Comparison of Flow Estimators for Rotary Blood Pumps: An In Vitro and In Vivo Study

被引:14
|
作者
Petrou, Anastasios [1 ]
Kuster, Daniel [1 ]
Lee, Jongseok [2 ]
Meboldt, Mirko [1 ]
Daners, Marianne Schmid [1 ]
机构
[1] Swiss Fed Inst Technol, Prod Dev Grp Zurich, Dept Mech & Proc Engn, CH-8092 Zurich, Switzerland
[2] German Aerosp Ctr DLR, Inst Robot & Mechatron, D-82234 Wessling, Germany
关键词
Ventricular assist device; Numerical model of cardiovascular system; Physiological control; Monitoring; Extended Kalman filter; Pressure sensor; VENTRICULAR ASSIST DEVICES; PRESSURE-HEAD; DIFFERENTIAL PRESSURE; SYSTEM; POWER;
D O I
10.1007/s10439-018-2106-7
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Various approaches for estimating the flow rate of a rotary blood pump have been proposed for monitoring and control purposes. They have been evaluated under different test conditions and, therefore, a direct comparisonamong them is difficult. Furthermore, a limited performance has been reported for the areas where the pump flow and motor current present a non-monotonic relationship. In this regard, we selected most approaches that have been presented in literature and added a modified one, resulting in four estimators, which are either non-invasive or invasive, i.e., inlet and outlet pump pressure sensors are used. Data from in vitro and in vivo studies with the Deltastream pump DP2 were used to compare the estimators under the same test conditions. These data included both constant and varying pre- and afterload, contractility, viscosity, as well as pump speed settings. Bland-Altman plots were used to evaluate the performance of the estimators. The mean error of the overall estimated flow in vitro ranged from 0.002 to 0.38L/min and the limits of agreement (LoA) between +/- 2L/min. During negative flows the mean error decreased by about 25% when the pump inlet pressure was addedas an input. In vivo, the mean errors increased, while the LoA remained in the same range. An estimator based on pump pressure difference improves the reliability in areas where flow and current relationship is not monotonic. A trade-off between estimation accuracy and number of sensors was identified. The estimation objective and the potential errors should be considered when selecting an estimation approach and designing the pump systems.
引用
收藏
页码:2123 / 2134
页数:12
相关论文
共 50 条
  • [41] Inlet and outlet devices for rotary blood pumps
    Song, XW
    Wood, HG
    Allaire, PE
    Antaki, JF
    Olsen, DB
    ARTIFICIAL ORGANS, 2004, 28 (10) : 911 - 915
  • [42] Hydraulic Characterization of Implantable Rotary Blood Pumps
    Boes, Stefan
    Thamsen, Bente
    Haas, Mattia
    Daners, Marianne Schmid
    Meboldt, Mirko
    Granegger, Marcus
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2019, 66 (06) : 1618 - 1627
  • [43] Hemolysis testing is not the adequate method to verify optimization of the flow profile of rotary blood pumps
    Mueller, JH
    Brodde, M
    Nuesser, P
    Mueller, J
    Graichen, K
    Brandes, K
    Kehrel, B
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2006, 29 (05): : 504 - 504
  • [44] Fluid Dynamics in Rotary Piston Blood Pumps
    Johannes Wappenschmidt
    Simon J. Sonntag
    Martin Buesen
    Sascha Gross-Hardt
    Tim Kaufmann
    Thomas Schmitz-Rode
    Ruediger Autschbach
    Andreas Goetzenich
    Annals of Biomedical Engineering, 2017, 45 : 554 - 566
  • [45] Accelerated Hemocompatibility Testing of Rotary Blood Pumps
    Mcnamee, Antony P.
    Griffith, Tia A.
    Smith, Amanda G.
    Kuck, Lennart
    Simmonds, Michael J.
    ASAIO JOURNAL, 2023, 69 (10) : 918 - 923
  • [46] Indications for rotary blood pumps in clinical practice
    Meyns, B
    ARTIFICIAL ORGANS, 2001, 25 (05) : 323 - 326
  • [47] Neurological Complications with Pulsatile and Rotary Blood Pumps
    Mussivand, T.
    Holmes, K. S.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2009, 28 (02): : S278 - S278
  • [48] COMPARATIVE HEMOLYSIS TESTS OF ROTARY BLOOD PUMPS
    ITKIN, GP
    MATVEEV, YG
    ROMANOV, O
    ARTIFICIAL ORGANS, 1995, 19 (07) : 616 - 619
  • [49] Rotary blood pumps for cardiac assistance: A "must"?
    Mesana, TG
    ARTIFICIAL ORGANS, 2004, 28 (02) : 218 - 225
  • [50] Development of a Pump Flow Estimator for Rotary Blood Pumps to Enhance Monitoring of Ventricular Function
    Granegger, Marcus
    Moscato, Francesco
    Casas, Fernando
    Wieselthaler, Georg
    Schima, Heinrich
    ARTIFICIAL ORGANS, 2012, 36 (08) : 691 - 699