Comparison of Flow Estimators for Rotary Blood Pumps: An In Vitro and In Vivo Study

被引:14
|
作者
Petrou, Anastasios [1 ]
Kuster, Daniel [1 ]
Lee, Jongseok [2 ]
Meboldt, Mirko [1 ]
Daners, Marianne Schmid [1 ]
机构
[1] Swiss Fed Inst Technol, Prod Dev Grp Zurich, Dept Mech & Proc Engn, CH-8092 Zurich, Switzerland
[2] German Aerosp Ctr DLR, Inst Robot & Mechatron, D-82234 Wessling, Germany
关键词
Ventricular assist device; Numerical model of cardiovascular system; Physiological control; Monitoring; Extended Kalman filter; Pressure sensor; VENTRICULAR ASSIST DEVICES; PRESSURE-HEAD; DIFFERENTIAL PRESSURE; SYSTEM; POWER;
D O I
10.1007/s10439-018-2106-7
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Various approaches for estimating the flow rate of a rotary blood pump have been proposed for monitoring and control purposes. They have been evaluated under different test conditions and, therefore, a direct comparisonamong them is difficult. Furthermore, a limited performance has been reported for the areas where the pump flow and motor current present a non-monotonic relationship. In this regard, we selected most approaches that have been presented in literature and added a modified one, resulting in four estimators, which are either non-invasive or invasive, i.e., inlet and outlet pump pressure sensors are used. Data from in vitro and in vivo studies with the Deltastream pump DP2 were used to compare the estimators under the same test conditions. These data included both constant and varying pre- and afterload, contractility, viscosity, as well as pump speed settings. Bland-Altman plots were used to evaluate the performance of the estimators. The mean error of the overall estimated flow in vitro ranged from 0.002 to 0.38L/min and the limits of agreement (LoA) between +/- 2L/min. During negative flows the mean error decreased by about 25% when the pump inlet pressure was addedas an input. In vivo, the mean errors increased, while the LoA remained in the same range. An estimator based on pump pressure difference improves the reliability in areas where flow and current relationship is not monotonic. A trade-off between estimation accuracy and number of sensors was identified. The estimation objective and the potential errors should be considered when selecting an estimation approach and designing the pump systems.
引用
收藏
页码:2123 / 2134
页数:12
相关论文
共 50 条
  • [21] A Mathematical Model for Estimating Physiological Parameters of Blood Flow through Rotary Blood Pumps
    Telyshev, D., V
    BIOMEDICAL ENGINEERING-MEDITSINSKAYA TEKNIKA, 2020, 54 (03): : 163 - 168
  • [22] A Mathematical Model for Estimating Physiological Parameters of Blood Flow through Rotary Blood Pumps
    D. V. Telyshev
    Biomedical Engineering, 2020, 54 : 163 - 168
  • [23] Hemodynamic Benefits of Counterpulsation, Implantable, Percutaneous, and Intraaortic Rotary Blood Pumps: An In-Silico and In Vitro Study
    Wang, Yu
    Koenig, Steven C.
    Sobieski, Michael A.
    Slaughter, Mark S.
    Giridharan, Guruprasad A.
    CARDIOVASCULAR ENGINEERING AND TECHNOLOGY, 2017, 8 (04) : 439 - 452
  • [24] Hemodynamic Benefits of Counterpulsation, Implantable, Percutaneous, and Intraaortic Rotary Blood Pumps: An In-Silico and In Vitro Study
    Yu Wang
    Steven C. Koenig
    Michael A. Sobieski
    Mark S. Slaughter
    Guruprasad A. Giridharan
    Cardiovascular Engineering and Technology, 2017, 8 : 439 - 452
  • [25] Measurement for implantable rotary blood pumps
    Bertram, CD
    PHYSIOLOGICAL MEASUREMENT, 2005, 26 (04) : R99 - R117
  • [26] Condition monitoring of rotary blood pumps
    Jammu, VB
    Malanoski, S
    Walter, T
    Smith, W
    ASAIO JOURNAL, 1997, 43 (05) : M639 - M643
  • [27] Control strategy for rotary blood pumps
    Ohuchi, K
    Kikugawa, D
    Takahashi, K
    Uemura, M
    Nakamura, M
    Murakami, T
    Sakamoto, T
    Takatani, S
    ARTIFICIAL ORGANS, 2001, 25 (05) : 366 - 370
  • [28] In vitro analysis of the flow fields of a rotary blood pump
    Morsi, Y
    Yang, W
    Chan, S
    COMPUTATIONAL MECHANICS, VOLS 1 AND 2, PROCEEDINGS: NEW FRONTIERS FOR THE NEW MILLENNIUM, 2001, : 875 - 880
  • [29] Rotary blood pumps: A new horizon
    Olsen, DB
    ARTIFICIAL ORGANS, 1999, 23 (08) : 695 - 696
  • [30] Purge system for rotary blood pumps
    Sipin, AJ
    Bender, B
    Fabrey, WJ
    Keller, R
    Liu, J
    Olsen, DB
    ARTIFICIAL ORGANS, 1997, 21 (07) : 611 - 619