Time dependent pre-treatment EPID dosimetry for standard and FFF VMAT

被引:42
|
作者
Podesta, Mark [1 ]
Nijsten, Sebastiaan M. J. J. G. [1 ]
Persoon, Lucas C. G. G. [1 ]
Scheib, Stefan G. [2 ]
Baltes, Christof [2 ]
Verhaegen, Frank [1 ,3 ]
机构
[1] MUMC, MAASTRO, Dept Radiat Oncol, Maastricht, Netherlands
[2] Varian Med Syst iLab GmbH, Res & Dev, Dattwil Baden, Switzerland
[3] McGill Univ, Dept Oncol, Med Phys Unit, Montreal, PQ, Canada
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2014年 / 59卷 / 16期
关键词
VMAT; time dependent; EPID dosimetry; control point; VOLUMETRIC-MODULATED ARC; RESPONSE CHARACTERISTICS; LAG CORRECTION; PORTAL IMAGES; VERIFICATION; IMRT; THERAPY; BEAM; QA; DECONVOLUTION;
D O I
10.1088/0031-9155/59/16/4749
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Methods to calibrate Megavoltage electronic portal imaging devices (EPIDs) for dosimetry have been previously documented for dynamic treatments such as intensity modulated radiotherapy (IMRT) using flattened beams and typically using integrated fields. While these methods verify the accumulated field shape and dose, the dose rate and differential fields remain unverified. The aim of this work is to provide an accurate calibration model for time dependent pre-treatment dose verification using amorphous silicon (a-Si) EPIDs in volumetric modulated arc therapy (VMAT) for both flattened and flattening filter free (FFF) beams. A general calibration model was created using a Varian TrueBeam accelerator, equipped with an aS1000 EPID, for each photon spectrum 6 MV, 10 MV, 6 MV-FFF, 10 MV-FFF. As planned VMAT treatments use control points (CPs) for optimization, measured images are separated into corresponding time intervals for direct comparison with predictions. The accuracy of the calibration model was determined for a range of treatment conditions. Measured and predicted CP dose images were compared using a time dependent gamma evaluation using criteria (3%, 3 mm, 0.5 sec). Time dependent pre-treatment dose verification is possible without an additional measurement device or phantom, using the on-board EPID. Sufficient data is present in trajectory log files and EPID frame headers to reliably synchronize and resample portal images. For the VMAT plans tested, significantly more deviation is observed when analysed in a time dependent manner for FFF and non-FFF plans than when analysed using only the integrated field. We show EPID-based pre-treatment dose verification can be performed on a CP basis for VMAT plans. This model can measure pre-treatment doses for both flattened and unflattened beams in a time dependent manner which highlights deviations that are missed in integrated field verifications.
引用
下载
收藏
页码:4749 / 4768
页数:20
相关论文
共 50 条
  • [1] Use of a commercial EPID dosimetry system in VMAT pre-treatment QA
    Parisotto, M.
    Carbonini, C.
    Brambilla, M. G.
    Monti, A. F.
    Ferrari, M. B.
    Zanni, D.
    Torresin, A.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S164 - S165
  • [2] Linking VMAT pre-treatment QA and patient treatment QA by EPID dosimetry
    Sassowsky, M.
    Manser, P.
    Fix, M. K.
    Frauchiger, D.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2012, 188 (08) : 754 - 754
  • [3] EPID-based 3D dosimetry for pre-treatment FFF VMAT stereotactic body radiotherapy plan verification using dosimetry Check™
    Chendi, Agnese
    Botti, Andrea
    Orlandi, Matteo
    Sghedoni, Roberto
    Iori, Mauro
    Cagni, Elisabetta
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 81 : 227 - 236
  • [4] Confidence in VMAT EPID pre-treatment QA
    Monti, A. F.
    Trombetta, L.
    Brambilla, M. G.
    Carbonini, C.
    Torresin, A.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S170 - S170
  • [5] 3D EPID based dosimetry for pre-treatment verification of VMAT - methods and challenges
    Greer, P. B.
    7TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2013, 444
  • [6] EPID dosimetry - configuration and pre-treatment IMRT verification
    Grzadziel, Aleksandra
    Smolinska, Barbara
    Rutkowski, Roman
    Slosarek, Krzysztof
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2007, 12 (06) : 307 - 312
  • [7] EPID Dosimetry - Configuration and Pre-Treatment IMRT Verification
    Grzadziel, A.
    Smolinska, B.
    Rutkowski, R.
    Slosarek, K.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2006, VOL 14, PTS 1-6, 2007, 14 : 1932 - 1936
  • [8] Pre-treatment and in vivo EPID dosimetry for head and neck cancer
    McDermott, L.
    Wendling, M.
    Sonke, J.
    van Herk, M.
    Mijnheer, B.
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S307 - S307
  • [9] Using EPID pre-treatment dosimetry to solve a clinical problem
    McDermott, LN
    Wendling, M
    Stroom, J
    van Asselen, B
    van Herk, M
    Mijnheer, BJ
    RADIOTHERAPY AND ONCOLOGY, 2005, 76 : S65 - S66
  • [10] VMAT Pre-treatment Dosimetry: Dependence on Plan and Delivery Parameters
    Masi, L.
    Doro, R.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (02): : S898 - S899