New Non-contact Tonometer HNT-1P Reliability: Comparing Intraocular Pressure, Central Corneal Thickness, and Corrected Intraocular Pressure

被引:2
|
作者
Eah, Kyu Sang [1 ]
Shin, Joong Won [1 ]
Sung, Kyung Rim [1 ]
机构
[1] Univ Ulsan, Coll Med, Dept Ophthalmol Asan, Asan Med Ctr, Seoul, South Korea
来源
关键词
Central corneal thickness; Corrected intraocular pressure; HNT-1P; Intraocular pressure; Tonometry; GOLDMANN APPLANATION TONOMETRY; DYNAMIC CONTOUR TONOMETRY; BIOMECHANICAL PROPERTIES; ULTRASOUND PACHYMETRY; AGREEMENT; PARAMETERS;
D O I
10.3341/jkos.2020.61.5.524
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To assess the reliability of a new non-contact tonometer, HNT-1P, by comparing its intraocular pressure (IOP), central corneal thickness (CCT), and corrected IOP (cIOP) measurements with the IOP measured by Goldmann applanation tonometry (GAT), the CCT measured with ultrasound pachymetry (USP), and the cIOP derived from GAT IOP and USP CCT. Methods: A total of 120 eyes of 65 healthy or glaucoma suspects were included in this study. The IOP was measured with the HNT-1P and GAT. The CCT was determined using the HNT-1P and USP. The IOP measured with GAT was corrected according to the CCT measured by USP. The correlations of the measurements obtained with the various instruments were evaluated using Pearson's correlation coefficient, and the agreement was assessed using Bland-Altman plots. Results: The average IOP measured with the HNT-1P and GAT was 14.8 +/- 5.1 mmHg and 15.6 +/- 5.2 mmHg, respectively, and the correlation coefficient between the two IOP measurements was 0.964 (p < 0.001). The mean CCT was 553.5 +/- 27.4 mu m using the HNT-1P and 550.6 +/- 26.3 mu m using USP, and the correlation coefficient between the two measurements was 0.913 (p < 0.001). The average cIOP was 14.6 +/- 5.0 and 13.4 +/- 6.5 mmHg according to the HNT-1P and GAT, respectively, and the correlation coefficient between the two corrected measurements was 0.956 (p < 0.001). Bland-Altman plots showed a high degree of agreement between the HNT-1P measurements and those of the other devices. Conclusions: The new non-contact HNT-1P tonometer provided reliable IOP, CCT, and cIOP measurements when compared with GAT and USP. HNT-1P could therefore be used clinically for reliable and accurate measurements of the IOP, CCT, and cIOP.
引用
收藏
页码:524 / 531
页数:8
相关论文
共 50 条
  • [41] Central corneal thickness and intraocular pressure in the Cameroonian nonglaucomatous population
    Eballe, Andre Omgbwa
    Koki, Godefroy
    Ellong, Augustin
    Owono, Didier
    Epee, Emilienne
    Bella, Lucienne Assumpta
    Mvogo, Come Ebana
    Kouam, Jeanne Mayouego
    CLINICAL OPHTHALMOLOGY, 2010, 4 : 717 - 724
  • [42] Central corneal thickness and measured intraocular pressure in diabetes.
    Kim, CY
    Kim, HK
    Lim, CS
    Oh, JH
    Hong, YJ
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1999, 40 (04) : S79 - S79
  • [44] Influence of Corneal Thickness on the Intraocular Pressure Readings for Maklakoff's Tonometer of Different Weight
    Franus, D. V.
    EIGHTH POLYAKHOV'S READING, 2018, 1959
  • [45] THE COMPARISON OF INTRAOCULAR PRESSURE MEASUREMENT BY GOLDMANN APPLANATION TONOMETER AND NON-CONTACT AIR PUFF TONOMETER: A CROSS-SECTIONAL RESEARCH
    Arooj, Bushra
    Nadeem, Asmarah
    Ali, Sobia
    INDO AMERICAN JOURNAL OF PHARMACEUTICAL SCIENCES, 2018, 5 (12): : 16875 - 16879
  • [46] Difference in intraocular pressure measurements between non-contact tonometry and Goldmann applanation tonometry and the role of central corneal thickness in affecting glaucoma referrals
    Kamel, Khalid
    Dervan, Edward
    Falzon, Kevin
    O'Brien, Colm
    IRISH JOURNAL OF MEDICAL SCIENCE, 2019, 188 (01) : 321 - 325
  • [47] Difference in intraocular pressure measurements between non-contact tonometry and Goldmann applanation tonometry and the role of central corneal thickness in affecting glaucoma referrals
    Khalid Kamel
    Edward Dervan
    Kevin Falzon
    Colm O’Brien
    Irish Journal of Medical Science (1971 -), 2019, 188 : 321 - 325
  • [48] Changes in Central Corneal Thickness, Corrected Intraocular Pressure, and Iridocorneal Angle in Unilateral Acute Anterior Uveitis
    Pandey, Sumit
    , Ranju Kharel
    Joshi, Sagun Narayan
    NEPALESE JOURNAL OF OPHTHALMOLOGY, 2023, 15 (29) : 53 - 60
  • [49] Effects of central corneal thickness and corneal curvature on the intraocular pressure measurement by Goldmann applanation tonometer and ocular blood flow pneumatonometer
    Saleh, Tarek A.
    Adams, Morag
    McDermott, Bill
    Claridge, Kate G.
    Ewings, Paul
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2006, 34 (06): : 516 - 520
  • [50] Measurement of intraocular pressure using the NT-4000: A new non-contact tonometer equipped with pulse synchronous measurement function
    Yaoeda, K
    Shirakashi, M
    Fukushima, A
    Funaki, S
    Funaki, H
    Ofuchi, N
    Nakatsue, T
    Abe, H
    JOURNAL OF GLAUCOMA, 2005, 14 (03) : 201 - 205