Implant placement accuracy in total knee arthroplasty: validation of a CT-based measurement technique

被引:8
|
作者
Campanelli, Valentina [1 ]
Lozano, Rocio [1 ]
Akhlaghpour, Hosna [1 ]
Brar, Abheetinder S. [1 ]
Maislin, David [2 ,3 ]
Nedopil, Alexander J. [4 ]
Zuhars, Joel [1 ]
机构
[1] THINK Surg Inc, Fremont, CA 94538 USA
[2] Biomed Stat Consulting, Wynnewood, PA USA
[3] Musculoskeletal Clin Regulatory Advisers LLC, Washington, DC USA
[4] Univ British Columbia, Dept Orthopaed Surg, Vancouver, BC, Canada
关键词
Active robotics; implant placement accuracy; knee; total knee arthroplasty (TKA); validation; ALIGNMENT; POSITION; ORIENTATION; NAVIGATION;
D O I
10.21037/qims.2020.01.02
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: The primary goal of many computer-assisted surgical systems like robotics for total knee arthroplasty (TKA) is to accurately execute a preoperative plan. To assess whether the preoperative plan was executed accurately in 3D, one option is to compare the planned and postoperative implant placement using a preoperative and postoperative CT scan of the patient's limb. This comparison requires a 3D-to-3D surface registration between the preoperative and postoperative 3D bone models and between the planned and postoperative 3D implants. Hence, the present study aimed at validating this measurement technique by determining (I) the anatomical regions that result in the lowest 6-degree of freedom (DoF) errors for 3 D-to-3D surface registration of bone models, (II) the 6-DoF errors for 3D-to-3D surface registration of the implant models, and (III) the 6-DoF of the complete measurement technique. Methods: Four different regions of the femur were tested to determine which one would result in the most accurate 3D-to-3D registration of the bone models using 12 cadaveric lower limb specimens. Next, total knee arthroplasties were performed on six specimens, and the accuracy of the 3D-to-3D implant registration was evaluated against a gold standard registration performed using fiducial markers. Results: The most accurate 3D-to-3D bone registration was obtained when using the largest anatomical regions available after TKA, being the full 3D femur model or the femur model without the distal femur which resulted in root mean square errors within 0.2 mm for translations and 0.2 degrees for rotation. The accuracy of the 3D-to-3D femoral and tibial implant registration was within 0.7 mm for translations and 0.4 degrees-0.6 degrees for rotations, respectively. The accuracy for the overall procedure was within 0.9 mm and 0.6 degrees for both femur and tibia when using femoral regions resulting in accurate 3D-to-3D bone registration. Conclusions: In conclusion, this measurement technique can be used in applications where measurement errors up to 0.9 nun in translations and up to 0.6 degrees in rotations in component placement are acceptable.
引用
收藏
页码:475 / 484
页数:10
相关论文
共 50 条
  • [1] Robotic total knee arthroplasty - The accuracy of CT-based component placement
    Decking, J
    Theis, C
    Achenbach, T
    Roth, E
    Nafe, B
    Eckardt, A
    [J]. ACTA ORTHOPAEDICA SCANDINAVICA, 2004, 75 (05): : 573 - 579
  • [2] Effect of improved navigation performance on the accuracy of implant placement in total hip arthroplasty with a CT-based navigation system
    Nakahara, Ichiro
    Kyo, Takayuki
    Kuroda, Yasuo
    Miki, Hidenobu
    [J]. JOURNAL OF ARTIFICIAL ORGANS, 2018, 21 (03) : 340 - 347
  • [3] Effect of improved navigation performance on the accuracy of implant placement in total hip arthroplasty with a CT-based navigation system
    Ichiro Nakahara
    Takayuki Kyo
    Yasuo Kuroda
    Hidenobu Miki
    [J]. Journal of Artificial Organs, 2018, 21 : 340 - 347
  • [4] Navigation in total-knee arthroplasty -: CT-based implantation compared with the conventional technique
    Perlick, L
    Bäthis, H
    Tingart, M
    Perlick, C
    Grifka, J
    [J]. ACTA ORTHOPAEDICA SCANDINAVICA, 2004, 75 (04): : 464 - 470
  • [5] Accuracy of Implant Placement Utilizing Customized Patient Instrumentation in Total Knee Arthroplasty
    Bugbee, William D.
    Mizu-uchi, Hideki
    Patil, Shantanu
    D'Lima, Darryl
    [J]. ADVANCES IN ORTHOPEDICS, 2013, 2013
  • [6] The Accuracy of CT-Based Three-Dimensional Templating in Predicting Implant Sizes in Patients Undergoing Robot-Assisted Total Knee Arthroplasty
    Adkar, Neeraj
    Patil, Mangesh
    Vaidya, Swapnil
    Kumbar, Rajendra
    Kerhalkar, Ravi
    Mote, Girish
    Thareja, Satwik
    Sadalagi, Prajwal
    Bajwa, Supreet
    [J]. INDIAN JOURNAL OF ORTHOPAEDICS, 2024, 58 (10) : 1388 - 1394
  • [7] Accuracy of CT-based measurements of glenoid version for total shoulder arthroplasty
    Hoenecke, Heinz R., Jr.
    Hermida, Juan C.
    Flores-Hernandez, Cesar
    D'Lima, Darryl D.
    [J]. JOURNAL OF SHOULDER AND ELBOW SURGERY, 2010, 19 (02) : 166 - 171
  • [8] Accuracy of 2D CT-Based Measurements of Rotational Alignment of the Femoral Component for Total Knee Arthroplasty
    Park, Jun Young
    Kwon, Hyuck Min
    Cho, Byung Woo
    Lee, Woo-Suk
    Yang, Ick Hwan
    Park, Kwan Kyu
    [J]. YONSEI MEDICAL JOURNAL, 2021, 62 (05) : 439 - 445
  • [9] CT Validation of Intraoperative Implant Position and Knee Alignment as Determined by the MAKO Total Knee Arthroplasty System
    Sires, James D.
    Wilson, Christopher J.
    [J]. JOURNAL OF KNEE SURGERY, 2021, 34 (10) : 1133 - 1137
  • [10] Accuracy of CT-based patient-specific guides for total knee arthroplasty in patients with post-traumatic osteoarthritis
    Schotanus M.G.M.
    van Haaren E.H.
    Hendrickx R.P.M.
    Jansen E.J.P.
    Kort N.P.
    [J]. European Journal of Orthopaedic Surgery & Traumatology, 2015, 25 (8) : 1313 - 1320