Comparative dosimetry of diode and diamond detectors in electron beams for intraoperative radiation therapy

被引:0
|
作者
Björk, P [1 ]
Knöös, T [1 ]
Nilsson, P [1 ]
机构
[1] Univ Lund Hosp, Dept Radiat Phys, S-22185 Lund, Sweden
关键词
diamond detector; silicon diode; intraoperative radiation therapy; electron beam dosimetry; dose distribution analysis;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of the present study is to examine the validity of using silicon semiconductor detectors in degraded electron beams with a broad energy spectrum and a wide angular distribution. A comparison is made with diamond detector measurements, which is the dosimeter considered to give the best results provided that dose rate effects are corrected for. Two-dimensional relative absorbed dose distributions in electron beams (6-20 MeV) for intraoperative radiation therapy (IORT) are measured in a water phantom. To quantify deviations between the detectors, a dose comparison tool that simultaneously examines the dose difference and distance to agreement (DTA) is used to evaluate the results in low- and high-dose gradient regions, respectively. Uncertainties of the experimental measurement setup (+/-1% and +/-0.5 mm) are taken into account by calculating a composite distribution that fails this dose-difference and DTA acceptance limit. Thus, the resulting area of disagreement should be related to differences in detector performance. The dose distributions obtained with the diode are generally in very good agreement with diamond detector measurements. The buildup region and the dose falloff region show good agreement with increasing electron energy, while the region outside the radiation field close to the water surface shows an increased difference with energy. The small discrepancies in the composite distributions are due to several factors: (a) variation of the silicon-to-water collision stopping-power ratio with electron energy, (b) a more pronounced directional dependence for diodes than for diamonds, and (c) variation of the electron fluence perturbation correction factor with depth. For all investigated treatment cones and energies, the deviation is within dose-difference and DTA acceptance criteria of +/-3% and +/-1 mm, respectively. Therefore, p-type silicon diodes are well suited, in the sense that they give results in close agreement with diamond detectors, for practical measurements of relative absorbed dose distributions in degraded electron beams used for IORT. (C) 2000 American Association of Physicists in Medicine. [S0094-2405(00)00611-8].
引用
收藏
页码:2580 / 2588
页数:9
相关论文
共 50 条
  • [31] The performance of a multi guard ring (MGR) diode for clinical electron beams dosimetry
    Nascimento, C. R.
    Asfora, V. K.
    Goncalves, J. A. C.
    Khoury, H. J.
    Barros, V. S. M.
    Kalil, L. F.
    Bueno, C. C.
    APPLIED RADIATION AND ISOTOPES, 2018, 141 : 112 - 117
  • [32] The abilities of LiF thermoluminescent detectors for dosimetry at boron neutron capture therapy beams
    Toivonen, M
    Chernov, V
    Jungner, H
    Aschan, C
    Toivonen, A
    RADIATION MEASUREMENTS, 1998, 29 (3-4) : 373 - 377
  • [33] Intraoperative Electron Radiation Therapy in Retroperitoneal Sarcoma
    Roeder, Falk
    Alldinger, Ingo
    Uhl, Matthias
    Saleh-Ebrahimi, Ladan
    Schimmack, Simon
    Mechtersheimer, Gunhild
    Buechler, Markus W.
    Debus, Juergen
    Krempien, Robert
    Ulrich, Alexis
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2018, 100 (02): : 516 - 527
  • [34] THE PHYSICS OF ELECTRON INTRAOPERATIVE RADIATION-THERAPY
    LANZL, LH
    KAO, M
    ROZENFELD, M
    SABBAS, A
    PAGNAMENTA, A
    SOUTH AFRICAN JOURNAL OF SCIENCE, 1987, 83 (09) : 570 - 570
  • [35] Real-time in vivo dosimetry using micro-MOSFET detectors during intraoperative electron beam radiation therapy in early-stage breast cancer
    Ciocca, M
    Piazzi, V
    Lazzari, R
    Vavassori, A
    Luini, A
    Veronesi, P
    Galimberti, V
    Intra, M
    Guido, A
    Tosi, G
    Veronesi, U
    Orecchia, R
    RADIOTHERAPY AND ONCOLOGY, 2006, 78 (02) : 213 - 216
  • [36] Application of diamond detectors to the dosimetry of 45 and 100 kVp therapy beams: comparison with a parallel-plate ionization chamber and Monte Carlo
    Hugtenburg, RP
    Johnston, K
    Chalmers, GJ
    Beddoe, AH
    PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (09): : 2489 - 2501
  • [37] Characterization of Fricke-gelatin dosimeters for intraoperative Radiation Therapy dosimetry
    Magugliani, Gabriele
    Liosi, Giulia M.
    Andreoli, Stefano
    De Crescenzo, Stefano
    Mossini, Eros
    Macerata, Elena
    Mariani, Mario
    RADIATION PHYSICS AND CHEMISTRY, 2020, 168
  • [38] Dosimetry verification of IntraOperative Radiation Therapy (IORT): a Monte Carlo Study
    Alhamada, H.
    Simon, S.
    Philippson, C.
    Vandekerkhove, C.
    Jourani, Y.
    Pauly, N.
    Reynaert, N.
    Van Gestel, D.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S976 - S976
  • [39] An investigation of the operating characteristics of two PTW diamond detectors in photon and electron beams
    De Angelis, C
    Onori, S
    Pacilio, M
    Cirrone, GAP
    Cuttone, G
    Raffaele, L
    Bucciolini, M
    Mazzocchi, S
    MEDICAL PHYSICS, 2002, 29 (02) : 248 - 254
  • [40] Characterization of a synthetic single crystal diamond Schottky diode for radiotherapy electron beam dosimetry
    Di Venanzio, C.
    Marinelli, Marco
    Milani, E.
    Prestopino, G.
    Verona, C.
    Verona-Rinati, G.
    Falco, M. D.
    Bagala, P.
    Santoni, R.
    Pimpinella, M.
    MEDICAL PHYSICS, 2013, 40 (02)