Method for simultaneous analysis of eight vitamin E isomers in various foods by high performance liquid chromatography and fluorescence detection

被引:83
|
作者
Pinheiro-Sant'Ana, Helena Maria [1 ]
Guinazi, Michele [1 ]
Oliveira, Daniela da Silva [2 ]
Della Lucia, Ceres Mattos [1 ]
Reis, Barbara de Lazzari [1 ]
Cardoso Brandao, Sebastiao Cesar [2 ]
机构
[1] Univ Fed Vicosa, Dept Nutr & Hlth, Lab Vitamins Anal, BR-36561000 Vicosa, MG, Brazil
[2] Univ Fed Vicosa, Dept Food Technol, BR-36561000 Vicosa, MG, Brazil
关键词
Tocopherols; Tocotrienols; Vegetables; Vegetable oils; Eggs; Seeds; ALPHA-TOCOPHEROL; TOCOTRIENOLS; ANTIOXIDANTS; VEGETABLES; FORMULAS; COLUMNS;
D O I
10.1016/j.chroma.2011.09.067
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The objective of this study was to optimize a method to investigate the occurrence and to quantify the full isomeric composition of vitamin E (alpha-, beta-, gamma- and delta-tocopherols and tocotrienols) in 6 vegetables (raw and cooked). 3 herbs/spices, raw and cooked eggs, vegetable oils (canola, olive and soybean), flaxseed and sorghum (flour and seeds) and soy (flour) by HPLC with fluorescence detection. Different conditions of extraction and analysis were tested. The optimized method consisted of direct extraction with solvent (hexane:ethyl acetate, 85:15, v/v). For analysis normal phase column was used with mobile phase consisting of hexane:isopropanol:acetic acid (98.9:0.6:0.5) with isocratic elution and fluorescence detection. Excellent separation of all isomers was obtained along with adequate quantification in the foods analyzed. Recovery rates of standards ranged from 91.3 to 99.4%. The linearity range for each isomer varied from 2.5 to 137.5 ng/mL (R(2) greater than 0.995 in all cases). Detection limits ranged from 21.0 to 48.0 ng/mL for tocopherols and from 56.0 to 67.0 ng/mL for tocotrienols, while quantification limits ranged from 105.0 to 240.0 ng/mL for tocopherols and from 280.0 to 335.0 ng/mL for tocotrienols. The optimized method was considered simple, fast and reliable, and also preserved vitamin E isomers when compared to validated methods involving saponification. (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:8496 / 8502
页数:7
相关论文
共 50 条
  • [1] Simultaneous determination of eight vitamin E isomers and vitamin A in foods by normal-phase high performance liquid chromatography
    Zheng Yibin
    Huang Baifen
    Ren Yiping
    CHINESE JOURNAL OF CHROMATOGRAPHY, 2016, 34 (07) : 692 - 696
  • [2] A Novel Ionic Liquid-Based Liquid-Liquid Microextraction Combined with High Performance Liquid Chromatography for Simultaneous Determination of Eight Vitamin E Isomers in Human Serum
    Yin, Shuo
    Yang, Yi
    Zhang, Jing
    Li, Yongxin
    Wu, Ling
    Sun, Chengjun
    JOURNAL OF AOAC INTERNATIONAL, 2020, 103 (04) : 989 - 996
  • [3] Simultaneous analysis of eight vitamin E isomers in Moringa oleifera Lam. leaves by ultra performance convergence chromatography
    Qi, Ningli
    Gong, Xiao
    Feng, Cuiping
    Wang, Xiaoxi
    Xu, Yongwei
    Lin, Lijing
    FOOD CHEMISTRY, 2016, 207 : 157 - 161
  • [4] Determination of vitamin E in aquatic organisms by high-performance liquid chromatography with fluorescence detection
    Huo, JZ
    Nelis, HJ
    Lavens, P
    Sorgeloos, P
    DeLeenheer, AP
    ANALYTICAL BIOCHEMISTRY, 1996, 242 (01) : 123 - 128
  • [5] An improved method for simultaneous analysis of aminothiols in human plasma by high-performance liquid chromatography with fluorescence detection
    Cevasco, Giorgio
    Piatek, Anna Maria
    Scapolla, Carlo
    Thea, Sergio
    JOURNAL OF CHROMATOGRAPHY A, 2010, 1217 (14) : 2158 - 2162
  • [6] Simultaneous determination of Vitamin E homologs in chicken meat by liquid chromatography with fluorescence detection
    Hewavitharana, AK
    Lanari, MC
    Becu, C
    JOURNAL OF CHROMATOGRAPHY A, 2004, 1025 (02) : 313 - 317
  • [7] Simultaneous determination of eight vitamin E isomers and α-tocopherol acetate in functional foods and nutritional supplements by gas chromatography - mass spectrometry
    Lu, Dan
    Yang, Yi
    Wu, Xin
    Zeng, Li
    Li, Yongxin
    Sun, Chengjun
    ANALYTICAL METHODS, 2015, 7 (08) : 3353 - 3362
  • [8] SIMULTANEOUS DETERMINATION OF SERUM VITAMIN-A AND VITAMIN-E BY LIQUID-CHROMATOGRAPHY WITH FLUORESCENCE DETECTION
    WILLIAMS, ATR
    JOURNAL OF CHROMATOGRAPHY, 1985, 341 (01): : 198 - 201
  • [9] Determination of Vitamin E Isomers of Grape Seeds By High-Performance Liquid Chromatography-UV Detection
    Seker, Mehmet E.
    Celik, Ali
    Dost, Kenan
    JOURNAL OF CHROMATOGRAPHIC SCIENCE, 2012, 50 (02) : 97 - 101
  • [10] SEPARATION OF VITAMIN-E ISOMERS BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY
    TANGNEY, CC
    DRISKELL, JA
    MCNAIR, HM
    JOURNAL OF CHROMATOGRAPHY, 1979, 172 (APR): : 513 - 515