Scatter Reduction and Correction for Dual-Source Cone-Beam CT Using Prepatient Grids

被引:16
|
作者
Ren, Lei [1 ,2 ]
Chen, Yingxuan [2 ]
Zhang, You [2 ]
Giles, William [1 ]
Jin, Jianyue [3 ]
Yin, Fang-Fang [1 ,2 ]
机构
[1] Duke Univ, Med Ctr, Dept Radiat Oncol, DUMC Box 3295, Durham, NC 27710 USA
[2] Duke Univ, Med Phys Grad Program, DUMC Box 3295, Durham, NC 27710 USA
[3] Georgia Regents Univ, Dept Radiat Oncol, Augusta, GA USA
基金
美国国家卫生研究院;
关键词
dual energy imaging; cross scatter; grid blocking ratio; scatter artifact; total variation regularization; CNR enhancement; simultaneous acquisition; COMPUTED-TOMOGRAPHY CBCT; IMPROVE IMAGE QUALITY; CORRECTION ALGORITHM; RADIATION-THERAPY; GEOMETRY; ARTIFACT; SYSTEM;
D O I
10.1177/1533034615587615
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: Scatter significantly limits the application of the dual-source cone-beam computed tomography by inducing scatter artifacts and degrading contrast-to-noise ratio, Hounsfield-unit accuracy, and image uniformity. Although our previously developed interleaved acquisition mode addressed the cross scatter between the 2 X-ray sources, it doubles the scanning time and doesn't address the forward scatter issue. This study aims to develop a prepatient grid system to address both forward scatter and cross scatter in the dual-source cone-beam computed tomography. Methods: Grids attached to both X-ray sources provide physical scatter reduction during the image acquisition. Image data were measured in the unblocked region, while both forward scatter and cross scatter were measured in the blocked region of the projection for postscan scatter correction. Complementary projections were acquired with grids at complementary locations and were merged to form complete projections for reconstruction. Experiments were conducted with different phantom sizes, grid blocking ratios, image acquisition modes, and reconstruction algorithms to investigate their effects on the scatter reduction and correction. The image quality improvement by the prepatient grids was evaluated both qualitatively through the artifact reduction and quantitatively through contrast-to-noise ratio, Hounsfield-unit accuracy, and uniformity using a CATphan 504 phantom. Results: Scatter artifacts were reduced by scatter reduction and were removed by scatter correction method. Contrast-to-noise ratio, Hounsfield-unit accuracy, and image uniformity were improved substantially. The simultaneous acquisition mode achieved comparable contrast-to-noise ratio as the interleaved and sequential modes after scatter reduction and correction. Higher grid blocking ratio and smaller phantom size led to higher contrast-to-noise ratio for the simultaneous mode. The iterative reconstruction with total variation regularization was more effective than the Feldkamp, Davis, and Kress method in reducing noise caused by the scatter correction to enhance contrast-to-noise ratio. Conclusion: The prepatient grid system is effective in removing the scatter effects in the simultaneous acquisition mode of the dual-source cone-beam computed tomography, which is useful for scanning time reduction or dual energy imaging.
引用
收藏
页码:416 / 427
页数:12
相关论文
共 50 条
  • [21] Scatter correction of cone-beam CT using a deep residual convolution neural network (DRCNN)
    Jiang, Yangkang
    Yang, Chunlin
    Yang, Pengfei
    Hu, Xi
    Luo, Chen
    Xue, Yi
    Xu, Lei
    Hu, Xiuhua
    Zhang, Luhan
    Wang, Jing
    Sheng, Ke
    Niu, Tianye
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (14):
  • [22] Robust scatter correction method for cone-beam CT using an interlacing-slit plate
    Huang, Kui-Dong
    Xu, Zhe
    Zhang, Ding-Hua
    Zhang, Hua
    Shi, Wen-Long
    CHINESE PHYSICS C, 2016, 40 (06) : 99 - 106
  • [23] Robust scatter correction method for cone-beam CT using an interlacing-slit plate
    黄魁东
    徐哲
    张定华
    张华
    时稳龙
    Chinese Physics C, 2016, 40 (06) : 99 - 106
  • [24] Improved Primary Modulation Method for Scatter Correction in Cone-Beam CT Using Local Filtration
    Zhu, L.
    MEDICAL PHYSICS, 2017, 44 (06) : 3283 - 3284
  • [25] X-ray scatter correction for cone-beam CT using moving blocker array
    Zhu, L
    Strobel, N
    Fahrig, R
    Medical Imaging 2005: Physics of Medical Imaging, Pts 1 and 2, 2005, 5745 : 251 - 258
  • [26] HELICAL DUAL SOURCE CONE-BEAM MICRO-CT
    Johnston, Samuel M.
    Johnson, G. Allan
    Badea, Cristian T.
    2014 IEEE 11th International Symposium on Biomedical Imaging (ISBI), 2014, : 177 - 180
  • [27] SCATTER REDUCTION IN CONE-BEAM CT USING A SCANNING CYLINDRICAL APERTURE TOMOGRAPHIC (SCAT) DEVICE
    RUDIN, S
    BEDNAREK, DR
    MEDICAL PHYSICS, 1980, 7 (04) : 428 - 428
  • [28] Single-scan scatter correction for cone-beam CT using a stationary beam blocker: a preliminary study
    Niu, Tianye
    Zhu, Lei
    MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING, 2011, 7961
  • [29] Scatter Correction for Clinical Cone-Beam CT System Using An Optimized Stationary Beam Blocker in a Single Scan
    Liang, X.
    Luo, C.
    Zhang, Z.
    Xie, Y.
    Niu, T.
    MEDICAL PHYSICS, 2018, 45 (06) : E170 - E170
  • [30] Scatter Correction for Clinical Cone-Beam CT System Using An Optimized Stationary Beam Blocker with a Single Scan
    Liang, X.
    Gong, S.
    Zhou, Q.
    Zhang, Z.
    Xie, Y.
    Niu, T.
    MEDICAL PHYSICS, 2016, 43 (06) : 3457 - 3457