FASCAL 2: a new NIST facility for the calibration of the spectral irradiance of sources

被引:17
|
作者
Yoon, HW [1 ]
Proctor, JE [1 ]
Gibson, CE [1 ]
机构
[1] Natl Inst Stand & Technol, Opt Technol Div, Gaithersburg, MD 20899 USA
关键词
D O I
10.1088/0026-1394/40/1/308
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The facility for automated spectroradiometric calibrations (FASCAL) is the primary facility for calibration of spectral irradiance and spectral radiance at NIST and has been in continuous use since the early 1970s. Due to the increasing demands for spectroradiometric calibration, especially for supporting the monitoring of global environmental changes, a new facility, FASCAL 2, dedicated to calibrating spectral irradiance has been built. This facility will enable faster responses to calibration requests and, ultimately, result in lower uncertainties in the disseminated spectral irradiances. The FASCAL 2 facility is designed with the objective of achieving a signal-to-noise ratio exceeding 1000: 1 from 250 nm to 2500 nm in a bandwidth of 4 nm to 8 nm when measuring a 1000 W FEL lamp at a distance of 50 cm with a receiving aperture of 1 cm(2). The facility will also be capable of calibrating deuterium lamps from 200 nm to 400 rim. The facility has six independent source stations, with four of the stations dedicated to measurements with spectral irradiance lamps and two stations reserved for the realization of spectral irradiance scales and checking the accuracy of automated wavelength. After verifying that the calibrations of spectral irradiance performed in FASCAL and FASCAL 2 agree within their combined uncertainties, FASCAL 2 will become the primary NIST facility for calibration of spectral irradiance.
引用
收藏
页码:S30 / S34
页数:5
相关论文
共 50 条
  • [41] SI-traceable Spectral Irradiance Radiometric Characterization and Absolute Calibration of the TSIS-1 Spectral Irradiance Monitor (SIM)
    Richard, Erik
    Harber, Dave
    Coddington, Odele
    Drake, Ginger
    Rutkowski, Joel
    Triplett, Matthew
    Pilewskie, Peter
    Woods, Tom
    REMOTE SENSING, 2020, 12 (11)
  • [42] Neutron calibration facility with radioactive neutron sources at KRISS
    Park, H.
    Kim, J.
    Choi, K. -O.
    RADIATION PROTECTION DOSIMETRY, 2007, 126 (1-4) : 159 - 162
  • [43] Facility for spectral radiance calibration at low light level
    Wu, Zhifeng
    Wang, Qiming
    Dai, Caihong
    Li, Ling
    Sun, Ruoduan
    Wang, Yanfei
    He, Sijie
    Guo, Qing
    OPTICS EXPRESS, 2024, 32 (08) : 14837 - 14846
  • [44] MEASUREMENT AND CALIBRATION OF LARGE-AREA ALPHA-PARTICLE SOURCES AT NIST
    UNTENWEGER, MP
    HUTCHINSON, JMR
    HEDGE, PA
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 339 (1-2): : 109 - 115
  • [45] ARTIFICIAL SUNTANNING - SPECTRAL IRRADIANCE AND HAZARD EVALUATION OF ULTRAVIOLET SOURCES
    GIES, HP
    ROY, CR
    ELLIOTT, G
    HEALTH PHYSICS, 1986, 50 (06): : 691 - 703
  • [46] Extension of the NIST spectral power-responsivity calibration service to 2500 nm
    Eppeldauer, G. P.
    Yoon, H. W.
    Zeng, J.
    Larason, T. C.
    Houston, J. M.
    Khromchenko, V.
    METROLOGIA, 2012, 49 (02) : S112 - S117
  • [47] Spectral irradiance measurement and actinic radiometer calibration for UV water disinfection
    Sperfeld, Peter
    Barton, Bettina
    Pape, Sven
    Towara, Anna-Lena
    Eggers, Jutta
    Hopfenmueller, Gabriel
    METROLOGIA, 2014, 51 (06) : S282 - S288
  • [48] SPECTRAL IRRADIANCE CALIBRATION IN THE INFRARED .3. THE INFLUENCE OF CO AND SIO
    COHEN, M
    WITTEBORN, FC
    CARBON, DF
    AUGASON, G
    WOODEN, D
    BREGMAN, J
    GOORVITCH, D
    ASTRONOMICAL JOURNAL, 1992, 104 (05): : 2045 - 2052
  • [49] Spectral photon irradiance and spectral photon radiance standards for the calibration of photon-counting imagers
    Xiao, Yike
    Gan, Haiyong
    Liu, Xiangliang
    He, Yingwei
    Wang, Weimin
    Xu, Yinuo
    Jing, Xufeng
    Shen, Changyu
    QUANTUM INFORMATION TECHNOLOGY (AOPC 2019), 2019, 11339
  • [50] New NIST-traceable standards for calibration and validation of DSC
    Archer, D. G.