Predictors of patient radiation exposure during transcatheter aortic valve replacement

被引:6
|
作者
Goldsweig, Andrew M. [1 ,2 ,3 ]
Kennedy, Kevin F. [4 ]
Kolte, Dhaval [2 ,3 ]
Abbott, J. Dawn [2 ,3 ]
Gordon, Paul C. [2 ,3 ]
Sharaf, Barry L. [2 ,3 ]
Sellke, Frank W. [2 ,3 ]
Ehsan, Afshin [2 ,3 ]
Sodha, Neel R. [2 ,3 ]
Rutar, Frank [5 ]
Aronow, Herbert D. [2 ,3 ]
机构
[1] Univ Nebraska Med Ctr, Div Cardiol, Omaha, NE USA
[2] Brown Univ, Warren Alpert Med Sch, Div Cardiol, Providence, RI 02912 USA
[3] Lifespan Cardiovasc Inst, Providence, RI USA
[4] St Lukes Mid Amer Heart Inst, Midwest Bioinformat, Kansas City, MO USA
[5] Univ Nebraska Med Ctr, Dept Radiat Safety, Omaha, NE USA
关键词
angiographic; fluoroscopic imaging; radiation physics; dosimetry; transcatheter valve implantation; INTERMEDIATE-RISK PATIENTS; CORONARY-ANGIOGRAPHY; INTERVENTION; TIME;
D O I
10.1002/ccd.27452
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Transcatheter aortic valve replacement (TAVR) exposes patients to radiation. ObjectivesMethodsWe sought to identify factors associated with higher radiation exposure and to quantify their relative influence, which may inform reduction of this hazard. All TAVR procedures at Rhode Island Hospital between March 20, 2012 and February 12, 2017 were included. Procedures were performed by two co-primary operators using a Siemens Artis Zeego system. Radiation metrics were generated by the imaging system. The primary metric was dose-area product (DAP, Gy*cm(2)), and secondary metrics were reference point air kerma (mGy) and fluoroscopy time (minutes). Data collected for the STS/ACC TVT Registry were utilized to develop a multivariable linear regression model predicting DAP. ResultsConclusionIn 294 TAVRs, median DAP was 169 Gy*cm(2) [interquartile range (IQR) 106-238]. The r(2) values for the full 27-variable DAP model and reduced eight-variable model were 0.457 and 0.420, respectively. Valve area, aortic insufficiency, and procedure year (suggesting absence of a learning curve) were non-significant predictors in the full model, while increasing weight, cutdown transfemoral access, higher pre-procedure creatinine and hemoglobin, and vascular complications predicted higher DAP in both models. Results were unchanged when DAP was log-transformed. Secondary models for air kerma and fluoroscopy time revealed similar predictors. Factors associated with increased procedural complexity and duration as well as radiation attenuation and scatter predict increased patient radiation exposure during TAVR. Modification of procedural technique, especially using percutaneous femoral vascular access, may facilitate reduction in exposure.
引用
收藏
页码:768 / 774
页数:7
相关论文
共 50 条
  • [1] Patient radiation exposure during transcatheter aortic valve replacement procedures
    Daneault, Benoit
    Balter, Stephen
    Kodali, Susheel K.
    Williams, Mathew R.
    Genereux, Philippe
    Reiss, George R.
    Paradis, Jean-Michel
    Green, Philip
    Kirtane, Ajay J.
    Smith, Craig
    Moses, Jeffrey W.
    Leon, Martin B.
    EUROINTERVENTION, 2012, 8 (06) : 679 - 684
  • [2] PROVIDER RADIATION EXPOSURE DURING TRANSCATHETER AORTIC VALVE REPLACEMENT
    Vishnevsky, Alec
    Savage, Rady Ho Michael
    Fischman, David
    Mehrotra, Praveen
    Entwhistle, John C., III
    Goldhammer, Jordan
    Marcantuono, Rebecca
    Moss, Lisa
    Keklak, John
    Davis, Marguerite
    Schmitz, Theresa
    Walinsky, Paul
    Morris, Rohinton
    Ruggiero, Nicholas
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2017, 69 (11) : 1030 - 1030
  • [3] Operator Radiation Exposure During Transfemoral Transcatheter Aortic Valve Replacement
    Goel, Sunny
    Casazza, Richard
    Pasam, Ravi Teja
    Montagna, Enrico
    Gotesman, Joseph
    Frankel, Robert
    Borgen, Elliot
    Crooke, Gregory
    Saunders, Paul
    Shani, Jacob
    STRUCTURAL HEART-THE JOURNAL OF THE HEART TEAM, 2022, 6 (01):
  • [4] Minimizing Radiation Exposure during Transcatheter Aortic Valve Replacement (TAVR) Procedures
    Ramsewak, Adesh
    Sharma, Divyesh
    O'Conaire, Sean
    Verghis, R.
    Manoharan, Ganesh
    Spence, Mark
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2012, 60 (17) : B265 - B265
  • [5] Transcatheter Aortic Valve Replacement Operator Radiation Exposure (TAVORE)
    Casazza, Richard
    Goel, Sunny
    Shani, Jacob
    JACC-CARDIOVASCULAR INTERVENTIONS, 2020, 13 (04) : S55 - S55
  • [6] Cerebral Microembolic Exposure during Transcatheter Aortic Valve Replacement
    Tom, Stephanie
    Trivedi, Jaimin
    Flaherty, Michael
    Grubb, Kendra
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2017, 70 (18) : B268 - B268
  • [7] Radiation Exposure During Transcatheter Aortic Valve Replacement: Impact of Arterial Approach and Prosthesis Type
    Faroux, Laurent
    Villecourt, Aurelien
    Guimaraes, Leonardo
    Wintzer-Wehekind, Jerome
    Junquera, Lucia
    Arsenault, Jean
    Blanpain, Thierry
    Tassan-Mangina, Sophie
    Heroguelle, Virginie
    Ruggieri, Vito Giovanni
    Metz, Damien
    Kalavrouziotis, Dimitri
    Dumont, Eric
    Paradis, Jean-Michel
    Delarochelliere, Robert
    Del Val, David
    Muntane-Carol, Guillem
    Mohammadi, Siamak
    Rodes-Cabau, Josep
    ANNALS OF THORACIC SURGERY, 2021, 111 (05): : 1601 - 1606
  • [8] Patient Selection for Transcatheter Aortic Valve Replacement
    不详
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2013, 62 (17) : S1 - S10
  • [9] HANDS DOWN: OPERATOR HAND VERSUS TORSO RADIATION EXPOSURE DURING TRANSCATHETER AORTIC VALVE REPLACEMENT
    Vishnevsky, Alec
    Ho, Rady
    Savage, Michael
    Fischman, David
    Mehrotra, Praveen
    Entwhistle, John C., III
    Goldhammer, Jordan
    Marcantuono, Rebecca
    Moss, Lisa
    Keklak, John
    Davis, Marguerite
    Schmitz, Theresa
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2017, 69 (11) : 1029 - 1029
  • [10] Predictors of Contrast Volume in Transcatheter Aortic Valve Replacement
    Goldsweig, Andrew M.
    Lyden, Elizabeth
    Aronow, Herbert D.
    Kolte, Dhaval
    Pavlides, Gregory
    Barton, David
    Chatzizisis, Yiannis
    Gumina, Richard J.
    Abbott, J. Dawn
    CARDIOLOGY, 2020, 145 (09) : 608 - 610