Physical characterization and comparison of two commercially available micro-MLCs

被引:6
|
作者
Podder, Tarun K. [1 ]
Bednarz, Greg [2 ]
Yu, Yan [1 ]
Galvin, James M. [1 ]
机构
[1] Thomas Jefferson Univ Hosp, Dept Radiat Oncol, Kimmel Canc Ctr NCI Designated, Jefferson Med Coll, Philadelphia, PA 19107 USA
[2] Univ Pittsburgh, Med Ctr, Dept Radiat Oncol, UPMC Canc Ctr, Pittsburgh, PA 15232 USA
关键词
Micromulti-leaf collimator; Penumbra; Interleaf leakage; Tongue-and-groove effect; Focal spot; IMRT; MULTILEAF COLLIMATOR SYSTEM;
D O I
10.1016/j.ejmp.2010.01.005
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In this study, the physical characteristics (penumbra width variation with the source size and shape, interleaf leakage, transmission through the leaves, and the tongue-and-groove effect) of two linear accelerators (BrainLAB's Novalis and Elekta's Synergy-S Beam Modulator) have been investigated. For similar square fields (about 4.5 cm x 4.5 cm) with source-to-surface/skin-distance (SSD) ranging from 90 cm to 115 cm and measurements taken at the depth of D(max) = 1.5 cm for 6 MV photon beam. The Novalis MLC has penumbra width of 2.4 +/- 0.11 mm-2.8 +/- 0.11 mm at the leaf-end and 2.2 +/- 0.1 mm-2.7 +/- 0.1 mm on the leaf-side; and those for the Synergy-S MLC are 4.4 +/- 0.17 mm-5.2 +/- 0.2 mm and 3.0 +/- 0.12 mm-3.5 +/- 0.12 mm. Upon rotating the Synergy-S collimator by 90 degrees (i.e., shifting the leaf movement to the gun target direction), significant reduction of the leaf-end penumbra width (17%) and increase of leaf-side penumbra width (28%) suggest an elliptical shape of the radiation source spot. Similar rotation of the collimator yielded reduction of the penumbras on both leaf-end (34%) and leaf-side (28%) for Novalis, indicating that the Novalis has a more symmetric source size. For all the field sizes and settings, BrainLAB's Novalis mu MLC produce a smaller penumbra for simple square fields compared to the Elekta's Synergy-S. However, this difference became less pronounced for leaf-side penumbra and also for circular fields. The tongue-and-groove effect of the Novalis (23 +/- 0.9%) is slightly smaller than that of the Synergy-S (25 +/- 1%); while the interleaf leakage and leakage directly through leaves for Synergy-S (1.6 +/- 0.07% a 0.9 +/- 0.04%) are lower than that of Novalis (2 +/- 0.08% & 1.3 +/- 0.05%). (C) 2010 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:52 / 57
页数:6
相关论文
共 50 条
  • [31] PHYSICAL AND SENSORY CHARACTERISTICS OF COMMERCIALLY AVAILABLE FRANKFURTERS
    MATULIS, RJ
    MCKEITH, FK
    BREWER, MS
    JOURNAL OF FOOD QUALITY, 1994, 17 (04) : 263 - 271
  • [32] Comparison of two commercially available gamma interferon blood tests for immunodiagnosis of tuberculosis
    Dominguez, Jose
    Ruiz-Manzano, Juan
    De Souza-Galvao, Malu
    Latorre, Irene
    Mila, Celia
    Blanco, Silvia
    Angeles Jimenez, M.
    Prat, Cristina
    Lacoma, Alicia
    Altet, Neus
    Ausina, Vicente
    CLINICAL AND VACCINE IMMUNOLOGY, 2008, 15 (01) : 168 - 171
  • [33] Comparison of the interfraction repositioning accuracy of two commercially available thermoplastic mask systems
    Cuijpers, J.
    Bronkhorstz, Y.
    van Werkhoven, A.
    Seniorz, P.
    Doornert, P.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S260 - S260
  • [34] A Comparison Of Two Commercially Available Practical Cooling Interventions On Cycling Capacity In The Heat
    Tyler, Christopher J.
    Kiri, Victor
    Cuttell, Saul
    MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2012, 44 : 479 - 479
  • [35] Comparison of pulmonary vein isolation between two commercially available cryoballoon systems
    Karlo Filipovic
    Arian Sultan
    Jakob Lüker
    Jan-Hendrik van den Bruck
    Jonas Wörmann
    Cornelia Scheurlen
    Jan-Hendrik Schipper
    Sebastian Dittrich
    Daniel Steven
    International Journal of Arrhythmia, 25 (1)
  • [36] A comparison of a Bayesian population method with two methods as implemented in commercially available software
    Bennett, JE
    Wakefield, JC
    JOURNAL OF PHARMACOKINETICS AND BIOPHARMACEUTICS, 1996, 24 (04): : 403 - 432
  • [37] Characterization of two Si3N4 commercially available ceramics
    Bougoin, M.
    Key Engineering Materials, 1994, 89-91
  • [38] A COMPARISON OF THE PHYSICAL-PROPERTIES OF SOME SULFACETAMIDE EYE OINTMENTS COMMERCIALLY AVAILABLE IN THE UK
    FORD, JL
    RUBINSTEIN, MH
    DUFFY, TD
    IRELAND, DS
    INTERNATIONAL JOURNAL OF PHARMACEUTICS, 1982, 12 (01) : 11 - 18
  • [39] Comparison Of Commercially-Available Pedometers For Potential Use For Promotion Of Lifestyle Physical Activity
    Prieto-Centurion, V.
    Zaidi, F.
    Coultas, D. B.
    Casaburi, R.
    Rand, C. S.
    Krishnan, J. A.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2015, 191
  • [40] Comparison of Two Commercially Available EPID QA Systems Tested On Two Different EPID Systems
    Alexandrian, A.
    McConnell, K.
    Stathakis, S.
    MEDICAL PHYSICS, 2017, 44 (06) : 2934 - 2934