Comparative dosimetry of diode and diamond detectors in electron beams for intraoperative radiation therapy

被引:0
|
作者
Björk, P [1 ]
Knöös, T [1 ]
Nilsson, P [1 ]
机构
[1] Univ Lund Hosp, Dept Radiat Phys, S-22185 Lund, Sweden
关键词
diamond detector; silicon diode; intraoperative radiation therapy; electron beam dosimetry; dose distribution analysis;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of the present study is to examine the validity of using silicon semiconductor detectors in degraded electron beams with a broad energy spectrum and a wide angular distribution. A comparison is made with diamond detector measurements, which is the dosimeter considered to give the best results provided that dose rate effects are corrected for. Two-dimensional relative absorbed dose distributions in electron beams (6-20 MeV) for intraoperative radiation therapy (IORT) are measured in a water phantom. To quantify deviations between the detectors, a dose comparison tool that simultaneously examines the dose difference and distance to agreement (DTA) is used to evaluate the results in low- and high-dose gradient regions, respectively. Uncertainties of the experimental measurement setup (+/-1% and +/-0.5 mm) are taken into account by calculating a composite distribution that fails this dose-difference and DTA acceptance limit. Thus, the resulting area of disagreement should be related to differences in detector performance. The dose distributions obtained with the diode are generally in very good agreement with diamond detector measurements. The buildup region and the dose falloff region show good agreement with increasing electron energy, while the region outside the radiation field close to the water surface shows an increased difference with energy. The small discrepancies in the composite distributions are due to several factors: (a) variation of the silicon-to-water collision stopping-power ratio with electron energy, (b) a more pronounced directional dependence for diodes than for diamonds, and (c) variation of the electron fluence perturbation correction factor with depth. For all investigated treatment cones and energies, the deviation is within dose-difference and DTA acceptance criteria of +/-3% and +/-1 mm, respectively. Therefore, p-type silicon diodes are well suited, in the sense that they give results in close agreement with diamond detectors, for practical measurements of relative absorbed dose distributions in degraded electron beams used for IORT. (C) 2000 American Association of Physicists in Medicine. [S0094-2405(00)00611-8].
引用
收藏
页码:2580 / 2588
页数:9
相关论文
共 50 条
  • [21] Response of diamond detectors in ultra-high dose-per-pulse electron beams for dosimetry at FLASH radiotherapy
    Kranzer, R.
    Schueller, A.
    Bourgouin, A.
    Hackel, T.
    Poppinga, D.
    Lapp, M.
    Looe, H. K.
    Poppe, B.
    PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (07):
  • [22] Evaluation of Dosimetry Formalisms in Intraoperative Radiation Therapy of Glioblastoma
    Alvarez, David Santiago Ayala
    Watson, Peter G. F.
    Popovic, Marija
    Heng, Veng Jean
    Evans, Michael D. C.
    Panet-Raymond, Valerie
    Seuntjens, Jan
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2023, 117 (03): : 763 - 773
  • [23] In vivo dosimetry using Gafchromic films during pelvic intraoperative electron radiation therapy (IOERT)
    Costa, Filipa
    Sarmento, Sandra
    Gomes, Dora
    Magalhaes, Helena
    Arrais, Rosario
    Moreira, Graciete
    Cruz, Maria Fatima
    Silva, Jose Pedro
    Santos, Lucio
    Sousa, Olga
    BRITISH JOURNAL OF RADIOLOGY, 2016, 89 (1063):
  • [24] In Vivo Dosimetry for Single Fraction Intraoperative Electron Radiation Therapy for Early Stage Breast Cancer
    Ash, R. B.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2017, 99 (02): : E2 - E2
  • [25] Diamond detectors for the monitoring of carbon-ion therapy beams
    Rebisz-Pomorska, M.
    Ciobanu, M.
    Kis, M.
    Pomorski, M.
    Voss, B.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 620 (2-3): : 534 - 539
  • [26] A diamond detector in the dosimetry of high-energy electron and photon beams
    Laub, WU
    Kaulich, TW
    Nüsslin, F
    PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (09): : 2183 - 2192
  • [27] Preliminary assessment of LiF and alanine detectors for the dosimetry of proton therapy beams
    Fattibene, P
    Calicchia, A
    dErrico, F
    DeAngelis, C
    Egger, E
    Onori, S
    RADIATION PROTECTION DOSIMETRY, 1996, 66 (1-4) : 305 - 309
  • [28] SELECTIVE INTERNAL RADIATION-THERAPY - VALIDATION OF INTRAOPERATIVE DOSIMETRY
    BURTON, MA
    GRAY, BN
    KELLEHER, DK
    KLEMP, PF
    RADIOLOGY, 1990, 175 (01) : 253 - 255
  • [29] In vivo dosimetry in cancer patients undergoing intraoperative radiation therapy
    Petoukhova, Anna
    Snijder, Roland
    Vissers, Thomas
    Ceha, Heleen
    Struikmans, Henk
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (18):
  • [30] DOSIMETRY, FIELD SHAPING AND OTHER CONSIDERATIONS FOR INTRAOPERATIVE ELECTRON THERAPY
    BIGGS, PJ
    EPP, ER
    LING, CC
    NOVACK, DH
    MICHAELS, HB
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1981, 7 (07): : 875 - 884