Hemodynamic Benefits of Counterpulsation, Implantable, Percutaneous, and Intraaortic Rotary Blood Pumps: An In-Silico and In Vitro Study

被引:0
|
作者
Yu Wang
Steven C. Koenig
Michael A. Sobieski
Mark S. Slaughter
Guruprasad A. Giridharan
机构
[1] Dalian University of Technology,Department of Biomedical Engineering, Faculty of Electronic Information and Electrical Engineering
[2] University of Louisville,Department of Bioengineering, Cardiovascular Innovation Institute
[3] University of Louisville,Department of Cardiovascular and Thoracic Surgery, Cardiovascular Innovation Institute
[4] University of Louisville,Department of Bioengineering, Cardiovascular Innovation Institute
关键词
Heart failure; Mechanical circulatory support devices; Cardiac and vascular hemodynamic responses; and ; modeling;
D O I
暂无
中图分类号
学科分类号
摘要
Mechanical circulatory support (MCS) devices have become a standard therapy for heart failure (HF) patients. MCS device designs may differ by level of support, inflow and/or outflow cannulation sites, and mechanism(s) of cardiac unloading and blood flow delivery. Investigation and direct comparison of hemodynamic parameters that help characterize performance of MCS devices has been limited. We quantified cardiac and vascular hemodynamic responses for different types of MCS devices. Continuous flow (CF) left ventricular (LV) assist devices (LVAD) with LV or left atrial (LA) inlet, counterpulsation devices, percutaneous CF LVAD, and intra-aortic rotary blood pumps (IARBP) were quantified using established computer simulation and mock flow loop models. Hemodynamic data were analyzed on a beat-to-beat basis at baseline HF and over a range of MCS support. Results demonstrated that all LVAD greatly diminished vascular pulsatility (P) and LV external work (LVEW). LVAD with LA inflow provided a greater reduction in LVEW compared to LVAD with LV inflow, but at the potential risk for blood stasis/thrombosis in the LV at high support. Counterpulsation provided greater coronary flow (CoF) augmentation, but had a lower reduction in LVEW compared to partial percutaneous LVAD support. IARBP diminished LVEW, but at the expense of diminished CoF due to coronary steal. The hemodynamic benefits for each type of mechanical circulatory support system are unique and clinical decisions on device selection to maximize end organ perfusion and minimize invasiveness needs to be considered for an individual patients’ presentation.
引用
收藏
页码:439 / 452
页数:13
相关论文
共 11 条
  • [1] Hemodynamic Benefits of Counterpulsation, Implantable, Percutaneous, and Intraaortic Rotary Blood Pumps: An In-Silico and In Vitro Study
    Wang, Yu
    Koenig, Steven C.
    Sobieski, Michael A.
    Slaughter, Mark S.
    Giridharan, Guruprasad A.
    CARDIOVASCULAR ENGINEERING AND TECHNOLOGY, 2017, 8 (04) : 439 - 452
  • [2] Physiologic control of rotary blood pumps:: An in vitro study
    Giridharan, GA
    Pantalos, GM
    Gillars, KJ
    Koenig, SC
    Skliar, M
    ASAIO JOURNAL, 2004, 50 (05) : 403 - 409
  • [3] Comparison of Flow Estimators for Rotary Blood Pumps: An In Vitro and In Vivo Study
    Anastasios Petrou
    Daniel Kuster
    Jongseok Lee
    Mirko Meboldt
    Marianne Schmid Daners
    Annals of Biomedical Engineering, 2018, 46 : 2123 - 2134
  • [4] Comparison of Flow Estimators for Rotary Blood Pumps: An In Vitro and In Vivo Study
    Petrou, Anastasios
    Kuster, Daniel
    Lee, Jongseok
    Meboldt, Mirko
    Daners, Marianne Schmid
    ANNALS OF BIOMEDICAL ENGINEERING, 2018, 46 (12) : 2123 - 2134
  • [5] Standardized Comparison of Selected Physiological Controllers for Rotary Blood Pumps: In Vitro Study
    Petrou, Anastasios
    Lee, Jongseok
    Dual, Seraina
    Ochsner, Gregor
    Meboldt, Mirko
    Daners, Marianne Schmid
    ARTIFICIAL ORGANS, 2018, 42 (03) : E29 - E42
  • [6] Correction to: Comparison of Flow Estimators for Rotary Blood Pumps: An In Vitro and In Vivo Study
    Anastasios Petrou
    Daniel Kuster
    Jongseok Lee
    Mirko Meboldt
    Marianne Schmid Daners
    Annals of Biomedical Engineering, 2019, 47 : 1827 - 1827
  • [7] Estimation of pump flow rate and abnormal condition of implantable rotary blood pumps during long-term in vivo study
    Nakata, K
    Yosikawa, M
    Takano, T
    Sankai, Y
    Ohtsuka, G
    Glueck, J
    Fujisawa, A
    Makinouchi, K
    Yokokawa, M
    Nosé, Y
    ARTIFICIAL ORGANS, 2000, 24 (04) : 315 - 319
  • [8] Comparison of Flow Estimators for Rotary Blood Pumps: An In Vitro and In Vivo Study (vol 46, pg 2123, 2018)
    Petrou, Anastasios
    Kuster, Daniel
    Lee, Jongseok
    Meboldt, Mirko
    Daners, Marianne Schmid
    ANNALS OF BIOMEDICAL ENGINEERING, 2019, 47 (08) : 1827 - 1827
  • [9] Identifying physiologically significant pumping state transitions in implantable rotary blood pumps used as left ventricular assist devices:: an in-vivo study
    Ayre, PJ
    Lovell, NH
    Morris, RW
    Wilson, ML
    Woodard, JC
    PROCEEDINGS OF THE 23RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4: BUILDING NEW BRIDGES AT THE FRONTIERS OF ENGINEERING AND MEDICINE, 2001, 23 : 445 - 448
  • [10] AN ARTIFICIAL NEURAL-NETWORK-BASED NONINVASIVE DETECTOR FOR SUCTION AND LEFT ATRIUM PRESSURE IN THE CONTROL OF ROTARY BLOOD PUMPS - AN IN-VITRO STUDY
    STOCKLMAYER, C
    DORFFNER, G
    SCHMIDT, C
    SCHIMA, H
    ARTIFICIAL ORGANS, 1995, 19 (07) : 719 - 724