Enhancement of soft-tissue contrast in cone-beam CT using an anti-scatter grid with a sparse sampling approach

被引:17
|
作者
Cho, Sanghoon [1 ]
Lim, Sunho [1 ]
Kim, Changhwan [1 ]
Wi, Sunhee [1 ]
Kwon, Taejin [1 ]
Youn, Won Sik [2 ]
Lee, Sang Hyun [2 ]
Kang, Bo Sun [3 ]
Cho, Seungryong [1 ,4 ]
机构
[1] Korea Adv Inst Sci & Technol, Dept Nucl & Quantum Engn, Daejeon, South Korea
[2] JPI Healthcare Co Ltd, Dept Res & Dev, Ansan, South Korea
[3] Konyang Univ, Dept Radiol Sci, Daejeon, South Korea
[4] KAIST Inst Hlth Sci & Technol IT Convergence & Ar, Daejeon, South Korea
基金
新加坡国家研究基金会;
关键词
Cone-beam CT (CBCT); Anti-scatter grid; Scatter correction; COMPUTED-TOMOGRAPHY; ANTISCATTER GRIDS; RECONSTRUCTION; ALGORITHM; OPTIMIZATION; STATISTICS;
D O I
10.1016/j.ejmp.2020.01.004
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Anti-scatter grids suppress the scatter substantially thus improving image contrast in radiography. However, its active use in cone-beam CT for the purpose of improving contrast-to-noise ratio (CNR) has not been successful mainly due to the increased noise related to Poisson statistics of photons. This paper proposes a sparse-view scanning approach to address the above issue. Method: Compared to the conventional cone-beam CT imaging framework, the proposed method reduces the number of projections and increases exposure in each projection to enhance image quality without an additional cost of radiation dose to patients. For image reconstruction from sparse-view data, an adaptive-steepest-descent projection-onto-convex-sets (ASD POCS) algorithm regularized by total-variation (TV) minimization was adopted. Contrast and CNR with various scattering conditions were evaluated in projection domain by a simulation study using GATE. Then we evaluated contrast, resolution, and image uniformity in CT image domain with Catphan phantom. A head phantom with soft-tissue structures was also employed for demonstrating a realistic application. A virtual grid-based estimation and reduction of scatter has also been implemented for comparison with the real anti-scatter grid. Results: In the projection domain evaluation, contrast and CNR enhancement was observed when using an anti-scatter grid compared to the virtual grid. In the CT image domain, the proposed method produced substantially higher contrast and CNR of the low-contrast structures with much improved image uniformity. Conclusion: We have shown that the proposed method can provide high-quality CBCT images particularly with an increased contrast of soft-tissue at a neutral dose for image-guidance.
引用
收藏
页码:1 / 9
页数:9
相关论文
共 50 条
  • [41] Robust scatter correction method for cone-beam CT using an interlacing-slit plate
    Huang, Kui-Dong
    Xu, Zhe
    Zhang, Ding-Hua
    Zhang, Hua
    Shi, Wen-Long
    CHINESE PHYSICS C, 2016, 40 (06) : 99 - 106
  • [42] Improved Primary Modulation Method for Scatter Correction in Cone-Beam CT Using Local Filtration
    Zhu, L.
    MEDICAL PHYSICS, 2017, 44 (06) : 3283 - 3284
  • [43] X-ray scatter correction for cone-beam CT using moving blocker array
    Zhu, L
    Strobel, N
    Fahrig, R
    Medical Imaging 2005: Physics of Medical Imaging, Pts 1 and 2, 2005, 5745 : 251 - 258
  • [44] Minimizing scatter artifact in cone-beam CT reconstruction using both kilovoltage and megavoltage beams
    Zhang, J.
    Yin, F.
    MEDICAL PHYSICS, 2006, 33 (06) : 1998 - 1998
  • [45] Single-scan scatter correction for cone-beam CT using a stationary beam blocker: a preliminary study
    Niu, Tianye
    Zhu, Lei
    MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING, 2011, 7961
  • [46] Scatter Correction for Clinical Cone-Beam CT System Using An Optimized Stationary Beam Blocker in a Single Scan
    Liang, X.
    Luo, C.
    Zhang, Z.
    Xie, Y.
    Niu, T.
    MEDICAL PHYSICS, 2018, 45 (06) : E170 - E170
  • [47] Scatter Correction for Clinical Cone-Beam CT System Using An Optimized Stationary Beam Blocker with a Single Scan
    Liang, X.
    Gong, S.
    Zhou, Q.
    Zhang, Z.
    Xie, Y.
    Niu, T.
    MEDICAL PHYSICS, 2016, 43 (06) : 3457 - 3457
  • [48] Scatter correction for a clinical cone-beam CT system using an optimized stationary beam blocker in a single scan
    Liang, Xiaokun
    Jiang, Yangkang
    Zhao, Wei
    Zhang, Zhicheng
    Luo, Chen
    Xiong, Jing
    Yu, Shaode
    Yang, Xiaoming
    Sun, Jihong
    Zhou, Qinxuan
    Niu, Tianye
    Xie, Yaoqin
    MEDICAL PHYSICS, 2019, 46 (07) : 3165 - 3179
  • [49] Investigation of 2D anti-scatter grid implementation in a gantry-mounted cone beam computed tomography system for proton therapy
    Pyakurel, Uttam
    Zhang, Yawei
    Sabounchi, Ryan
    Bayat, Farhang
    Brousmiche, Sebastien
    Bryant, Curtis
    Mendenhall, Nancy
    Johnson, Perry
    Altunbas, Cem
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2025, 33
  • [50] Achieving high-resolution, soft-tissue imaging with cone-beam CT: A two-pronged approach for modulation of x-ray fluence and detector gain
    Graham, SA
    Siewerdsen, JH
    Moseley, DJ
    Keller, H
    Shkumat, NA
    Jaffray, DA
    Medical Imaging 2005: Physics of Medical Imaging, Pts 1 and 2, 2005, 5745 : 198 - 206