New Non-contact Tonometer HNT-1P Reliability: Comparing Intraocular Pressure, Central Corneal Thickness, and Corrected Intraocular Pressure

被引:2
|
作者
Eah, Kyu Sang [1 ]
Shin, Joong Won [1 ]
Sung, Kyung Rim [1 ]
机构
[1] Univ Ulsan, Coll Med, Dept Ophthalmol Asan, Asan Med Ctr, Seoul, South Korea
来源
关键词
Central corneal thickness; Corrected intraocular pressure; HNT-1P; Intraocular pressure; Tonometry; GOLDMANN APPLANATION TONOMETRY; DYNAMIC CONTOUR TONOMETRY; BIOMECHANICAL PROPERTIES; ULTRASOUND PACHYMETRY; AGREEMENT; PARAMETERS;
D O I
10.3341/jkos.2020.61.5.524
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To assess the reliability of a new non-contact tonometer, HNT-1P, by comparing its intraocular pressure (IOP), central corneal thickness (CCT), and corrected IOP (cIOP) measurements with the IOP measured by Goldmann applanation tonometry (GAT), the CCT measured with ultrasound pachymetry (USP), and the cIOP derived from GAT IOP and USP CCT. Methods: A total of 120 eyes of 65 healthy or glaucoma suspects were included in this study. The IOP was measured with the HNT-1P and GAT. The CCT was determined using the HNT-1P and USP. The IOP measured with GAT was corrected according to the CCT measured by USP. The correlations of the measurements obtained with the various instruments were evaluated using Pearson's correlation coefficient, and the agreement was assessed using Bland-Altman plots. Results: The average IOP measured with the HNT-1P and GAT was 14.8 +/- 5.1 mmHg and 15.6 +/- 5.2 mmHg, respectively, and the correlation coefficient between the two IOP measurements was 0.964 (p < 0.001). The mean CCT was 553.5 +/- 27.4 mu m using the HNT-1P and 550.6 +/- 26.3 mu m using USP, and the correlation coefficient between the two measurements was 0.913 (p < 0.001). The average cIOP was 14.6 +/- 5.0 and 13.4 +/- 6.5 mmHg according to the HNT-1P and GAT, respectively, and the correlation coefficient between the two corrected measurements was 0.956 (p < 0.001). Bland-Altman plots showed a high degree of agreement between the HNT-1P measurements and those of the other devices. Conclusions: The new non-contact HNT-1P tonometer provided reliable IOP, CCT, and cIOP measurements when compared with GAT and USP. HNT-1P could therefore be used clinically for reliable and accurate measurements of the IOP, CCT, and cIOP.
引用
收藏
页码:524 / 531
页数:8
相关论文
共 50 条
  • [31] Effect of central corneal thickness on intraocular pressure and comparison of Topcon CT-80 non-contact tonometry with Goldmann applanation tonometry
    Mansoori, Tarannum
    Balakrishna, Nagalla
    CLINICAL AND EXPERIMENTAL OPTOMETRY, 2018, 101 (02) : 206 - 212
  • [32] The Association of Central Corneal Thickness and Intraocular Pressure Measures by Non-Contact Tonometry and Goldmann Applanation Tonometry among Glaucoma Patients.
    Kyei, Samuel
    Assiamah, Frank
    Kwarteng, Michael Agyemang
    Gboglu, Cynthia Pakyennu
    ETHIOPIAN JOURNAL OF HEALTH SCIENCES, 2020, 30 (06) : 999 - 1004
  • [33] The influence of central corneal thickness and corneal curvature on Intraocular pressure measured by Tono-Pen and rebound tonometer in children
    Sahin, Afsun
    Basmak, Hikmet
    Yidirim, Nigun
    JOURNAL OF GLAUCOMA, 2008, 17 (01) : 57 - 61
  • [34] Comparability of Icare Pro Rebound Tonometer with Goldmann Applanation and Noncontact Tonometer in a Wide Range of Intraocular Pressure and Central Corneal Thickness
    Tamcelik, Nevbahar
    Atalay, Eray
    Cicik, Erdogan
    Ozkok, Ahmet
    OPHTHALMIC RESEARCH, 2015, 54 (01) : 18 - 25
  • [35] Central corneal thickness and its relationship to intraocular pressure in children
    Muir, KW
    Jin, J
    Freedman, SF
    OPHTHALMOLOGY, 2004, 111 (12) : 2220 - 2223
  • [36] Intraocular pressure and central corneal thickness in a healthy student population
    Stamenkovic, Miroslav
    Marjanovic, Ivan
    Maric, Vesna
    Kalezic, Tanja
    Bozic, Marija
    SRPSKI ARHIV ZA CELOKUPNO LEKARSTVO, 2024, 152 (3-4) : 175 - 178
  • [37] Methods of measuring intraocular pressure independently of central corneal thickness
    Hager, A.
    Wiegand, W.
    OPHTHALMOLOGE, 2008, 105 (09): : 840 - 844
  • [38] The intraocular pressure and central corneal thickness in healthy premature infants
    Acar, Damla Erginturk
    Acar, Ugur
    Tunay, Zuhal Ozen
    Ozdemir, Ozdemir
    Dolgun, Anil
    Erdurmus, Mesut
    JOURNAL OF AAPOS, 2015, 19 (02): : 108 - 111
  • [39] Alterations in intraocular pressure and central corneal thickness in premature newborns
    Zengin, Neslihan
    Zengin, Mehmet Ozgur
    Karahan, Eyyup
    Tuncer, Ibrahim
    IZMIR DR BEHCET UZ COCUK HASTANESI DERGISI, 2014, 4 (01): : 20 - 24
  • [40] Non-contact terahertz spectroscopic measurement of the intraocular pressure through corneal hydration mapping
    Chen, Andrew
    Virk, Arjun
    Harris, Zachery
    Abazari, Azin
    Honkanen, Robert
    Arbab, M. Hassan
    BIOMEDICAL OPTICS EXPRESS, 2021, 12 (06) : 3438 - 3449