Scatter Reduction and Correction for Dual-Source Cone-Beam CT Using Prepatient Grids

被引:16
|
作者
Ren, Lei [1 ,2 ]
Chen, Yingxuan [2 ]
Zhang, You [2 ]
Giles, William [1 ]
Jin, Jianyue [3 ]
Yin, Fang-Fang [1 ,2 ]
机构
[1] Duke Univ, Med Ctr, Dept Radiat Oncol, DUMC Box 3295, Durham, NC 27710 USA
[2] Duke Univ, Med Phys Grad Program, DUMC Box 3295, Durham, NC 27710 USA
[3] Georgia Regents Univ, Dept Radiat Oncol, Augusta, GA USA
基金
美国国家卫生研究院;
关键词
dual energy imaging; cross scatter; grid blocking ratio; scatter artifact; total variation regularization; CNR enhancement; simultaneous acquisition; COMPUTED-TOMOGRAPHY CBCT; IMPROVE IMAGE QUALITY; CORRECTION ALGORITHM; RADIATION-THERAPY; GEOMETRY; ARTIFACT; SYSTEM;
D O I
10.1177/1533034615587615
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: Scatter significantly limits the application of the dual-source cone-beam computed tomography by inducing scatter artifacts and degrading contrast-to-noise ratio, Hounsfield-unit accuracy, and image uniformity. Although our previously developed interleaved acquisition mode addressed the cross scatter between the 2 X-ray sources, it doubles the scanning time and doesn't address the forward scatter issue. This study aims to develop a prepatient grid system to address both forward scatter and cross scatter in the dual-source cone-beam computed tomography. Methods: Grids attached to both X-ray sources provide physical scatter reduction during the image acquisition. Image data were measured in the unblocked region, while both forward scatter and cross scatter were measured in the blocked region of the projection for postscan scatter correction. Complementary projections were acquired with grids at complementary locations and were merged to form complete projections for reconstruction. Experiments were conducted with different phantom sizes, grid blocking ratios, image acquisition modes, and reconstruction algorithms to investigate their effects on the scatter reduction and correction. The image quality improvement by the prepatient grids was evaluated both qualitatively through the artifact reduction and quantitatively through contrast-to-noise ratio, Hounsfield-unit accuracy, and uniformity using a CATphan 504 phantom. Results: Scatter artifacts were reduced by scatter reduction and were removed by scatter correction method. Contrast-to-noise ratio, Hounsfield-unit accuracy, and image uniformity were improved substantially. The simultaneous acquisition mode achieved comparable contrast-to-noise ratio as the interleaved and sequential modes after scatter reduction and correction. Higher grid blocking ratio and smaller phantom size led to higher contrast-to-noise ratio for the simultaneous mode. The iterative reconstruction with total variation regularization was more effective than the Feldkamp, Davis, and Kress method in reducing noise caused by the scatter correction to enhance contrast-to-noise ratio. Conclusion: The prepatient grid system is effective in removing the scatter effects in the simultaneous acquisition mode of the dual-source cone-beam computed tomography, which is useful for scanning time reduction or dual energy imaging.
引用
收藏
页码:416 / 427
页数:12
相关论文
共 50 条
  • [31] Scatter correction for a clinical cone-beam CT system using an optimized stationary beam blocker in a single scan
    Liang, Xiaokun
    Jiang, Yangkang
    Zhao, Wei
    Zhang, Zhicheng
    Luo, Chen
    Xiong, Jing
    Yu, Shaode
    Yang, Xiaoming
    Sun, Jihong
    Zhou, Qinxuan
    Niu, Tianye
    Xie, Yaoqin
    MEDICAL PHYSICS, 2019, 46 (07) : 3165 - 3179
  • [32] Scatter Reduction and Correction for Dual-Source CBCT Using the Synchronized Moving Grid (SMOG) System
    Chen, Y.
    Zhang, Y.
    Giles, W.
    Jin, J.
    Yin, F.
    Ren, L.
    MEDICAL PHYSICS, 2014, 41 (06) : 541 - 541
  • [33] Scatter Correction for On-Board Cone-Beam CT in Radiation Therapy Using Planning MDCT Images
    Niu, T.
    Gao, H.
    Star-Lack, J.
    Sun, M.
    Fan, Q.
    Zhu, L.
    MEDICAL PHYSICS, 2010, 37 (06) : 3393 - +
  • [34] Parameterization of Scatter Dependent Artifacts in Cone-Beam CT
    Mail, N.
    Moseley, D.
    Jaffray, D.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [35] Compensators for management of dose and scatter in cone-beam CT
    Graham, S.
    Moseley, D.
    Siewerdsen, J.
    Jaffray, D.
    MEDICAL PHYSICS, 2006, 33 (06) : 1998 - 1999
  • [36] Scatter correction method for cone-beam CT based on interlacing-slit scan
    黄魁东
    张华
    史仪凯
    张亮
    徐哲
    Chinese Physics B, 2014, (09) : 519 - 525
  • [37] Optimization of the Design of a Moving Blocker for Cone-Beam CT Scatter Correction: Experimental Evaluation
    Chen, X.
    Ouyang, L.
    Yan, H.
    Jia, X.
    Zhang, Y.
    Wang, J.
    MEDICAL PHYSICS, 2016, 43 (06) : 3799 - 3799
  • [38] Scatter measurement and correction method for cone-beam CT based on single grating scan
    Huang, Kuidong
    Shi, Wenlong
    Wang, Xinyu
    Dong, Yin
    Chang, Taoqi
    Zhang, Hua
    Zhang, Dinghua
    OPTICAL ENGINEERING, 2017, 56 (06)
  • [39] EXPERIMENTAL SCATTER CORRECTION METHODS IN INDUSTRIAL X-RAY CONE-BEAM CT
    Schoerner, K.
    Goldammer, M.
    Stephan, J.
    REVIEW OF PROGRESS IN QUANTITATIVE NONDESTRUCTIVE EVALUATION, VOLS 30A AND 30B, 2011, 1335 : 485 - 492
  • [40] Planning CT-guided robust and fast cone-beam CT scatter correction using a local filtration technique
    Cui, Hehe
    Jiang, Xiao
    Fang, Chengyijue
    Zhu, Lei
    Yang, Yidong
    MEDICAL PHYSICS, 2021, 48 (11) : 6832 - 6843