Time dependent pre-treatment EPID dosimetry for standard and FFF VMAT

被引:42
|
作者
Podesta, Mark [1 ]
Nijsten, Sebastiaan M. J. J. G. [1 ]
Persoon, Lucas C. G. G. [1 ]
Scheib, Stefan G. [2 ]
Baltes, Christof [2 ]
Verhaegen, Frank [1 ,3 ]
机构
[1] MUMC, MAASTRO, Dept Radiat Oncol, Maastricht, Netherlands
[2] Varian Med Syst iLab GmbH, Res & Dev, Dattwil Baden, Switzerland
[3] McGill Univ, Dept Oncol, Med Phys Unit, Montreal, PQ, Canada
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2014年 / 59卷 / 16期
关键词
VMAT; time dependent; EPID dosimetry; control point; VOLUMETRIC-MODULATED ARC; RESPONSE CHARACTERISTICS; LAG CORRECTION; PORTAL IMAGES; VERIFICATION; IMRT; THERAPY; BEAM; QA; DECONVOLUTION;
D O I
10.1088/0031-9155/59/16/4749
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Methods to calibrate Megavoltage electronic portal imaging devices (EPIDs) for dosimetry have been previously documented for dynamic treatments such as intensity modulated radiotherapy (IMRT) using flattened beams and typically using integrated fields. While these methods verify the accumulated field shape and dose, the dose rate and differential fields remain unverified. The aim of this work is to provide an accurate calibration model for time dependent pre-treatment dose verification using amorphous silicon (a-Si) EPIDs in volumetric modulated arc therapy (VMAT) for both flattened and flattening filter free (FFF) beams. A general calibration model was created using a Varian TrueBeam accelerator, equipped with an aS1000 EPID, for each photon spectrum 6 MV, 10 MV, 6 MV-FFF, 10 MV-FFF. As planned VMAT treatments use control points (CPs) for optimization, measured images are separated into corresponding time intervals for direct comparison with predictions. The accuracy of the calibration model was determined for a range of treatment conditions. Measured and predicted CP dose images were compared using a time dependent gamma evaluation using criteria (3%, 3 mm, 0.5 sec). Time dependent pre-treatment dose verification is possible without an additional measurement device or phantom, using the on-board EPID. Sufficient data is present in trajectory log files and EPID frame headers to reliably synchronize and resample portal images. For the VMAT plans tested, significantly more deviation is observed when analysed in a time dependent manner for FFF and non-FFF plans than when analysed using only the integrated field. We show EPID-based pre-treatment dose verification can be performed on a CP basis for VMAT plans. This model can measure pre-treatment doses for both flattened and unflattened beams in a time dependent manner which highlights deviations that are missed in integrated field verifications.
引用
收藏
页码:4749 / 4768
页数:20
相关论文
共 50 条
  • [41] Multi-centre audit of VMAT planning and pre-treatment verification
    Jurado-Bruggeman, Diego
    Hernandez, Victor
    Saez, Jordi
    Navarro, David
    Pino, Francisco
    Martinez, Tatiana
    Alayrach, Maria-Elena
    Ailleres, Norbert
    Melero, Alejandro
    Jornet, Nuria
    RADIOTHERAPY AND ONCOLOGY, 2017, 124 (02) : 302 - 310
  • [42] A Statistical Approach to Manage Quality for Pre-Treatment Verification in IMRT/VMAT
    Jassal, K.
    Sarkar, B.
    Mohanti, B.
    Roy, S.
    Ganesh, T.
    Munshi, A.
    Chougule, A.
    Sachdev, K.
    MEDICAL PHYSICS, 2015, 42 (06) : 3348 - 3348
  • [43] Pre-treatment phantom dosimetry: effects in different phantoms and detectors
    Van Asselen, B.
    Wolthaus, J. W. H.
    Hackett, S. L.
    Kok, J. G. M.
    Woodings, S. J.
    Raaymakers, B. W.
    RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S311 - S312
  • [44] Sensitivity of ASi EPID to Delivery Errors for Pre Treatment Verification of Lung SBRT VMAT Plans
    Alharthi, T.
    Arumugam, S.
    Vial, P.
    Holloway, L.
    Thwaites, D.
    MEDICAL PHYSICS, 2018, 45 (06) : E324 - E324
  • [45] Comparison of time-benefit ratio between in vivo dosimetry and pre-treatment verification in IMRT breast treatment
    Bermejo, S.
    Ventosa, N.
    Domenech, X.
    Flores, R.
    De Segura, G. Gomez
    Jimenez, J.
    Nolla, X.
    Martin, L. San
    Carrasco, P.
    Latorre, A.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S898 - S899
  • [46] Pre-treatment and in-vivo dosimetry of Helical Tomotherapy treatment plans using the Dosimetry Check system
    Mezzenga, E.
    Cagni, E.
    Botti, A.
    Orlandi, M.
    Iori, M.
    JOURNAL OF INSTRUMENTATION, 2014, 9
  • [47] Feasibility of Time-Resolved DVH Based Treatment Delivery Verification of VMAT with EPID
    Greer, P.
    David, R.
    Wolf, J.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [48] Dosimetric pre-treatment IMRT verification with a new CCD camera based fluoroscopic EPID
    Pasma, KL
    Bolt, RA
    Mestrom, LHM
    Huizenga, H
    Bal, M
    Spies, L
    Visser, AG
    RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S115 - S115
  • [49] PRE-TREATMENT DOSIMETRIC VERIFICATION OF RAPIDARC TREATMENTS WITH AN ASI EPID AND THE GLAAS ALGORITHM.
    Nicolini, G.
    Fogliata, A.
    Vanetti, E.
    Clivio, A.
    Korreman, S.
    Cozzi, L.
    RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S25 - S25
  • [50] Pre-treatment and in vivo fetal dosimetry in brain radiotherapy treatment during pregnancy
    Valenti, M.
    Parisotto, M.
    Ardito, C.
    Puschi, M.
    Costantini, S.
    Cardinali, M.
    Maggi, S.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S752 - S753