Comparison of forward- and back-projection in vivo EPID dosimetry for VMAT treatment of the prostate

被引:29
|
作者
Bedford, James L. [1 ]
Hanson, Ian M.
Hansen, Vibeke N.
机构
[1] Inst Canc Res, Joint Dept Phys, Sutton SM2 5PT, Surrey, England
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2018年 / 63卷 / 02期
关键词
portal dosimetry; EPID; VMAT; treatment plan; MODULATED ARC THERAPY; DOSE VERIFICATION; PORTAL DOSIMETRY; STATISTICS NOTES; RADIOTHERAPY; IMRT; IMPLEMENTATION; OPTIMIZATION; IMAGES; MODE;
D O I
10.1088/1361-6560/aa9c60
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In the forward-projection method of portal dosimetry for volumetric modulated arc therapy (VMAT), the integrated signal at the electronic portal imaging device (EPID) is predicted at the time of treatment planning, against which the measured integrated image is compared. In the back-projection method, the measured signal at each gantry angle is back-projected through the patient CT scan to give a measure of total dose to the patient. This study aims to investigate the practical agreement between the two types of EPID dosimetry for prostate radiotherapy. The AutoBeam treatment planning system produced VMAT plans together with corresponding predicted portal images, and a total of 46 sets of gantry-resolved portal images were acquired in 13 patients using an iViewGT portal imager. For the forward-projection method, each acquisition of gantry-resolved images was combined into a single integrated image and compared with the predicted image. For the back-projection method, iViewDose was used to calculate the dose distribution in the patient for comparison with the planned dose. A gamma index for 3% and 3 mm was used for both methods. The results were investigated by delivering the same plans to a phantom and repeating some of the deliveries with deliberately introduced errors. The strongest agreement between forward-and back-projection methods is seen in the isocentric intensity/dose difference, with moderate agreement in the mean gamma. The strongest correlation is observed within a given patient, with less correlation between patients, the latter representing the accuracy of prediction of the two methods. The error study shows that each of the two methods has its own distinct sensitivity to errors, but that overall the response is similar. The forward-and back-projection EPID dosimetry methods show moderate agreement in this series of prostate VMAT patients, indicating that both methods can contribute to the verification of dose delivered to the patient.
引用
收藏
页数:16
相关论文
共 50 条
  • [21] Development of geometrically ideal dose distribution as a reference for treatment planning in VMAT using filtered back-projection method
    Miki, Kentaro
    Hioki, Kazunari
    Nakashima, Takeo
    Saito, Akito
    Murakami, Yuji
    Kimura, Tomoki
    Nishibuchi, Ikuno
    Nagata, Yasushi
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 57 : 115 - 122
  • [22] Linking VMAT pre-treatment QA and patient treatment QA by EPID dosimetry
    Sassowsky, M.
    Manser, P.
    Fix, M. K.
    Frauchiger, D.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2012, 188 (08) : 754 - 754
  • [23] Use of a commercial EPID dosimetry system in VMAT pre-treatment QA
    Parisotto, M.
    Carbonini, C.
    Brambilla, M. G.
    Monti, A. F.
    Ferrari, M. B.
    Zanni, D.
    Torresin, A.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S164 - S165
  • [24] Time dependent pre-treatment EPID dosimetry for standard and FFF VMAT
    Podesta, Mark
    Nijsten, Sebastiaan M. J. J. G.
    Persoon, Lucas C. G. G.
    Scheib, Stefan G.
    Baltes, Christof
    Verhaegen, Frank
    PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (16): : 4749 - 4768
  • [25] A CONVOLUTIONAL FRAMEWORK FOR FORWARD AND BACK-PROJECTION IN FAN-BEAM GEOMETRY
    Zhang, Kai
    Entezari, Alireza
    2019 IEEE 16TH INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING (ISBI 2019), 2019, : 1455 - 1458
  • [26] FPGA-based Forward and Back-Projection Operators for Tomographic Reconstruction
    Jin, Kyungchan
    Song, Sangyup
    MEDICAL IMAGING 2013: PHYSICS OF MEDICAL IMAGING, 2013, 8668
  • [27] VMAT Case Study for Patient Specific Virtual Compensator Algorithm EPID In-Vivo Dosimetry
    Barbiere, J.
    Forbang, R. Teboh
    Ndlovu, A.
    MEDICAL PHYSICS, 2020, 47 (06) : E672 - E672
  • [28] Back-projection algorithm characteristic analysis in forward-looking bistatic SAR
    Zeng, DaZhi
    Hu, Cheng
    Zeng, Tao
    Long, Teng
    PROCEEDINGS OF 2006 CIE INTERNATIONAL CONFERENCE ON RADAR, VOLS 1 AND 2, 2006, : 147 - +
  • [29] In vivo dosimetry with electronic portal imaging device in VMAT for prostate cancer
    Inui, S.
    Ueda, Y.
    Ono, S.
    Ohira, S.
    Isono, M.
    Nitta, Y.
    Murata, S.
    Miyazaki, M.
    Teshima, T.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S942 - S942
  • [30] Clinical pilot study for EPID-based in vivo dosimetry using EPIgray™ for head and neck VMAT
    Sophie D. Halliday
    Rebecca A. Day
    Lynne Greig
    Robert J. W. Louwe
    Physical and Engineering Sciences in Medicine, 2022, 45 : 1335 - 1340