New Non-contact Tonometer HNT-1P Reliability: Comparing Intraocular Pressure, Central Corneal Thickness, and Corrected Intraocular Pressure

被引:2
|
作者
Eah, Kyu Sang [1 ]
Shin, Joong Won [1 ]
Sung, Kyung Rim [1 ]
机构
[1] Univ Ulsan, Coll Med, Dept Ophthalmol Asan, Asan Med Ctr, Seoul, South Korea
来源
关键词
Central corneal thickness; Corrected intraocular pressure; HNT-1P; Intraocular pressure; Tonometry; GOLDMANN APPLANATION TONOMETRY; DYNAMIC CONTOUR TONOMETRY; BIOMECHANICAL PROPERTIES; ULTRASOUND PACHYMETRY; AGREEMENT; PARAMETERS;
D O I
10.3341/jkos.2020.61.5.524
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To assess the reliability of a new non-contact tonometer, HNT-1P, by comparing its intraocular pressure (IOP), central corneal thickness (CCT), and corrected IOP (cIOP) measurements with the IOP measured by Goldmann applanation tonometry (GAT), the CCT measured with ultrasound pachymetry (USP), and the cIOP derived from GAT IOP and USP CCT. Methods: A total of 120 eyes of 65 healthy or glaucoma suspects were included in this study. The IOP was measured with the HNT-1P and GAT. The CCT was determined using the HNT-1P and USP. The IOP measured with GAT was corrected according to the CCT measured by USP. The correlations of the measurements obtained with the various instruments were evaluated using Pearson's correlation coefficient, and the agreement was assessed using Bland-Altman plots. Results: The average IOP measured with the HNT-1P and GAT was 14.8 +/- 5.1 mmHg and 15.6 +/- 5.2 mmHg, respectively, and the correlation coefficient between the two IOP measurements was 0.964 (p < 0.001). The mean CCT was 553.5 +/- 27.4 mu m using the HNT-1P and 550.6 +/- 26.3 mu m using USP, and the correlation coefficient between the two measurements was 0.913 (p < 0.001). The average cIOP was 14.6 +/- 5.0 and 13.4 +/- 6.5 mmHg according to the HNT-1P and GAT, respectively, and the correlation coefficient between the two corrected measurements was 0.956 (p < 0.001). Bland-Altman plots showed a high degree of agreement between the HNT-1P measurements and those of the other devices. Conclusions: The new non-contact HNT-1P tonometer provided reliable IOP, CCT, and cIOP measurements when compared with GAT and USP. HNT-1P could therefore be used clinically for reliable and accurate measurements of the IOP, CCT, and cIOP.
引用
收藏
页码:524 / 531
页数:8
相关论文
共 50 条
  • [21] Central corneal thickness and intraocular pressure in ocular hypertension
    Gutierrez, C
    Casaponsa, JRV
    Munoz, F
    Rebolleda, G
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2001, 239 (06) : 463 - 463
  • [22] Central corneal thickness change with intraocular pressure lowering
    Williams, AJ
    Song, J
    Allingham, RR
    Herndon, L
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2004, 45 : U449 - U449
  • [23] Influence of central corneal thickness on applanation intraocular pressure
    Damji, KF
    Munger, R
    JOURNAL OF GLAUCOMA, 2000, 9 (03) : 205 - 207
  • [24] Central corneal thickness and intraocular pressure in a Mongolian population
    Foster, PJ
    Baasanhu, J
    Alsbirk, PH
    Munkhbayar, D
    Uranchimeg, D
    Johnson, GJ
    OPHTHALMOLOGY, 1998, 105 (06) : 969 - 973
  • [25] Central corneal thickness and intraocular pressure in premature infants
    Sekeroglu, Mehmet Ali
    Hekimoglu, Emre
    Petricli, Ikbal Seza
    Karakaya, Jale
    Ozcan, Beyza
    Yucel, Husniye
    Kavurt, Aysen Sumru
    Bas, Ahmet Yagmur
    INTERNATIONAL OPHTHALMOLOGY, 2015, 35 (06) : 847 - 851
  • [26] Comparison of intraocular pressure measurements obtained by icare pro tonometer, non-contact tonometer and Goldmann applanation tonometer in healthy individuals
    Duran, M.
    JOURNAL FRANCAIS D OPHTALMOLOGIE, 2023, 46 (10): : 1195 - 1203
  • [27] Implications of Central Corneal Thickness-corrected Intraocular Pressure in Normal-tension Glaucoma with Reduced Central Corneal Thickness
    Lee, Sang Muk
    Baek, Sung Uk
    JOURNAL OF THE KOREAN OPHTHALMOLOGICAL SOCIETY, 2023, 64 (09): : 811 - 818
  • [28] Non-contact tonometry: predicting intraocular pressure using a material-corneal thickness-independent methodology
    Redaelli, Elena
    Calvo, Begona
    Rodriguez Matas, Jose Felix
    Luraghi, Giulia
    Grasa, Jorge
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2024, 12
  • [29] Relationship between intraocular pressure, central corneal thickness and corneal topography
    Rinaldi, G
    Baccara, F
    Vattovani, O
    Gioulis, D
    Iacono, P
    Cecchini, P
    Ravalico, G
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2002, 43 : U967 - U967
  • [30] Influence of Corneal Biomechanical Properties on Intraocular Pressure Differences Between a non-contact ORA Tonometer and the Goldmann Applanation Tonometer in primary congenital glacomatous children
    Zareei, Athar
    Razeghinejad, Mohamad Reza
    Salouti, Ramin
    JOURNAL OF RESEARCH IN MEDICAL AND DENTAL SCIENCE, 2018, 6 (03): : 109 - 113