The dose response of normoxic polymer gel dosimeters measured using X-ray CT

被引:55
|
作者
Hill, B
Venning, A
Baldock, C
机构
[1] Canberra Hosp, Biomed Engn Serv, Med Phys Sect, Woden, ACT 2606, Australia
[2] Queensland Univ Technol, Sch Phys & Chem Sci, Brisbane, Qld, Australia
[3] Univ Sydney, Sch Phys, Inst Med Phys, Sydney, NSW 2006, Australia
来源
BRITISH JOURNAL OF RADIOLOGY | 2005年 / 78卷 / 931期
关键词
D O I
10.1259/bjr/46029447
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
X-ray CT was used to determine the dose response of normoxic polymer gel dosimeters. Normoxic polymer gel dosimeters were manufactured and irradiated up to 150 Gy. Up to 50 CT images were acquired oil a Toshiba Aquilion Multislice CT scanner using protocols for 80 kV and 135 kV to determine dose response. HU-dose sensitivity, the linear regression of data for the HU versus dose for the linear part of the curve LIP to 60 Gy was 0.38 +/- 0.07 HU Gy(-1) for 135 kV and 0.37 +/- 0.01 H U Gy(-1) for 80 kV. Dose resolution was found to be < 1.3 Gy for all absorbed dose range up to 70 Gy for 135 kV, similar to that measured previously for polyacrylamide gel (PAG). Although the HU-dose sensitivity was lower than that previously measured for PAG gel dosimeters it had a greater range of absorbed dose indicating that normoxic polymer gel dosimeters have potential in CT gel dosimetry.
引用
收藏
页码:623 / 630
页数:8
相关论文
共 50 条
  • [1] The effect of x-ray CT dose on normoxic polymer gel dosimeters
    Baxter, P.
    Jirasek, A.
    Hilts, M.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S56 - S56
  • [2] Optimization of the imaging protocol of an X-ray CT scanner for evaluation of normoxic polymer gel dosimeters
    Subramanian, Brindha
    Ravindran, Paul B.
    Baldock, Clive
    JOURNAL OF MEDICAL PHYSICS, 2006, 31 (02) : 72 - 77
  • [3] X-ray CT dose in normoxic polyacrylamide gel dosimetry
    Baxter, P.
    Jirasek, A.
    Hilts, M.
    MEDICAL PHYSICS, 2007, 34 (06) : 1934 - 1943
  • [4] Cosolvent-free polymer gel dosimeters with improved dose sensitivity and resolution for x-ray CT dose response
    Chain, J. N. M.
    Jirasek, A.
    Schreiner, L. J.
    McAuley, K. B.
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (07): : 2091 - 2102
  • [5] Investigation of X-ray CT dose in normoxic polyacrylamide gel dosimetry
    Jirasek, A.
    Baxter, P.
    Hilts, M.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2006, VOL 14, PTS 1-6, 2007, 14 : 1873 - +
  • [6] Dose rate properties of NIPAM-based x-ray CT polymer gel dosimeters
    Jirasek, A.
    Johnston, H.
    Hilts, M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (11): : 4399 - 4411
  • [7] A preliminary study of the novel application of normoxic polymer gel dosimeters for the measurement of CTDI on diagnostic x-ray CT scanners
    Hill, B
    Venning, AJ
    Baldock, C
    MEDICAL PHYSICS, 2005, 32 (06) : 1589 - 1597
  • [8] Polymer gel dosimeters with enhanced sensitivity for use in x-ray CT polymer gel dosimetry
    Jirasek, A.
    Hilts, M.
    McAuley, K. B.
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (18): : 5269 - 5281
  • [9] An experimental study of the dose response of polymer gel dosimeters imaged with x-ray computed tomography
    Trapp, JV
    Bäck, SÅJ
    Lepage, M
    Michael, G
    Baldock, C
    PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (11): : 2939 - 2951
  • [10] Characterization of the essential dosimetric properties of cosolvent-free polymer gel dosimeters: Recent progress in x-ray CT based normoxic polymer gel dosimetry
    Johnston, H.
    Hilts, M.
    Carrick, J.
    Jirasek, A.
    7TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2013, 444