Time dependent pre-treatment EPID dosimetry for standard and FFF VMAT

被引:42
|
作者
Podesta, Mark [1 ]
Nijsten, Sebastiaan M. J. J. G. [1 ]
Persoon, Lucas C. G. G. [1 ]
Scheib, Stefan G. [2 ]
Baltes, Christof [2 ]
Verhaegen, Frank [1 ,3 ]
机构
[1] MUMC, MAASTRO, Dept Radiat Oncol, Maastricht, Netherlands
[2] Varian Med Syst iLab GmbH, Res & Dev, Dattwil Baden, Switzerland
[3] McGill Univ, Dept Oncol, Med Phys Unit, Montreal, PQ, Canada
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2014年 / 59卷 / 16期
关键词
VMAT; time dependent; EPID dosimetry; control point; VOLUMETRIC-MODULATED ARC; RESPONSE CHARACTERISTICS; LAG CORRECTION; PORTAL IMAGES; VERIFICATION; IMRT; THERAPY; BEAM; QA; DECONVOLUTION;
D O I
10.1088/0031-9155/59/16/4749
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Methods to calibrate Megavoltage electronic portal imaging devices (EPIDs) for dosimetry have been previously documented for dynamic treatments such as intensity modulated radiotherapy (IMRT) using flattened beams and typically using integrated fields. While these methods verify the accumulated field shape and dose, the dose rate and differential fields remain unverified. The aim of this work is to provide an accurate calibration model for time dependent pre-treatment dose verification using amorphous silicon (a-Si) EPIDs in volumetric modulated arc therapy (VMAT) for both flattened and flattening filter free (FFF) beams. A general calibration model was created using a Varian TrueBeam accelerator, equipped with an aS1000 EPID, for each photon spectrum 6 MV, 10 MV, 6 MV-FFF, 10 MV-FFF. As planned VMAT treatments use control points (CPs) for optimization, measured images are separated into corresponding time intervals for direct comparison with predictions. The accuracy of the calibration model was determined for a range of treatment conditions. Measured and predicted CP dose images were compared using a time dependent gamma evaluation using criteria (3%, 3 mm, 0.5 sec). Time dependent pre-treatment dose verification is possible without an additional measurement device or phantom, using the on-board EPID. Sufficient data is present in trajectory log files and EPID frame headers to reliably synchronize and resample portal images. For the VMAT plans tested, significantly more deviation is observed when analysed in a time dependent manner for FFF and non-FFF plans than when analysed using only the integrated field. We show EPID-based pre-treatment dose verification can be performed on a CP basis for VMAT plans. This model can measure pre-treatment doses for both flattened and unflattened beams in a time dependent manner which highlights deviations that are missed in integrated field verifications.
引用
下载
收藏
页码:4749 / 4768
页数:20
相关论文
共 50 条
  • [21] Evaluation of a new Electronic Portal Imaging Device (EPID) for pre-treatment and in vivo dosimetry
    Reilly, A. J.
    Van Esch, A.
    Carver, A.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S182 - S183
  • [22] 4D dosimetry and its applications to pre-treatment quality control and real-time in vivo dosimetry of VMAT treatments
    Nordstrom, F.
    af Wetterstedt, S.
    Back, S. A. J.
    7TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2013, 444
  • [23] Comparing the error detection performance of Portal Dosimetry & PerFraction in pre-treatment VMAT QA
    Moloney, S.
    RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S734 - S735
  • [24] Pre-treatment VMAT verification with SunCHECK Fraction 0 and Varian Portal Dosimetry - a comparison
    Hoven, N. I.
    Kalsnes, J.
    Lonne, P.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S955 - S956
  • [25] Retrospective analysis of portal dosimetry pre-treatment quality assurance of intracranial SRS/SRT VMAT treatment plans
    Osei, Ernest
    Graves, Sarah
    Darko, Johnson
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2022, 21 (04) : 540 - 552
  • [26] A High Resolution Pre-Treatment VMAT QA Technique Based On EPID for Single Isocenter Multiple Mets Stereotactic Radiosurgery
    Wu, Q.
    Adamson, J.
    Liang, B.
    Liu, B.
    Zhou, F.
    MEDICAL PHYSICS, 2016, 43 (06) : 3685 - 3685
  • [27] FFF pre-treatment QA using TrueBeam Portal Dosimetry with DMI panel and comparison with PTW Octavius 1500
    Beyer, G.
    Gago, P.
    Ruiz, G.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S182 - S182
  • [28] Epid-based in vivo dosimetry for SBRT-VMAT treatment dose verification
    Cilla, S.
    Ianiro, A.
    Craus, M.
    Viola, P.
    Fidanzio, A.
    Azario, L.
    Greco, F.
    Grusio, M.
    Deodato, F.
    Macchia, G.
    Valentini, V.
    Morganti, A.
    Piermattei, A.
    RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S772 - S773
  • [29] Pre-treatment verification of the patient dose with an amorphous silicon EPID
    Wendling, M
    McDermott, LN
    Louwe, RJW
    van Herk, MB
    Mijnheer, B
    RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S113 - S114
  • [30] A model-based 3D patient-specific pre-treatment QA method for VMAT using the EPID
    McCowan, P. M.
    Asuni, G.
    van Beek, T.
    van Uytven, E.
    Kujanpaa, K.
    McCurdy, B. M. C.
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (04): : 1600 - 1612