Comparison between pulse waveform analysis and thermodilution cardiac output determination in patients with severe pre-eclampsia

被引:29
|
作者
Dyer, R. A. [1 ]
Piercy, J. L. [1 ,2 ]
Reed, A. R. [1 ]
Strathie, G. W. [2 ]
Lombard, C. J. [5 ]
Anthony, J. A. [3 ,4 ]
James, M. F. [1 ]
机构
[1] Univ Cape Town, Dept Anaesthesia D23, ZA-7925 Cape Town, South Africa
[2] Univ Cape Town, Dept Crit Care, ZA-7925 Cape Town, South Africa
[3] Univ Cape Town, Dept Obstet & Gynaecol, ZA-7925 Cape Town, South Africa
[4] Groote Schuur Hosp, ZA-7925 Cape Town, South Africa
[5] MRC, Biostat Unit, Cape Town, South Africa
关键词
cardiac output; monitoring; pre-eclampsia; INJECTION; AGREEMENT; PRECISION; PRESSURE;
D O I
10.1093/bja/aeq292
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background. This study compared cardiac output (CO) measurements derived from pulse waveform analysis with values obtained by thermodilution (TD), in patients with postpartum complications of severe pre-eclampsia. Methods. Eighteen patients were recruited, 24-96 h post-delivery. After central venous calibration of the pulse waveform analysis monitor (LiDCOplus), CO readings were compared with those obtained by the TD method and repeated twice at 15 min intervals. The comparison was repeated after peripheral venous calibration. Further comparisons were made in eight patients at 120 and 240 min after peripheral venous calibration. Results. Data were pooled for measurements at 0, 15, and 30 min after calibration. For the comparison between TD and LiDCOplus using central venous calibration, TD exhibited a significant positive bias of 0.58 litre min(-1) [95% confidence interval (CI): 0.77 to 0.39]. After peripheral venous calibration, there was no significant bias [0.16 litre min(-1) (95% CI: -0.37 to 0.06)]. The estimated limits of agreement for central and peripheral venous calibrations were -2.12 to 0.96 and -1.50 to 1.20 litre min(-1), respectively. When comparing LiDCOplus and TD, there was no time-based effect at 120 or 240 min post-peripheral calibration. Conclusions. Central and peripheral venous calibrations of the LiDCOplus monitor were associated with clinically insignificant bias when compared with TD. Limits of agreement were within the recommendation of 30% for acceptance of a new CO technique when compared with current reference methods. This form of minimally invasive CO monitoring may have a valuable role in obstetric critical care.
引用
收藏
页码:77 / 81
页数:5
相关论文
共 50 条
  • [21] Delivery of patients with early onset, severe pre-eclampsia
    Hall, DR
    Odendaal, HJ
    Steyn, DW
    INTERNATIONAL JOURNAL OF GYNECOLOGY & OBSTETRICS, 2001, 74 (02) : 143 - 150
  • [22] Magnesium sulphate prophylaxis in patients with severe pre-eclampsia
    Odendaal, H
    MATERNAL-FETAL RISK IN GESTOSIS: PREECLAMPSIA AND PREGNANCY INDUCED HYPERTENSION, 1996, : 441 - 443
  • [23] Reproducibility of thermodilution cardiac output determination in critically ill patients: Comparison between bolus and continuous method
    LeTulzo, Y
    Belghith, M
    Seguin, P
    DallAva, J
    Monchi, M
    Thomas, R
    Dhainaut, JF
    JOURNAL OF CLINICAL MONITORING, 1996, 12 (05): : 379 - 385
  • [24] Continuous cardiac output by femoral arterial thermodilution calibrated pulse contour analysis: Comparison with pulmonary arterial thermodilution
    Goedje, O
    Hoeke, K
    Lichtwarck-Aschoff, M
    Faltchauser, A
    Lamm, P
    Reichart, B
    CRITICAL CARE MEDICINE, 1999, 27 (11) : 2407 - 2412
  • [25] Comparison of Cardiac Output in OPCAB: Bolus Thermodilution Technique versus Pulse Contour Analysis
    Sujatha, P.
    Mehta, Y.
    Dhar, A.
    Sarkar, D.
    Meharwal, Z. S.
    Trehan, N.
    ANNALS OF CARDIAC ANAESTHESIA, 2006, 9 (01) : 44 - 48
  • [26] Analysis of the accuracy of continuous thermodilution cardiac output measurement - Comparison with intermittent thermodilution and Fick cardiac output measurement
    Jacquet, L
    Hanique, G
    Glorieux, D
    Matte, P
    Goenen, M
    INTENSIVE CARE MEDICINE, 1996, 22 (10) : 1125 - 1129
  • [27] Assessment of a cardiac output device using arterial pulse waveform analysis, Vigileo™, in cardiac surgery compared to pulmonary arterial thermodilution
    Hamm, J. -B.
    Nguyen, B. -V.
    Kiss, G.
    Wargnier, J. -P.
    Jauffroy, A.
    Helaine, L.
    Arvieux, C. C.
    Gueret, G.
    ANAESTHESIA AND INTENSIVE CARE, 2010, 38 (02) : 295 - 301
  • [28] Comparison of Transpulmonary Thermodilution and Calibrated Pulse Contour Analysis with Pulmonary Artery Thermodilution Cardiac Output Measurements in Anesthetized Dogs
    Garofalo, N. A.
    Teixeira-Neto, F. J.
    Rodrigues, J. C.
    Cerejo, S. A.
    Aguiar, A. J. A.
    Becerra-Velasquez, D. R.
    JOURNAL OF VETERINARY INTERNAL MEDICINE, 2016, 30 (04) : 941 - 950
  • [29] MEASUREMENT OF CARDIAC OUTPUT IN PATIENTS WITH NON-TRAUMATIC INTRACRANIAL HEMORRHAGE: A COMPARISON BETWEEN INTERMITTENT THERMODILUTION TECHNIQUE AND UNCALIBRATED ARTERIAL PRESSURE WAVEFORM ANALYSIS
    Junttila, E.
    Koskenkari, J.
    Ohtonen, P.
    Ala-Kokko, T.
    INTENSIVE CARE MEDICINE, 2010, 36 : S97 - S97
  • [30] COMPARISON OF ARTERIAL WAVEFORM ANALYSIS AND THERMODILUTION FOR ESTIMATION OF CARDIAC OUTPUT IN A MODEL OF ABDOMINAL AORTIC ANEURYSM RUPTURE
    Suk, P.
    Cundrle, I., Jr.
    Hruda, J.
    Helan, M.
    Krbusik, J.
    Vlasin, M.
    Matejovic, M.
    Pavlik, M.
    Sramek, V.
    INTENSIVE CARE MEDICINE, 2010, 36 : S98 - S98