY AND ONCOLOGY, 2006, 80 (03) : 288 - 295
  • [9] The implementation of mosfet detectors for clinical dosimetry in electron beams
    Bloemen-Van Gurp, E
    Minken, A
    Dehing-Oberije, C
    Lambin, P
    Mijnheer, B
    RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S470 - S470
  • [10] Towards a Guidance System to Aid in the Dosimetry Calculation of Intraoperative Electron Radiation Therapy
    Portales, Cristina
    Gimeno, Jesus
    Vera, Lucia
    Fernandez, Marcos
    JOURNAL OF IMAGING, 2015, 1 (01) : 180 - 192